UGVCL Mehsana HPLC System Tender 2025 Q3 - High Performance Liquid Chromatography Procurement
N/a
Bid Publish Date
09-Jan-2026, 7:06 pm
Bid End Date
30-Jan-2026, 7:00 pm
EMD
₹9,50,000
Value
₹3,16,46,548
Location
Progress
Quantity
2
Bid Type
Two Packet Bid
The procurement opportunity targets a GC-MS system for the Forensic Science Laboratory in the Rohini, New Delhi area, under the Home Department Delhi. Estimated value is ₹31,646,548 with an EMD of ₹9,50,000. The scope includes comprehensive GC configurations with detectors, auto sampler, head space and pyrolysis compatibility, desktop workstation, UPS, and full installation. Unique aspects include multi-detector compatibility, dual-detector capability, and inclusion of required safety and EMI/EMC certifications. Bidders should align with government supply norms and local delivery to the Delhi address.
GC-MS configuration with inclusive GC detectors (FID, TCD, ECD, NPD/FTD, FPD, PID, PFPD)
Number of GC inlets and detectors supported simultaneously
Capillary column quantity, length, and type specification
Mass spectrometer with defined ionization modes and sensitivity; full-scan S/N ratio
Head space sampler and pyrolyzer compatibility included
Desktop workstation with required graphics memory; built-in display
Online UPS (kVA rating and backup time) and gas supply cylinders with regulators
EMD of ₹9.5 Lakh; ensure bid security as per terms
Delivery and installation at Rohini NABL Lab, Delhi
2% to 25% quantity variation and extended delivery timelines governed by option clause
Payment terms aligned with government procurement; GST invoicing to consignee with GSTIN; advance or milestone payments only as per contract
On-site delivery and installation at Forensic Science Laboratory, Rohini within contracted delivery period; option clause allows extension for increased quantity
Penalties may apply for delays; LD clauses to be specified in contract per delivery schedule
Minimum 3 years of govt GC-MS supply experience
Not under liquidation or bankruptcy
OEM authorization if the bidder is an authorized distributor
N/a
N/a
ETAWAH, UTTAR PRADESH
Directorate Of Employment & Training
NAVSARI, GUJARAT
Directorate Of Training And Employment, Lucknow
Hamirpur, UTTAR PRADESH
Kerala Agricultural University
THIRUVANANTHAPURAM, KERALA
Tender Results
Loading results...
| Category | Specification | Requirement |
|---|---|---|
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "GC Detectors inclusive in the scope of supply | Flame Ionization Detector (FID), Thermal Conductivity Detector (TCD or Hot Wire Detector), Electron Capture Detector (ECD), Nitrogen–Phosphorus Detector (NPD) / Flame Thermionic Detector (FTD), Flame Photometric Detector (FPD), Photo Ionization Detector (PID), Pulsed Flame Photometric Detector (PFPD) |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of FID in the System - inclusive in the scope of Supply | 0, 1, 2 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of TCD in the System - inclusive in the scope of Supply | 0, 1, 2 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of ECD in the System - inclusive in the scope of Supply | 0, 1, 2 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of NPD/FTD in the System - inclusive in the scope of Supply | 0, 1, 2 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of FPD in the System - inclusive in the scope of Supply | 0, 1, 2 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of PID in the System - inclusive in the scope of Supply | 0, 1, 2 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of PFPD in the System - inclusive in the scope of Supply | 0, 1, 2 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | Number of Inlets GC can Simultaneously Support | 1, 2, 3 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | Number of Detectors (Ports) GC can Simultaneously Support | 1, 2, 3, 4 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | Number of Capillary Columns provided in GC - inclusive in the scope of Supply | 0, 1, 2 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | Type of Capillary Column | Polar, Non-Polar, Semi-Polar |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "GC Capillary Column Length (Hint:-Select '0' if Not Applicable)" | 0, 30, 60, 100 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | GC Packed Column Length | 0, 3, 6, 9, 12 |
| PERFORMANCE | GC Detector Minimum Detection Level (MDL) | < 1.4 pg C/s for FID, < 1.8 pg C/s for FID, < 3.0 pg C/s for FID < 400 pg/ml for TCD, < = 1 microgram/ml for TCD, < = 10 microgram/ml for TCD, < 4.5 fg /sec for ECD, < 5.5 fg /sec for ECD, < 0.1pg N /sec for NPD/FTD, < 0.03 pg P /sec for NPD/FTD, < 45 fg P /sec for FPD, < 100 fg P /sec for FPD, <2.5 pg S /sec for FPD, <5 pg S /sec for FPD |
