All India Institute Of Medical Sciences (aiims) invites bids for Arterial Blood Gas Analyzer (V2) (Q2) in PATNA, BIHAR. Quantity: 1. Submission Deadline: 13-11-2025 14: 00: 00. Submit your proposal before the deadline.
Categories
3
Technical Specifications
1 Item
Item #1 Details
View Catalog| Category | Specification | Requirement |
|---|---|---|
| GENERAL | Product Description | Arterial Blood Gas Analyzer |
| PRODUCT INFORMATION | Type of blood gas analyser | Bench Top |
| PRODUCT INFORMATION | Type of Technology | Cartridge (Multiple-Use) |
| PRODUCT INFORMATION | Parameters measured by blood gas analyser | pCO₂, pO₂, Na⁺, K⁺, pH and calculated parameters like HCO₃⁻, Base excess, TCO₂ |
| PRODUCT INFORMATION | Additional Parameters measurable by the blood gas analyser | Cl-, iCa, Glu, Hct, Lacate, Co-oximetry |
| PRODUCT INFORMATION | Cardiac Parameters measurable | Not available |
| PRODUCT INFORMATION | Neonatal Parameters | nBil |
| PRODUCT INFORMATION | Monitor type | LCD, LED Or higher |
| PRODUCT INFORMATION | Type of printer | Inbuilt Printer |
| PRODUCT INFORMATION | Cartridge / reagent to be used can be stored mostly at room temperature between 2-30 Degree C, Automatic calibration (hint: seller shall inform the exact details to buyer at the time of making the supplies) | Yes |
| PRODUCT INFORMATION | Shall have patient Bar Code facility, Ability to run a wide range of patient panels on one system | Yes |
| PRODUCT INFORMATION | Capability to transfer patient data to LIS/HIS | Yes |
| PRODUCT INFORMATION | Sample Volume in micro litre | ≤ 200 |
| PRODUCT INFORMATION | Power source | AC Mains |
| PRODUCT INFORMATION | Battery backup provided | Through UPS for 30 mins |
| CERTIFICATION AND REPORTS | Compliance to Medical Device Rules (MDR) 2017 as amended till date | Yes |
| CERTIFICATION AND REPORTS | Availability of valid drug license for the product issued from the competent authority defined under Drugs and Cosmetic Act 1940 and Rules made there under as amended till date | Yes |
| CERTIFICATION AND REPORTS | Manufacturing unit certification | ISO:13485 (Latest) |
| CERTIFICATION AND REPORTS | Availability of Test Report for product as per Medical Device Rules (MDR) 2017 as amended till date | Yes |
| CERTIFICATION AND REPORTS | Electrical Safety Standards | IEC/EN 60601-1 or equivalent BIS Standard |
| CERTIFICATION AND REPORTS | Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission and/or along with supplies as per buyer requirement | Yes |
| WARRANTY & SHELF LIFE | Warranty (including electrode and membrane etc as applicable for the analyzer offered) in years | 5.0 |
| WARRANTY & SHELF LIFE | Shelf life of the reagent / cartridges as applicable in months | 6 |
| WARRANTY & SHELF LIFE | On-board shelf life of reagent in weeks | 4 |
| MISCELLANEOUS REQUIREMENTS | OEM/Reseller shall ensure uninterrupted availability of all spares for 10 years | Yes |
| MISCELLANEOUS REQUIREMENTS | Availability of toll free facility for technical support maintened by OEM or authorized agencies | Yes |
| MISCELLANEOUS REQUIREMENTS | User/Technical/Maintenance manuals to be supplied in English in hard and soft copy | Yes |
| MISCELLANEOUS REQUIREMENTS | Details of equipments and procedures required for local calibration and routine maintenance to be supplied and advanced maintenance task documentation also to be furnished | Yes |
| MISCELLANEOUS REQUIREMENTS | List of important spares and accessories, with their part numbers to be supplied to the buyer at the time of supplying the equipment | Yes |
| MISCELLANEOUS REQUIREMENTS | Installation and Demonstration of equipment and training to be provided after completing supplies before acceptance | Yes |
| MISCELLANEOUS REQUIREMENTS | Principal Manufacturer must have direct Presence/approved service center In India | Yes |
| MISCELLANEOUS REQUIREMENTS | Calibration certificates as per NABH requirement | Yes |
AI-Powered Bidder Prediction
Companies most likely to bid
Unlock Bidder Insights
AI predictions on likely bidders
Required Documents
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover
Additional Doc 1 (Requested in ATC)
Compliance of BoQ specification and supporting document *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
Similar Tenders
Arterial Blood Gas Analyzer (V2)
COAL INDIA LIMITED
BARDHAMAN, WEST BENGAL
Arterial Blood Gas Analyzer (V2)
Medical Education Department Uttar Pradesh
LUCKNOW, UTTAR PRADESH
Arterial Blood Gas Analyzer (V2)
Employees State Insuarnce Corporation
HYDERABAD, TELANGANA
Blood Collection Monitor (V2)
Medical Education Department Uttar Pradesh
LUCKNOW, UTTAR PRADESH
Blood Collection Monitor (V2)
Health and Family Welfare Department Assam
KAMRUP, ASSAM
Download FREE!
Access all tender documents at no cost
Documents 3
GeM-Bidding-8503258.pdf
Main Document
CATALOG-Specification-1
CATALOG Specification
GEM General Terms and Conditions Document
GEM_GENERAL_TERMS_AND_CONDITIONS
GeM-Bidding-8503258.pdf
Main Document
CATALOG-Specification-1
CATALOG Specification
GEM General Terms and Conditions Document
GEM_GENERAL_TERMS_AND_CONDITIONS