Progress
Public procurement opportunity for Junagadh Agricultural University Atomic Absorption Spectrophotometer (AAS) (Q3) in PORBANDAR, GUJARAT. Quantity: 1 issued by. Submission Deadline: 07-09-2026 09: 00: 00. View full details and respond.
Quantity
1
Bid Type
Two Packet Bid
Bid Validity
180 (Days)
Bid Type
Service
Evaluation
Total value wise evaluation
Inspection Required
No
Tech Clarification Time
2 Days
EMD Required
Yes
MII Purchase Preference
No
MSE Purchase Preference
No
MSE Exemption/Relaxation
No
Startup Exemption/Relaxation
No
Bid Splitting Applied
No
Arbitration Clause
No
Mediation Clause
No
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
No
Item Category
Atomic Absorption Spectrophotometer (AAS) (Q3)
Advisory Bank
State Bank of India
Max Delivery Days
15
Delivery Locations
1
Delivery Cities
Porbandar
Delivery Pincodes
360579
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Vimalkumar Natubhai Patel | 360579,COLLEGE OF AGRICULTURE, JAU, OPP. SAINT JOSEPH SCHOOL, ADITYANA ROAD, KHAPAT | Porbandar | Gujarat | 360579 | 1 | 15 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
OTHER
GEM_GENERAL_TERMS_AND_CONDITIONS
Extended Deadline
07-Sep-2026, 3:30 am
Opening Date
07-Sep-2026, 4:00 am
Discover companies most likely to bid on this tender
Experience Criteria
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
Additional Doc 1 (Requested in ATC) *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
| Category | Specification | Requirement |
|---|---|---|
| TYPE | Type of Atomic Absorption Spectrometer (AAS) | Double beam |
| TYPE | Monochromator type | Czerny-turner |
| TYPE | Detector Type | Photo Multiplier Tube (PMT) |
| SPECTRAL | Source Wavelength - Minimum | 185.0 |
| SPECTRAL | Source Wavelength - Maximum | 900.0 |
| SPECTRAL | Accuracy (+/-) | 0.1 |
| SPECTRAL | Repeatability | 0.1 |
| SPECTRAL | Diffraction grating (lines/mm) | 1800 |
| SPECTRAL | Wavelength setting and scanning | Automatic |
| SPECTRAL | Slit width Adjustment | Automatic |
| SPECTRAL | Resolution (nm/mm) | 0.1 |
| SPECTRAL | Bandwidth (nm) | 0.1 - 2.0 |
| SPECTRAL | Focal length (mm) | 250 to 299 |
| LIGHT SOURCE | Type of HCL | Both Single and Mulitiple Element |
| LIGHT SOURCE | Number of Single element HCLs supplied (inclusive in the scope of supply) | 6 |
| LIGHT SOURCE | Number of multiple element (alloy) HCLs supplied (inclusive in the scope of supply) | N/A |
| LIGHT SOURCE | Measuring modes (K factor and standards) | Transmittance, Absorbance, Concentration (Wavelength, time), Emission |
| LIGHT SOURCE | Detectable Elements using single element HCL | Cu, Fe, Zn, Mn, Mg, B, K, Mo, Ca, As, Na, Al, Au, Be, Ba, Bi, Cd, Ce, Co, Cr, lr, Li, Ni, Pb, Pd, Pt, Rh, Se, Si, Sn, Sr, Te, Ti, Tl, V, Zr, Hg, Sb, Ag |
| LIGHT SOURCE | Detectable Elements using multi element HCL | Na-K, Ca-Mg, Si-Al, Fe-Ni, Sr-Ba, Al-Ca-Mg, Ca-Mg-Zn, Cu-Mo-Co-Zn, Cd-Cu-Pb-Zn, Cu-Fe-Zn-Mn, Co-Cr-Cu-Fe-Mn-Ni, Cd - Hg - Pb |
| LIGHT SOURCE | Lifespan of Single element lamp (mA-Hr) | 5000 |
| LIGHT SOURCE | Lifespan of multi-element/Alloys lamp (mA-Hr) | N/A |
| LIGHT SOURCE | Number of sockets on Turret | 6 |
| ATOMIZER | Type of atomizer | Flame Burner |
| ATOMIZER | Nebulizer | Pt/Ir nebulizer |
| ATOMIZER | Background Correction | D2 Lamp method |
| ATOMIZER | Burner diameter (mm) | 100 |
| ATOMIZER | Time for analysis (sample) | 30.0 |
| GENERIC | Display | through PC |
| GENERIC | Display type | LED |
| GENERIC | Test result printing | Through external printer |
| GENERIC | If inbuilt printer, Type of printer | N/A |
| GENERIC | Safety sensors | Pressure sensor, Flame Sensor, Flame Ignition Sensor, Burner interlock, N/A, Power failure protection, Nebulizer Sensor, Auto ignition |
| GENERIC | Burner movement (lateral,horizontal & height) | Automatic |
| OPERATING CONDITIONS | Minimum Operating Temperature | 10.0 |
| OPERATING CONDITIONS | Maximum Operating Temperature | 35.0 |
| OPERATING CONDITIONS | Operating Humidity (RH) (%) at 40 degree c | 80.0 |
| ACCESSORIES | Type of gas cylinders | Acetylene |
| ACCESSORIES | Type of Regulators supplied | Without Heater |
| ACCESSORIES | Number of Regulators Supplied (inclusive in the scope of supply) | 3 |
| ACCESSORIES | Number of Acetylene Cylinders supplied (inclusive in the scope of supply) | 1 |
| ACCESSORIES | Capacity of each Acetylene cylinder (Litres) | 47 |
| ACCESSORIES | Number of Nitrous oxide Cylinders supplied (inclusive in the scope of supply) | 1 |
| ACCESSORIES | Capacity of each Nitrous oxide cylinder (Litres) | 47 |
| ACCESSORIES | Gas cylinders having PESO certificate | Yes |
| ACCESSORIES | Other accessories (inclusive in the scope of supply) | Gas flow control box, Oil Moisture Filters, SS Exhaust fume hood with blower, Schematic diagram and instruction manual, Hydride generator, Auto sampler, N/A |
| ACCESSORIES | Suitable software for connecting PC & Printer (inclusive in the scope of supply) | Yes |
| WARRANTY / AVAILABILITY OF SPARES | warranty (Year) | 2.0 |
| WARRANTY / AVAILABILITY OF SPARES | Spares and Consumables inclusive in the scope of supply | Yes |
| CERTIFICATION | Availability of UL/CE Certification as per EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control, and laboratory use General requirements | Yes |
| CERTIFICATION | Conformance to EMC/EMI as per EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use | Yes |
| CERTIFICATION | Availability of Test Reports from Central Govt / NABL approved / ILAC accredited Lab to prove conformity to the specification | No |