A tender has been published for N/a Automated HPLC System for Separation of complex mixtures (Q3) in LUCKNOW, UTTAR PRADESH. Quantity: 1 by. Submission Deadline: 31-10-2025 18: 00: 00. Check eligibility and apply.
Tender Notice for Automated HPLC System for Separation of complex mixtures in LUCKNOW, UTTAR PRADESH
Posted
16 Oct 2025, 05:56 pm
Deadline
31 Oct 2025, 06:00 pm
EMD
₹80,000
Location
Progress
Quantity
1
Bid Type
Two Packet Bid
Categories
5
Technical Specifications
1 Item
Item #1 Details
View Catalog| Category | Specification | Requirement |
|---|---|---|
| Standard parameter | Instrument calibration standards | IEC 17025, NA |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Components of unit | Quaternary Low-Pressure Gradient pump& Parallel Double Plunger, Quaternary Low Pressure Gradient pump & Serial Double Plunger |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Number of solvents delivery | 4 |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Flow rate setting range in ml/min | 0.001 to 10 |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Flow rate accuracy | ± 1 %, ± 0.1%, ± 1 % or better |
| UV-Vis Detector | UV-Vis Detector provided | Yes, No Or higher |
| UV-Vis Detector | Light source | Deuterium lamp, NA if not provided, Tungsten Lamp |
| UV-Vis Detector | Spectrum Band width in nm | 4, 5, 6, 8, NA if not provided |
| UV-Vis Detector | Wavelength range in nm | 190 - 700, NA if not provided |
| UV-Vis Detector | Sampling Rate in Hz | 80, 90, 100, NA if not provided |
| UV-Vis Detector | The flow cell must be temperature Controlled | Yes, NA if not provided |
| UV-Vis Detector | Wavelength accuracy in nm | ± 1, NA if not provided |
| UV-Vis Detector | Wavelength reproducibility in nm | ± 0.1, NA if not provided |
| UV-Vis Detector | Drift in AU/Hour | less than 1 x 10-4, NA if not provided |
| UV-Vis Detector | Noise level in AU | ± 2.5 x 10-6, NA if not provided, ±5 x 10-6 |
| UV-Vis Detector | Ability to monitor and quantitative 2 wavelengths simultaneously | Yes, NA if not provided |
| UV-Vis Detector | Linearity in AU | 2.5 or more, NA if not provided |
| UV-Vis Detector | Temperature control provided for Flow cell | Yes, NA if not provided |
| UV-Vis Detector | Temperature control range for flow cell in Degree Celsius | 5 degree above room temp to 50 degree, NA if not provided |
| UV-Vis Detector | flow cell path length in mm | 10, NA if not provided |
| UV-Vis Detector | Flow cell volume in ?l | 10 - 15, NA if not provided, 16 - 20 |
| Auto Sample Injector with cooler | Injection Method | Total volume sample injection |
| Auto Sample Injector with cooler | Carry over of sample | 0.005% or lower, 0.025% or lower |
| Auto Sample Injector with cooler | Injection precision in RSD | less than 0.20 % RSD, less than 0.50 % RSD |
| Software | Operating Windows | Win 7 or latest, Win 10 or latest |
| Photo Diode Array (PDA) Detector | Provision of Photo diode array | Yes, No Or higher |
| Photo Diode Array (PDA) Detector | PDA detector Wavelength range in nm | 190- 800 or better, NA if not provided |
| Photo Diode Array (PDA) Detector | Number of Diodes | 512, 1024, NA if not provided |
| Photo Diode Array (PDA) Detector | PDA detector light source | Deuterium lamp and Tungsten lamp, NA if not provided, Deuterium lamp, Tungsten Lamp |
| Photo Diode Array (PDA) Detector | PDA detector Wavelength accuracy in nm | ± 1, NA if not provided |
| Photo Diode Array (PDA) Detector | Data Rate in Hz | 100, NA if not provided, <100 |
| Photo Diode Array (PDA) Detector | Noise in AU | less than ± 0.6 x 10-5 or lower, NA if not provided, less than ± 1 x 10-5 or lower |
| Photo Diode Array (PDA) Detector | PDA detector Drift in AU/Hour | less than 1 x 10-4, NA if not provided, less than 1 x 10-3 |
| Photo Diode Array (PDA) Detector | Temperature controlled flow cell in C | 5 degree above room temp to 50 degree, NA if not provided |
| Photo Diode Array (PDA) Detector | Flow cell | Flow cell, NA if not provided |
| Warranty & Reports | Warranty (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) | 2, 3, 4, 5, 1 Or higher |
| Warranty & Reports | Number of installations in Central /State/PSU Govt Hospitals | 0, <3, >3 |
| Warranty & Reports | Availability of toll free facility for technical support maintained by OEM or authorized agencies | Yes, No Or higher |
AI-Powered Bidder Prediction
Companies most likely to bid
Unlock Bidder Insights
AI predictions on likely bidders
Required Documents
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover
Additional Doc 1 (Requested in ATC)
Additional Doc 2 (Requested in ATC) *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
Similar Tenders
Automated HPLC System for Separation of complex mixtures
Education Department Chandigarh
CHANDIGARH, CHANDIGARH
Download FREE!
Access all tender documents at no cost
Documents 4
GeM-Bidding-8478869.pdf
Main Document
CATALOG-Specification-1
CATALOG Specification
Buyer uploaded ATC document
ATC
GEM General Terms and Conditions Document
GEM_GENERAL_TERMS_AND_CONDITIONS
GeM-Bidding-8478869.pdf
Main Document
CATALOG-Specification-1
CATALOG Specification
Buyer uploaded ATC document
ATC
GEM General Terms and Conditions Document
GEM_GENERAL_TERMS_AND_CONDITIONS