| COLUMN OVEN (GAS CHROMATOGRAPH - GC) | Maximum Operating Temperature (GC Oven) | upto 400 deg. C, upto 450 deg. C (2) Or higher |
| COLUMN OVEN (GAS CHROMATOGRAPH - GC) | Number of Oven Ramps / Plateaus for Temperature programming | 6/7 or more, 9/10 or more, 20/21 or more |
| PNEUMATIC CONTROL | Number of Injection Ports | 1, 2 |
| PNEUMATIC CONTROL | Maximum Injection Port heating Temperature | upto 400 deg. C (1), upto 450 deg. C (2) Or higher |
| PNEUMATIC CONTROL | Pressure Range | upto100 psi (1), upto 120 psi (2) Or higher |
| PNEUMATIC CONTROL | Pressure set point Resolution (±)(psi) | 0.001, 0.01, 0.1 Or lower |
| PNEUMATIC CONTROL | Type of Sample Injection Port | Split, Splitless, split / Splitless, Purged packed Inlet |
| PNEUMATIC CONTROL | Split Ratio | 6000:1, 7000:1, 7500:1 |
| AUTO INJECTOR / AUTO SAMPLER | Compatible Liquid Auto Sampler - inclusive in the scope of supply | Yes, No |
| AUTO INJECTOR / AUTO SAMPLER | Capacity of the Liquid Auto Sampler (Number of Vials) | 0, 100, 150 Or higher |
| AUTO INJECTOR / AUTO SAMPLER | Number of Injection Ports in the Liquid Auto Sampler | 0, 1, 2 |
| AUTO INJECTOR / AUTO SAMPLER | Number of Channels in the Liquid Auto Sampler | 0, 2 |
| AUTO INJECTOR / AUTO SAMPLER | Liquid Auto Sampler Features | Fully automated Liquid Auto Sampler, Injects with variable speed & varying syringe sizes, Operates through Software (Software inclusive in the scope of supply), NA |
| HEAD SPACE SAMPLER | Compatible Head Space Sampler- inclusive in the scope of supply | Yes, No |
| HEAD SPACE SAMPLER | Capacity of the Head Space Sampler Carousel (number of Vials) | 12 or better, NA |
| HEAD SPACE SAMPLER | Head Space Sampler Temperature Range | Upto 200 deg. C or better, NA |
| PYROLYZER SYSTEM | Compatible Pyrolyzer System for pyrolysis of solid and liquid samples- inclusive in the scope of supply | Yes, No |
| MASS SPECTROPHOTOMETER (MS) SYSTEM | Ionization modes | Electron Ionization (EI), Chemical Ionization (CI), EI/PCI/NCI |
| MASS SPECTROPHOTOMETER (MS) SYSTEM | Detector Type | Electron Multiplier, Photo Multiplier, Triple-Axis, HED-EM |
| MASS SPECTROPHOTOMETER (MS) SYSTEM | Operating modes | Full Scan Mode, Selective Ion Monitoring (SIM) Mode, Both Full Scan & SIM Modes |
| PERFORMANCE OF THE MASS SPECTROPHOTOMETER (MS) SYSTEM | Sensitivity (S/N Ratio) in full scan | 1500:1 with 1 pg/microliter OFN, 800:1 (EI) or better OFN, 700:1 (EI) or better OFN, 250:1 (EI) or better OFN, 300:1 (PCI) or better BZP, 2000:1 (NCI) or better OFN |
| SALIENT FEATURES OF GC-MS | "Salient Featutes of GC-MS (Select applicable features Only)" | Automatic leak testing/detection of Pneumatics, Direct setting of split flow rates and ratios, Flow sensor for control & storage of split ratio, Split and Septum Purge gases, Able to direct inject onto packed and wide-bore capillary columns, Oven power turns-off automatically when the lid/door is opened, Over Heat Protection, Display Unit with built-in Key Board, Controlled by External PC System, Programmable ion source temperature, System includes variety of auto tune algorithms to tune the instrument for maximum sensitivity or for specific target compounds, EI source is inert to active compounds and programmable, EI Source have a dual filament design with automatic selection of other filament if one fails during analysis;, NA |
| LIBERARIES | Libraries for GC-MS | NIST Library for MS - Licensed version - 2017 or latest, Wiley Library, NA |
| SOFTWARE | "Software Features (Select Applicable Features only)" | Compatible with Windows 7 or higher version operating system, Seamless integration & control of all the GC parameters, Interface with Multiple GCs through the comprehensive real-time monitoring and diagnostic software, Real time notification via advisories and indicators, Counters, Chromatographic attributes, Instrument diagnostics, Access to maintenance information, Access to maintenance, run and event logs, Retention Time - Locking module for analyzing target compounds in complex matrices |
| DESKTOP WORKSTATION | Desktop workstation Suitable to GC-MS - inclusive in the scope of supply | Yes, No |
| DESKTOP WORKSTATION | Graphics Card Memory Size | 0, 1, 2, 4 |
| ONLINE UPS | UPS Provided in the Scope of Supply | Yes, No |
| ONLINE UPS | Rating of online UPS(KVA) | 5.0 KVA, 7.5 KVA, 10 KVA, NA |
| ONLINE UPS | Backup time of online UPS | 0, 60, 90, 120 |
| GENERIC | Built-in Display | Yes, No |
| GENERIC | Connectivity interface | USB, WIFI, ETHERNET, Bluetooth, RS 232, RJ-45 |
| GENERIC | Provision to split the sample from single injection to two different detectors and get results for two detectors simultaneously | Yes, No |
