Sher-e-kashmir University Of Agricultural Sciences And Technology Of Jammu has released a public tender for Gas Chromatography (Q3) in JAMMU, JAMMU & KASHMIR. Quantity: 1. Submission Deadline: 02-09-2026 13: 00: 00. Download documents and apply online.
Quantity
1
Bid Type
Two Packet Bid
Bid Validity
30 (Days)
Bid Type
Service
Evaluation
Total value wise evaluation
Inspection Required
No
RA Qualification Rule
H1-Highest Priced Bid Elimination
Tech Clarification Time
2 Days
EMD Required
Yes
MII Purchase Preference
No
MSE Purchase Preference
No
MSE Exemption/Relaxation
No
Startup Exemption/Relaxation
No
Bid Splitting Applied
No
Min. Avg. Annual Turnover
10
OEM Avg. Turnover
10
Past Performance
10 %
Experience Required
3 Year (s)
Arbitration Clause
No
Mediation Clause
No
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
Yes
Item Category
Gas Chromatography (Q3)
Advisory Bank
State Bank of India
ePBG Percentage
0.01%
ePBG Duration (Months)
12
Payment Timelines
Payments shall be made to the Seller within 45 days of issue of consignee receipt-cum-acceptance certificate (CRAC) and on-line submission of bills (This is in supersession of 10 days time as provided in clause 12 of GeM GTC)
Max Delivery Days
15
Delivery Locations
1
Delivery Cities
Jammu
Delivery Pincodes
180009
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Devinder Sharma | 180009,Sher-e-Kashmir University of Agricultural Sciences and Technology of Jammu Main Campus- Chatha, Jammu (J&K) | Jammu | Jammu And Kashmir | 180009 | 1 | 15 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
OTHER
GEM_GENERAL_TERMS_AND_CONDITIONS
Discover companies most likely to bid on this tender
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover
Compliance of BoQ specification and supporting document *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
| Category | Specification | Requirement |
|---|---|---|
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "GC Detectors inclusive in the scope of supply (Select Only those Detectors, which are part of supply)" | Flame Ionization Detector (FID), Thermal Conductivity Detector (TCD or Hot Wire Detector), Electron Capture Detector (ECD) |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of FID in the System - inclusive in the scope of Supply (Select '0' if this Detector is not a part of supply)" | 1 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of TCD in the System - inclusive in the scope of Supply (Select '0' if this Detector is not a part of supply)" | 1 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of ECD in the System - inclusive in the scope of Supply (Select '0' if this Detector is not a part of supply)" | 0 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of NPD/FTD in the System - inclusive in the scope of Supply (Select '0' if this Detector is not a part of supply)" | 0 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of FPD in the System - inclusive in the scope of Supply (Select '0' if this Detector is not a part of supply)" | 0 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of PID in the System - inclusive in the scope of Supply (Select '0' if this Detector is not a part of supply)" | 0 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of PFPD in the System - inclusive in the scope of Supply (Select '0' if this Detector is not a part of supply)" | 0 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | Number of Inlets GC can Simultaneously Support | 2 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | Number of Detectors (Ports) GC can Simultaneously Support | 2 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | Number of Capillary Columns provided in GC - inclusive in the scope of Supply | 2 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | Type of Capillary Column | NA |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "GC Capillary Column Length (Hint:-Select '0' if Not Applicable)" | 0.0 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "GC Capillary Column Internal Diameter (Hint:-Select '0' if Not Applicable)" | 0.0 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "GC Capillary Column Film Thickness (Hint:-Select '0' if Not Applicable)" | 0.0 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of Packed Columns provided in GC - inclusive in the scope of Supply (Hint:-Select '0' if Not Applicable)" | 0 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "GC Packed Column Length (Hint:-Select '0' if Not Applicable)" | 0.0 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "GC Packed Column Internal Diameter (Hint:-Select '0' if Not Applicable)" | 0 |
| PERFORMANCE | "GC Detector Minimum Detection Level (MDL) " | < 1.4 pg C/s for FID, < 3.0 pg C/s for FID < 400 pg/ml for TCD |
| PERFORMANCE | GC Detector Linear Dynamic Range (± 10%) | >= 10e7 for FID, >= 10e5 for TCD, >= 5x10e4 for ECD, >= 10e3 for NPD/FTD, >= 10e3 for FPD |
| PERFORMANCE | GC Detector Data acquisition rate | 50 Hz or more |
| COLUMN OVEN (GAS CHROMATOGRAPH - GC) | Maximum Operating Temperature (GC Oven) | upto 450 deg. C (2) Or higher |
| COLUMN OVEN (GAS CHROMATOGRAPH - GC) | Oven cool down time from Max operating temperature to 50 Deg c | 3.5 Or lower |
| COLUMN OVEN (GAS CHROMATOGRAPH - GC) | Number of Oven Ramps / Plateaus for Temperature programming | 9/10 or more |
| COLUMN OVEN (GAS CHROMATOGRAPH - GC) | Temperature Ramp Rate - Maximum | 120 Deg. C/minute or better |
| COLUMN OVEN (GAS CHROMATOGRAPH - GC) | Retention time repeatability | 0.01 |
| COLUMN OVEN (GAS CHROMATOGRAPH - GC) | Temperature set point resolution (±) | 0.1 |
| PNEUMATIC CONTROL | "Type of Inlet Pneumatics (Select applicable parameters only)" | Electronic Pneumatic Control (EPC), Programmed Pneumatic Control (PPC), Advanced Flow Controller (AFC), Advanced Pressure Controller (APC) |
| PNEUMATIC CONTROL | Number of Injection Ports | 2 |
| PNEUMATIC CONTROL | Maximum Injection Port heating Temperature | upto 450 deg. C (2) Or higher |
| PNEUMATIC CONTROL | Pressure Range | upto 120 psi (2) Or higher |
| PNEUMATIC CONTROL | Pressure set point Resolution (±)(psi) | 0.001 Or lower |
| PNEUMATIC CONTROL | Type of Sample Injection Port | split / Splitless, Purged packed Inlet |
| PNEUMATIC CONTROL | Split Ratio | 6000:1 |
| PNEUMATIC CONTROL | Modes of Operation | Constant flow, Constant Pressure, Programmable Flow, Programmable Pressure, Large volume - 200 microliters or more for capillary column |
| AUTO INJECTOR / AUTO SAMPLER | Compatible Liquid Auto Sampler - inclusive in the scope of supply | Yes |
| AUTO INJECTOR / AUTO SAMPLER | Capacity of the Liquid Auto Sampler (Number of Vials) | 150 Or higher |
| AUTO INJECTOR / AUTO SAMPLER | Volume of each Vial of the Liquid Auto Sampler | 2.0 |
| AUTO INJECTOR / AUTO SAMPLER | Number of Injection Ports in the Liquid Auto Sampler | 1 |
| AUTO INJECTOR / AUTO SAMPLER | Number of Channels in the Liquid Auto Sampler | 2 |
| AUTO INJECTOR / AUTO SAMPLER | Minimum Volume of the injected liquid sample | 0.1 |
| AUTO INJECTOR / AUTO SAMPLER | Maximum Volume of the injected liquid sample | 5.0 |
| AUTO INJECTOR / AUTO SAMPLER | Operating Voltage of the Liquid Auto Sampler | 230 V ± 10% |
| AUTO INJECTOR / AUTO SAMPLER | "Liquid Auto Sampler Features (Select applicable features Only)" | Fully automated Liquid Auto Sampler, Injects with variable speed & varying syringe sizes, Operates through Software (Software inclusive in the scope of supply) |
| HEAD SPACE SAMPLER | Compatible Head Space Sampler- inclusive in the scope of supply | Yes |
| HEAD SPACE SAMPLER | Capacity of the Head Space Sampler Carousel (number of Vials) | 12 or better |
