National Rural Health Mission announces a tender for Immunoassay Analyzer (V3) (Q2) in RAYAGADA, ODISHA. Quantity: 11. Submission Deadline: 02-09-2026 18: 00: 00. Last date to apply is approaching fast!
Quantity
11
Bid Type
Two Packet Bid
Bid Validity
180 (Days)
Bid Type
Service
Evaluation
Total value wise evaluation
Inspection Required
No
RA Qualification Rule
H1-Highest Priced Bid Elimination
Tech Clarification Time
2 Days
EMD Required
Yes
MII Purchase Preference
No
MSE Purchase Preference
No
MSE Exemption/Relaxation
No
Startup Exemption/Relaxation
No
Bid Splitting Applied
No
Min. Avg. Annual Turnover
25
OEM Avg. Turnover
100
Past Performance
50 %
Warranty Period
1 years
Experience Required
3 Year (s)
Arbitration Clause
No
Mediation Clause
No
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
Yes
Item Category
Immunoassay Analyzer (V3) (Q2)
Advisory Bank
State Bank of India
ePBG Percentage
2%
ePBG Duration (Months)
14
Max Delivery Days
15
Delivery Locations
1
Delivery Cities
Rayagada
Delivery Pincodes
765001
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| RAJABALA PRIYAMBADA | 765001,DHH HOSPITAL MAIN ROAD RAYAGADA | Rayagada | Odisha | 765001 | 11 | 15 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
OTHER
GEM_GENERAL_TERMS_AND_CONDITIONS
Discover companies most likely to bid on this tender
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover
Additional Doc 1 (Requested in ATC)
Additional Doc 2 (Requested in ATC)
Additional Doc 3 (Requested in ATC)
Additional Doc 4 (Requested in ATC)
Compliance of BoQ specification and supporting document *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
| Category | Specification | Requirement |
|---|---|---|
| PRODUCT INFORMATION | Type of Configuration | Benchtop |
| PRODUCT INFORMATION | Type of system configuration | Standalone |
| PRODUCT INFORMATION | Type of system | Closed |
| PRODUCT INFORMATION | Processing modes | Batch Mode |
| PRODUCT INFORMATION | Type of Automation | Semi automated |
| PRODUCT INFORMATION | Throughput of the system per hour | ≥ 50 |
| PRODUCT INFORMATION | Sample Loading Capacity at a time | ≥ 10 |
| PRODUCT INFORMATION | Type of samples | Serum, Plasma, Whole blood, Urine |
| PRODUCT INFORMATION | Sample volume per test in µL | ≥ 1 µL |
| PRODUCT INFORMATION | Type of method | Chemiluminescence (CLIA), Enzyme immunoassays (EIA) |
| Test Menu | Cardiac Markers | Yes |
| Test Menu | Tests which can be performed by Cardiac Markers | CK-MB, Troponin I / Troponin-T, NT-proBNP |
| Test Menu | Infectious Bio Markers | No |
| Test Menu | Tests which can be performed by Bio Markers | NA |
| Test Menu | Tumor Markers | No |
| Test Menu | Tests which can be performed by Tumor Markers | NA |
| Test Menu | Anemia profile | Yes |
| Test Menu | Tests which can be performed by Anemia profile | Ferritin |
| Test Menu | Bone Metabolism | Yes |
| Test Menu | Tests which can be performed by Bone metabolism | 25-OH Vitamin D |
| Test Menu | Thyroid profile | Yes |
| Test Menu | Tests which can be performed by thyroid profile | TSH, T4, T3, FT4, FT3 |
| Test Menu | Infertility | Yes |
| Test Menu | Tests which can be performed for infertility | FSH, LH, HCG/ß-HCG, Estradiol, Testosterone, AMH |
| Test Menu | Infectious Test | No |
| Test Menu | Tests which can be performed for Infectious Test | NA |
| Test Menu | Dual, Triple or Quad prenatal screening tests | No |
| Test Menu | Tests which can be performed for prenatal screening | NA |
| Test Menu | Liver Fibrosis Tests | No |
| Test Menu | Tests which can be performed for Liver Fibrosis | NA |
| Test Menu | Alzheimer Screening | No |
| Test Menu | Tests which can be performed for Alzheimer Screening | NA |
| Test Menu | Neuro Markers | No |
| Test Menu | Tests which can be performed for Neuro Markers | NA |
| Test Menu | Immuno Suppressive drugs | No |
| Test Menu | Tests which can be performed for Immuno Suppressive drugs | NA |
| SYSTEM FEATURES | Closed-tube sampling | Yes, No Or higher |
| SYSTEM FEATURES | Direct tube sampling | Yes, No Or higher |
| SYSTEM FEATURES | Liquid-level sensing | Yes |
| SYSTEM FEATURES | Clot, Bubble, Viscosity detection | Yes |
| SYSTEM FEATURES | Type of Calibration | Automatic and Manual |
| DATA MANAGEMENT | Type of display | LED/LCD/TFT |
| DATA MANAGEMENT | Size of Display in inches | ≥ 10 Inches |
| DATA MANAGEMENT | Touch Screen display | Yes |
| DATA MANAGEMENT | Handheld barcode reader | No |
| DATA MANAGEMENT | HIS/LIS Interface | RS 232 serial, Ethernet |
| DATA MANAGEMENT | Data Storage of sample | 1000 or more |
| ACCESSORIES | Startup Reagent Kit to be provided | Yes |
| POWER REQUIREMENTS | Type of power supply in VAC,Hz | 200-240 VAC, 50-60 Hz |
| CERTIFICATION AND REPORTS | Compliance to Drugs and Cosmetic Act 1940 and Medical Device Rules (MDR) 2017 as amended till date | Yes |
| CERTIFICATION AND REPORTS | Availability of valid Medical Device license for the product issued from the competent authority | Yes |