Digital Radiography System (V2)
N/a
LUCKNOW, UTTAR PRADESH
A tender has been published for The issuing authority Digital Radiography System (V2) in RANCHI, JHARKHAND. Quantity: 3 by. Submission Deadline: 03-03-2026 17: 00: 00. Check eligibility and apply.
Main Document
CATALOG Specification
GEM_GENERAL_TERMS_AND_CONDITIONS
N/a
LUCKNOW, UTTAR PRADESH
N/a
VELLORE, TAMIL NADU
Employees State Insurance Corporation
Civil Hospital Amdavad
AHMEDABAD, GUJARAT
Hll Lifecare Limited
THIRUVANANTHAPURAM, KERALA
Tender Results
Loading results...
| Category | Specification | Requirement |
|---|---|---|
| General | Product | Digital Radiography System |
| General | Description | Digital Radiographic system with flat panel detector |
| Table and Table top parameters | Type of table | Elevating Type, Floating Top Type |
| Table and Table top parameters | Type of table movement | Motorized |
| Table and Table top parameters | Number of movements of Table | 6 Way |
| Table and Table top parameters | Portability of table | No, Yes |
| Table and Table top parameters | Longitudinal Movement (with tolerance ± 10%) | ±50 |
| Table and Table top parameters | Lateral movement (with tolerance ± 10%) | ±15 |
| Table and Table top parameters | Length x Width (± 5%) in cm | 210 x 70 |
| Table and Table top parameters | X Ray density | < 1 mm A I |
| Table and Table top parameters | Minimum and maximum adjustable heights (± 10%) | 55-85 |
| Table and Table top parameters | Load bearing capacity of table in kg | 300 or more |
| Table Bucky System | Type of Table bucky movement | Auto tracking and auto synchronization, Auto tracking and auto positioning |
| Table Bucky System | Bucky rotatable | No, Yes |
| Table Bucky System | Size in inches | 17x17, 14x17 |
| Table Bucky System | AEC | 3 Field or higher |
| Table Bucky System | Grid Ratios | 10:1 or higher |
| Table Bucky System | Lines in mm | >=92 |
| Table Bucky System | Longitudinal travel | Motorized |
| Upright Bucky | Type | Motorized, Motorized tilting |
| Upright Bucky | Sizes in inches | 17x17, 14x17 |
| Upright Bucky | A E C | 3 Field or higher |
| Upright Bucky | Movable grid with Grid Ratios | 10:1 or higher |
| Upright Bucky | Line in inches | >=92 |
| Upright Bucky | Tilt | Yes |
| Digital Detector | Radiography Mode | Digital |
| Digital Detector | Number of detectors | Dual |
| Digital Detector | Detector Material | Cesium Iodide |
| Digital Detector | Detector Size in inches | 17 X 17 and 17 x 17, 14 x 17 and 14 x 17 |
| Digital Detector | Matrix size | 3kx3k, 2.8k x 2.8k |
| Digital Detector | Pixel size in µm | <=175 |
| Digital Detector | Grey Scale resolution, bits/pixel | 14 or better |
| Digital Detector | DQE at 0 LP/mm | 65% or more |
| Control panel for image acquisition | Image preview time in seconds | <=8 |
| Control panel for image acquisition | Control panel | Microprocessor controlled |
| Control panel for image acquisition | Automatic parameter selection | Yes |
| Control panel for image acquisition | Control panel entry | Touch screen, Alphanumeric key board |
| Control panel for image acquisition | Virtual grid software | Yes |
| Control panel for image acquisition | Grid removable | Yes |
| Control panel for image acquisition | Anatomic specific post processing | Yes |
| X Ray Generator and Tubes | High Frequency X-ray Generator power in KW | 80 KW |
| X Ray Generator and Tubes | Maximum mA at 100 KVP | 800 |
| X Ray Generator and Tubes | Focal spot | 0.6 and 1.2 |
| X Ray Generator and Tubes | Heat Capacity, KHU | 400-600 |
| X Ray Generator and Tubes | Control panel and X Ray tube | Yes |
| X Ray Generator and Tubes | Post Processing Features | Zoom, contrast and brightness, adjustment, panning annotate, mark and reporting |
| X Ray Generator and Tubes | Automatic tube positioning | Yes |
| Tube suspension | Tube mounted X Ray control | Yes |