| GENERIC | Back Flash - inclusive in the scope of supply | Post Column Back Flash, Mid Column Back Flash, NA |
| ACESSORIES | Gas Supply inclusive in the scope of supply | Nitrogen, Hydrogen, Argon, Helium, Zero Air, Methane, Isobutane, Not offered |
| ACESSORIES | Number of Cyliners of each type supplied - inclusive in the scope of supply | 0, 1, 2 |
| ACESSORIES | Gas cylinders certified by PESO | Yes, NA |
| ACESSORIES | Regulator - inclusive in the scope of supply | Dual Stage Regulator, NA |
| ACESSORIES | "Other Accessories inclusive in the scope of supply (Select only those Accessories which are part of supply)" | Gas purification panel for all Gases with moisture trap, Hydrocarbon trap & Oxygen trap, Syringe for manual injection x 05 nos, Auto Sampler Vials x 1000 vials with screw cap, Graphite ferrules x 10 nos, Vessel ferrules x 10 nos, Septa for injector x 100 nos, Inlet liner for splitles x 4 nos, Inlet liner for split x 4 nos, O-rings x 10 nos. of each type, Column Cutter x 2 nos, Gas tube Cutter x 2 nos, Tool Kit x 1 no., User / Technical / Maintenance Manual, Standard Samples - NIST / IPL Traceable, Direct Injection Port, Splitter, Multishot Pyrolyser, Singleshot Pyrolyser, Syringe for Auto Sampler injection x 10 nos, EI filaments x 5 nos, NA |
| ENVIRONMENTAL | Operating Humidity (RH) (%) at 40 degree C | Up to 95 |
| INSTALLATION & COMMISSIONING / WARRANTY / AVAILABILITY OF SPARES / TRAINING | Warranty on equipment | 1, 2, 3, 5 |
| INSTALLATION & COMMISSIONING / WARRANTY / AVAILABILITY OF SPARES / TRAINING | Number of days onsite training provided | 3, 5, 7, 10 |
| CERTIFICATION / TEST REPORT | "Availability of UL/CE Certificate as per EN 61010- 1:2010 Safety requirements for electrical equipment for measurement,control, and laboratory use General requirements" | Yes, No |
| CERTIFICATION / TEST REPORT | Conformance to EMC/EMI as per EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use | Yes, No |
| CERTIFICATION / TEST REPORT | Availability of Test Reports from Central Govt / NABL approved / ILAC accredited Lab to prove conformity to the specification | Yes, No |
Discover companies most likely to bid on this tender
GST registration certificate
PAN card
Experience certificates of supplying GC-MS systems to Govt bodies (3 years)
Financial statements demonstrating financial capacity
EMD payment receipt or bank guarantee
OEM authorization or manufacturer BIDDER certification
Technical bid documents and conformity certificates
UL/CE Certifications and EMC/EMI test reports
NABL/ILAC accredited test reports if available
Bidders must meet eligibility criteria including 3 years of govt GC-MS supply experience, provide OEM authorization, and submit EMD of ₹9,50,000. Include GST registration, PAN, financial statements, and NABL/ILAC test reports. Installation will be at Rohini, New Delhi with on-site training.
Submit GST certificate, PAN, experience certificates for govt GC-MS supplies, financial statements, EMD payment proof, OEM authorization, technical conformity certificates, and safety/EMC/NABL test reports. Invoices must be in consignee GSTIN name.
Equipment must comply with EN 61010-1 safety and EN 61326-1 EMC standards; NABL/ILAC accredited test reports should confirm conformity; UL/CE certificates should be available per the tender requirements.
The Earnest Money Deposit is ₹9,50,000, to be submitted as per bid terms; ensure the payment method aligns with the tender rules and is refundable if the bid is unsuccessful.
Delivery and installation are to be completed at the Forensic Science Laboratory, Rohini, New Delhi, within the contracted timeline; option clause allows quantity-based extension up to 25% with extended delivery time calculations.
The scope includes multiple detectors (FID, TCD, ECD, NPD/FTD, FPD, PID, PFPD), multiple GC inlets/ports, capillary columns, head space sampler, auto sampler, pyrolyzer compatibility, and a desktop workstation with UPS and data libraries.
Yes, OEM authorization or manufacturer certificate is required when bidding as an authorized distributor; include complete details and contact information of the OEM in the bid submission.
Warranty terms are specified in the contract with on-site commissioning; ensure spare parts availability and post-warranty AMC options as part of the bid proposal to meet long-term support expectations.
Council of Scientific and Industrial Research
Directorate Of Purchase And Stores
📍 MUMBAI, MAHARASHTRA
Indian Space Research Organization
📍 THIRUVANANTHAPURAM, KERALA
National Institute Of Technology (nit)
📍 RAIPUR, CHHATTISGARH
Indian Space Research Organization
📍 NELLORE, ANDHRA PRADESH
Sign up now to access all documents
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS