GEM

Procurement Opportunity: Neonatal Ventilators in AGRA, UTTAR PRADESH

Posted

15 Oct 2025, 02:47 pm

Deadline

05 Nov 2025, 03:00 pm

EMD

₹75,000

Value

₹75,00,000

Progress

Issue15 Oct 2025, 02:47 pm
AwardPending
Explore all 4 tabs to view complete tender details

Quantity

5

Bid Type

Two Packet Bid

Directorate Of Medical Education And Training invites bids for Neonatal Ventilators (Q2) in AGRA, UTTAR PRADESH. Quantity: 5. Submission Deadline: 05-11-2025 15: 00: 00. Submit your proposal before the deadline.

Technical Specifications 1 Item

Item #1 Details

View Catalog
Category Specification Requirement
GENERAL Product Description Neonatal Ventilator
PERFORMANCE PARAMETERS Patient Type Neonatal
PERFORMANCE PARAMETERS Tidal Volume Range In ml 2 to 300
PERFORMANCE PARAMETERS Respiratory Rate in Range,Breath/min 1 to 120
PERFORMANCE PARAMETERS Trigger Sensitivity Mechanism Flow
PERFORMANCE PARAMETERS FiO2 Range in % 21 to 100
PERFORMANCE PARAMETERS Inspiratory Flow Rate range L/min 0.4 to 100
PERFORMANCE PARAMETERS Flow Pattern/Waveform Adjustment Yes
PERFORMANCE PARAMETERS Inspiratory Pressure range in cmH20 0 to 50
PERFORMANCE PARAMETERS IE Ratio range 1:10 to 4:1
PERFORMANCE PARAMETERS Sigh Breath Function Yes
PERFORMANCE PARAMETERS Inspiratory Time in sec 0.15 to 5
PERFORMANCE PARAMETERS Flow Cycle in Percentage 0 to 45
PERFORMANCE PARAMETERS Pressure Control in cmH20 0 to 50
PERFORMANCE PARAMETERS Inspiration Pause time in sec 0.1 to 6
PERFORMANCE PARAMETERS Pressure Trigger cmH20 0.1 to 20
PERFORMANCE PARAMETERS Apnea time in sec 10 to 60
PERFORMANCE PARAMETERS Expiratory Time in millisec ≥ 200
PERFORMANCE PARAMETERS PEEP/CPAP Range in cmH20 0 to 45
PERFORMANCE PARAMETERS Flow Trigger in LPM 0.1 to 15
PERFORMANCE PARAMETERS Bias Flow in LPM Yes
PERFORMANCE PARAMETERS Volume Limit in Ltrs 0.02 to 2.5
PERFORMANCE PARAMETERS Leak Compensation Available Yes
PERFORMANCE PARAMETERS Auto 100% increase O2 Button Yes
PERFORMANCE PARAMETERS Inspiratory Pause Available Yes
PERFORMANCE PARAMETERS Expiratory Hold Available Yes, No
PERFORMANCE PARAMETERS Suction Control Option Available Yes
PERFORMANCE PARAMETERS Nasal CPAP Option Available Yes
PERFORMANCE PARAMETERS Auto PEEP Manoeuvre Yes
PERFORMANCE PARAMETERS Control Panel Lock Yes
PERFORMANCE PARAMETERS Facility for double lung ventilation Yes
PERFORMANCE PARAMETERS Integrated Air Compressor Yes, No
VENTILATION MODES PSV(Pressure supported ventilation) Yes
VENTILATION MODES P-CMV(Inspiration pressure control) Yes
VENTILATION MODES PC-AC With Expiratory Minimal volume Yes
VENTILATION MODES SIMV(Synchronized intermittent Mandatory Ventilation) with PCMV Yes
VENTILATION MODES ITT - Inspiratory Time Termination Yes
VENTILATION MODES CPAP:Continuos Positive Airway pressure Ventilation CPAP With backup Ventilation Yes
VENTILATION MODES High Frequency Oscillation Ventilator(HFOV) No
VENTILATION MODES NIV(Non Invasive Ventilation) Mode Yes
HIGH FREQUENCY OSCILLATIONS Mean Airway Pressure (MAP) 5 to 50
HIGH FREQUENCY OSCILLATIONS Frequency (Hz) 5 to 15
HIGH FREQUENCY OSCILLATIONS Amplitude 0 to 100
PATIENT ASSESSMENT TOOLS Maximum Waveform Displayed 3
PATIENT ASSESSMENT TOOLS Maximum Trending Time in Hours ≥72
PATIENT ASSESSMENT TOOLS Number of Loops 2 loops(p-v,f-v)
PATIENT ASSESSMENT TOOLS Capnography/Co2 Monitoring Yes, No
PATIENT ASSESSMENT TOOLS Volume Guarantee Yes, No
PATIENT ASSESSMENT TOOLS APRV Yes
PATIENT ASSESSMENT TOOLS Oxygen Therapy Yes
ADDITIONAL CAPABILITIES Nebulizer with particle size less than 3 micron Yes
ADDITIONAL CAPABILITIES ETCO2 Inbuilt available Yes, No
ADDITIONAL CAPABILITIES Humidifier Yes
MONITORED OR DISPLAYED Peak Inspiratory Pressure Monitored Yes
MONITORED OR DISPLAYED Mean Airway Pressure Yes
MONITORED OR DISPLAYED PEEP Pressure Yes
MONITORED OR DISPLAYED Tidal Volume Yes
MONITORED OR DISPLAYED Minute Volume Yes
MONITORED OR DISPLAYED Spontaneous Minute Volume Yes
MONITORED OR DISPLAYED FiO2 Monitored Yes
MONITORED OR DISPLAYED Display Respiratory rate Yes
MONITORED OR DISPLAYED Display Inspiratory Time Yes
MONITORED OR DISPLAYED Display Expiratory Time Yes
MONITORED OR DISPLAYED Display I:E Ratio Yes
ALARMS Low/High Fio2 Yes
ALARMS Low Minute Volume Yes
ALARMS High Minute Volume Yes
ALARMS Low Inspiratory Pressure Yes
ALARMS High Pressure Yes
ALARMS Loss of PEEP Yes
ALARMS Apnea Yes
ALARMS Continuous High pressure/Occlusion Yes
ALARMS Inverse I:E Ratio Yes
ALARMS High Respiratory Rate Yes
ALARMS High PEEP Yes
ALARMS Breathing Circuit Disconnect Yes
ALARMS Gas Supply Failure Yes
ALARMS Power Failure Yes
ALARMS Low Battery Yes
HUMIDIFIER Reusable Heated wire Provided for inspiratory Limb of breathing circuit & Water trap in expiratory limb available 2
HUMIDIFIER Neonatal Humidifier Jar 2
HUMIDIFIER Disposable Neonatal breathing circuit (dual heated wire) 20
HUMIDIFIER Auto Feed Humidification chamber Available Yes
HUMIDIFIER Disposable Neonatal breathing circuit with single heated wire provided for inspiratory Limb with Water trap in expiratory limb available 20
ADDITIONAL PARAMETERS Output Ports Type HDMI
ADDITIONAL PARAMETERS Remote Alarm Display Port Yes
ADDITIONAL PARAMETERS Reporting (Vent alarms and Patient Status) Yes
ADDITIONAL PARAMETERS View Reports on Display Yes
ADDITIONAL PARAMETERS Save Data to USB Yes
ADDITIONAL PARAMETERS Send Data Via Network Yes, No
ADDITIONAL PARAMETERS Display Available Yes
ADDITIONAL PARAMETERS Display Type LCD/LED/TFT
ADDITIONAL PARAMETERS Display Size in Inches ≥ 10, ≥ 12, ≥ 15 Or higher
ADDITIONAL PARAMETERS Can be used For Patient Transport Yes, No
ACCESSORIES Nasal CPAP Kit 10
ACCESSORIES Number of Disposable flow sensor 20
ACCESSORIES Number of Reusable flow sensor NA if Disposable
ACCESSORIES Trolley with ventillator with supporting arm from same manufacturer Yes
ACCESSORIES Autoclavable exhalation valve/expiratory cassette 2
ACCESSORIES Test Lung Available Yes
POWER REQUIREMENTS Type of power supply in VAC,Hz 220-240, 50Hz Single phase
POWER REQUIREMENTS Mains cable to be atleast 3m length Yes
POWER REQUIREMENTS Operating time of battery in min 120
CERTIFICATIONS Compliance to Medical Device Rules (MDR) 2017 as amended till date Yes
CERTIFICATIONS Availability of valid Medical Device license for the product issued from the competent authority defined under Drugs and Cosmetic Act 1940 and Rules made there under as amended till date Yes
CERTIFICATIONS Certification for manufacturing unit ISO:13485 (Latest)
CERTIFICATIONS Availability of Test Report for each supplied batch/product as per Medical Device Rule (MDR) 2017 as amended till date Yes
CERTIFICATIONS Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission or along with supplies as per buyer requirement Yes
CERTIFICATIONS Electrical Safety Standard IEC 60601-1-2 or Equivaent BIS
CERTIFICATIONS Safety and Essential Performance IS 13450 Part 2 Section 12
MISCELLANEOUS PARAMETERS OEM/Reseller (if supplied by reseller shall ensure uninterrupted availability of all spares for 10 years) Yes
MISCELLANEOUS PARAMETERS Availability of toll free facility for technical support maintened by OEM or authorized agencies Yes
MISCELLANEOUS PARAMETERS During the warranty period it shall be ensured that all the break down calls are attended with in 24 hrs and the complete details of service agents / contact details should be furnished to buyer and consignee at time of supplies Yes
MISCELLANEOUS PARAMETERS User/Technical/Maintenance manuals to be supplied in English in hard and soft copy Yes
MISCELLANEOUS PARAMETERS Details of equipments and procedures required for local calibration and routine maintenance to be supplied and advanced maintenance task documentation also to be furnished Yes
MISCELLANEOUS PARAMETERS List of important spares and accessories, with their part numbers and price lists to be supplied to the buyer at the time of supplying the equipment Yes
MISCELLANEOUS PARAMETERS Demonstration of equipment and training to be provided after completing supplies before acceptance Yes
MISCELLANEOUS PARAMETERS The Principal Manufacturer must have direct Presence/approved service center In India Yes
MISCELLANEOUS PARAMETERS Copies of reports and certifications to be furnished to buyer on demand at time of supplies Yes
WARRANTY Warranty in Years (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) 5.0

AI-Powered Bidder Prediction

Companies most likely to bid

Unlock Bidder Insights

AI predictions on likely bidders

Required Documents

1

Experience Criteria

2

Past Performance

3

Bidder Turnover

4

Certificate (Requested in ATC)

5

OEM Authorization Certificate

6

OEM Annual Turnover

7

Compliance of BoQ specification and supporting document *In case any bidder is seeking exemption from Experience / Turnover Criteria

8

the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer

Similar Tenders

Neonatal Ventilators

NA

PUNE, MAHARASHTRA

EMD: ₹80,000
Est: ₹40.0 L
View Details

Neonatal Ventilators

Employees State Insuarnce Corporation

WEST DELHI, DELHI

EMD: ₹6.8 L
Est: ₹2.3 Cr
View Details

Neonatal Ventilators

Health and Family Welfare Department Delhi

CENTRAL DELHI, DELHI

EMD: ₹56,000
View Details

Neonatal Ventilators

Health & Family Welfare Department Gujarat

,

EMD: ₹70,000
Est: ₹25.0 L
View Details

Portable Ventilators (V2)

Department of Health and Family Welfare

CENTRAL DELHI, DELHI

EMD: ₹40,000
View Details