| Tube suspension | Type of suspension | Ceiling |
| Tube suspension | Collimator Type | Automatic |
| Workstation | Monitor size in inches | 19, 21 |
| Workstation | Processor | i 3, i 5 |
| Workstation | RAM in GB | 8 |
| Workstation | HDD | 1 TB |
| Workstation | DiCOM compatibility of work station | Yes |
| Workstation | Work station should be fully network ready and it should transfer images and patient data from and to hospital network using LAN connectivity or wireless LAN | Yes |
| Workstation | Work station integrated with system | Yes |
| Workstation | Work station with advance measurement tool should be provided | Yes |
| Workstation | DICOM Printing software to be provided | Yes |
| Workstation | Patient study on film | Yes |
| Accessories | Hand grips | Yes |
| Accessories | Foot switches for adjusting longitudinal/side to side movements, locking, and light adjustment | Yes |
| Accessories | Patient fixing belts and compression device (for performing excretory urography, etc) | Yes |
| Accessories | Stitching Hardware and Software | Yes |
| Accessories | Integration | DICOM 3.0 storage SOP Classes |
| Accessories | DAP Meter provided | Yes |
| Accessories | Lead Apron | Yes |
| Accessories | Light weight ‘zero lead’ radiation protection aprons provided | Yes |
| Accessories | Lead Aprons - Full body and Skirt type | 2 no. each (0.5pb) |
| Accessories | Thyroid shields provided | Yes |
| Accessories | Gonadsheet Male provided | Yes |
| Accessories | Gonad sheet Female provided | Yes |
| Accessories | Wall mounted stand for radiation protection aprons supplied | Yes |
| Accessories | Lead glass window provided | Yes |
| Accessories | Footsteps for the table provided | Yes |
| Accessories | Power | 440 V 50 HZ 3 Phase 16 Ampls socket |
| Accessories | Voltage stablizer servo controlled in KVA | >=50 |
| Certifications | Compliance to Medical Device Rules (MDR) 2017 as amended till date | Yes |
| Certifications | Availability of valid Medical Device license for the product issued from the competent authority defined under Drugs and Cosmetic Act 1940 and Rules made there under as amended till date | Yes |
| Certifications | Certification for manufacturing unit | ISO:13485 (Latest) |
| Certifications | Availability of Test Report for each supplied batch/product as per Medical Device Rule (MDR) 2017 as amended till date | Yes |
| Certifications | Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission or along with supplies as per buyer requirement | Yes |
| Certifications | AERB Type Approved Equipment | Yes |
| Warranty | Warranty in Years | 2, 3, 4, 5 Or higher |
| Miscellaneous Parameters | OEM/Reseller ( if supplied to reseller shall ensure uninterputed availabilty of all spares for 10 years) | Yes |
| Miscellaneous Parameters | Availability of toll free facility for technical support maintained by OEM or authorized agencies | Yes |
| Miscellaneous Parameters | Availability of toll free facility for technical support maintened by OEM or authorized agencies | Yes |
| Miscellaneous Parameters | User/Technical/Maintenance manuals to be supplied in English in hard and soft copy | Yes |
| Miscellaneous Parameters | Demonstration of equipment and training to be provided after completing supplies before acceptance | Yes |
| Miscellaneous Parameters | Copies of reports and certifications to be furnished to buyer on demand at time of supplies | Yes |
| Miscellaneous Parameters | QA, QC and other tests as per NABH shall be carried out by OEM or authorized agency of OEM | Yes |
| Miscellaneous Parameters | Copies of all calibration and test reports to be submitted at the time of Supply of Machine | Yes |
Discover companies most likely to bid on this tender
N/a
📍 NAINITAL, UTTARAKHAND
N/a
📍 ALMORA, UTTARAKHAND
Sign up now to access all documents
Main Document
CATALOG Specification
GEM_GENERAL_TERMS_AND_CONDITIONS