FTIR Spectrometer,Should collect IR spectrum in different modes such as transmittance, absorbance p
Central Manufacturing Technology Institute
BANGALORE, KARNATAKAPosted 16 Dec
GEMGoods4 Documents
Category: FTIR Spectrometer , Should collect IR spectrum in different
modes such as transmittance, absorbance pellets, stand-
alone films and ATR mode , Spectral range- Mid IR - 7800 to
350 cm-1 , Spectral resolution- At least 0.6 cm-1 or lower in
Mid IR range without using any external J-stop card ,
Wavenumber precision- lower than 0.002 cm-1 at 2000 cm-
1 or better , Wavenumber accuracy- lower than 0.06 cm-1
at 2000 cm-1 or better , Signal to Noise ratio , Beam splitter
, Detectors , ATR Assembly- Single-reflection ATR with
diamond , Calibration standards and performance
verification , Sample holder , Spectral library , Software for
data acquisition, processing, and evaluation. , PC
requirement , Work Bench for the system , Warranty