Bid Publish Date
25-Aug-2026, 11:33 am
Bid End Date
14-Sep-2026, 7:00 pm
Location
Progress
Public procurement opportunity for Directorate Of Medical Education Government Of Madhya Pradesh CO2 Incubator (Q3) in Satna, MADHYA PRADESH. Quantity: 2 issued by. Submission Deadline: 14-09-2026 19: 00: 00. View full details and respond.
Quantity
2
Bid Type
Two Packet Bid
Bid Validity
120 (Days)
Bid Type
Service
Evaluation
Total value wise evaluation
Inspection Required
No
Tech Clarification Time
2 Days
EMD Required
No
MII Purchase Preference
No
MSE Purchase Preference
Yes
MSE Preference Band
L1+15%
MSE Exemption/Relaxation
Yes
Startup Exemption/Relaxation
Yes
Bid Splitting Applied
No
Min. Avg. Annual Turnover
2
OEM Avg. Turnover
8
Past Performance
50 %
Warranty Period
7 days
Experience Required
3 Year (s)
Arbitration Clause
No
Mediation Clause
No
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
No
Item Category
CO2 Incubator (Q3)
Advisory Bank
HDFC Bank
ePBG Percentage
5%
ePBG Duration (Months)
62
Payment Timelines
Payments shall be made to the Seller within 45 days of issue of consignee receipt-cum-acceptance certificate (CRAC) and on-line submission of bills (This is in supersession of 10 days time as provided in clause 12 of GeM GTC)
Max Delivery Days
15
Delivery Locations
1
Delivery Cities
Satna
Delivery Pincodes
485001
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Dr Siddharth Vijay Banode | 485001,Government Medical College, Satna Near Kendriya Vidyalaya No.2 Kripalpur | Satna | Madhya Pradesh | 485001 | 2 | 15 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
GEM_GENERAL_TERMS_AND_CONDITIONS
Discover companies most likely to bid on this tender
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover
Additional Doc 1 (Requested in ATC)
Compliance of BoQ specification and supporting document *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
| Category | Specification | Requirement |
|---|---|---|
| Performance Parameters | Temperature management range in degree Celsius | Range 3°C above ambient to 55°C |
| Performance Parameters | Temperature accuracy | ±0.1°C |
| Performance Parameters | Gas uniformity | ±0.1% at set 5.0% CO2 concentration at 37 ºC set temperature |
| Performance Parameters | Minimum Time taken for CO2 recovery rate after door opening | Under 6 min after door opening and closing event to attain 5% CO2. |
| Performance Parameters | Availability of Fan less design to reduce chance of contamination, reduce noise | Yes, Fan with in chamber HEPA for better recovery and ISO class 5 clean air Or higher |
| Reports | Warranty in years | 5 Or higher |