| HEAD SPACE SAMPLER | Head Space Sampler Temperature Range | Upto 200 deg. C or better |
| HEAD SPACE SAMPLER | "Head Space Sampler Features (Select applicable features Only)" | Built-in Sample Shaker, PC Controlled Static Heated Tubing Transfer Line, Operates through Software with single keyboard controls (Software inclusive in the scope of supply);, Singel Vial incubation, Headspace with Syringe with XYZ movement Sampler |
| PYROLYZER SYSTEM | Compatible Pyrolyzer System for pyrolysis of solid and liquid samples- inclusive in the scope of supply | No |
| PYROLYZER SYSTEM | Temperature in the Pyrolyzer System | NA |
| SALIENT FEATURES OF GAS CHROMATOGRAPH (GC) | "Salient Featutes of GC (Select applicable features Only)" | Automatic leak testing/detection of Pneumatics |
| SOFTWARE | "Software Features (Select Applicable Features only)" | Compatible with Windows 7 or higher version operating system, Seamless integration & control of all the GC parameters, Interface with Multiple GCs through the comprehensive real-time monitoring and diagnostic software, Real time notification via advisories and indicators, Counters, Chromatographic attributes, Instrument diagnostics, Access to maintenance information, Access to maintenance, run and event logs, Retention Time - Locking module for analyzing target compounds in complex matrices |
| GENERIC | Power Supply | 230 V ± 10% |
| GENERIC | Display | Built-in Display |
| GENERIC | Display Size(In case of Built-in Display) | 7.0 - 10.0 Or higher |
| GENERIC | Connectivity interface | ETHERNET, RS 232 |
| GENERIC | Provision to split the sample from single injection to two different detectors and get results for two detectors simultaneously | Yes, No |
| ACESSORIES | Gas Supply inclusive in the scope of supply | Nitrogen, Hydrogen, Argon, Helium, Zero Air |
| ACESSORIES | Capacity of Gas cylinder | 47 liters water capacity |
| ACESSORIES | Number of Cyliners of each type supplied - inclusive in the scope of supply | 2 |
| ACESSORIES | Gas cylinders certified by PESO | No for Gas Cylinder Not Offered |
| ACESSORIES | Regulator - inclusive in the scope of supply | Dual Stage Regulator |
| ACESSORIES | "Other Accessories inclusive in the scope of supply (Select only those Accessories which are part of supply)" | Gas purification panel for all Gases with moisture trap, Hydrocarbon trap & Oxygen trap, Syringe for manual injection x 05 nos, Auto Sampler Vials x 1000 vials with screw cap, Graphite ferrules x 10 nos, Vessel ferrules x 10 nos, Septa for injector x 100 nos, Inlet liner for splitles x 4 nos, Inlet liner for split x 4 nos, O-rings x 10 nos. of each type |
| ENVIRONMENTAL | Minimum Operating Temperature | 0.0 - 5.0 Or lower |
| ENVIRONMENTAL | Maximum Operating Temperature | 35.0 - 65.0 Or higher |
| ENVIRONMENTAL | Operating Humidity (RH) (%) at 40 degree C | 74.0 - 95.0 Or higher |
| INSTALLATION & COMMISSIONING / WARRANTY / AVAILABILITY OF SPARES / TRAINING | Warranty on equipment | 2.0 |
| INSTALLATION & COMMISSIONING / WARRANTY / AVAILABILITY OF SPARES / TRAINING | Number of days onsite training provided | 7.0 |
| INSTALLATION & COMMISSIONING / WARRANTY / AVAILABILITY OF SPARES / TRAINING | Number of Year Availability of Spares from the date of installation is assured | 7.0 |
| CERTIFICATION / TEST REPORT | "Availability of UL/CE Certificate as per IEC 61010- 1 latest - Safety requirements for electrical equipment for measurement,control, and laboratory use General requirements" | Yes |
| CERTIFICATION / TEST REPORT | Conformance to EMC/EMI as per IEC 61326-1 latest - Electrical equipment for measurement, control and laboratory use | Yes |
| CERTIFICATION / TEST REPORT | Availability of Test Reports from Central Govt / NABL approved / ILAC accredited Lab to prove conformity to the specification | No |