Ultrasound Machine (V2)
N/a
LUCKNOW, UTTAR PRADESH
Bid Publish Date
23-Apr-2026, 7:21 pm
Bid End Date
04-May-2026, 7:00 pm
EMD
₹31,480
Value
₹31,48,000
Location
Progress
Quantity
1
Bid Type
Two Packet Bid
The procurement opportunity is issued by the Chhattisgarh Medical Services Corporation Limited for an advanced Ultrasound Machine (V2) in Bilaspur, Chhattisgarh with an estimated value of ₹31,48,000 and an EMD of ₹31,480. The scope includes a wide array of probes and features (Convex, Microconvex, Linear, TVS, Cardiac, Hockey Stick, Volume variants, Endorectal), elastography options (E-Breast, E-Thyroid, Liver), and Cine/3D-4D capabilities. The device must support harmonics, CDI, PW/Power Doppler, triplex imaging, CEUS, tissue colorization, and DICOM interfaces, with PCI-PC connectivity and optional offline storage. The tender emphasizes real-time 3D imaging, high frame rates, and a robust diagnostic workflow, suitable for hospital use in government procurement. bid evaluation will consider compliance with listed scan modes, imaging processing, and data interoperability requirements.
Ultrasound system with complete probe portfolio: convex, microconvex, linear, TVS, cardiac, hockey stick, volume variants, endorectal
Elastography support: E-Breast, E-Thyroid, Liver elastography
Imaging features: CEUS, harmonic imaging, 3D/4D, triplex, tissue Doppler, color Doppler, PDI, 2D/3D cine loops
Display/processing: DICOM 3 compliant, automated B-mode and CDI image optimization, touch screen, HDMI/PCI interfaces
Interoperability: DICOM to PACS and RIS interfaces; USB/DVD writer; PC with dedicated software; network connectivity
EMD of ₹31,480 and total estimated value ₹31,48,000; delivery may vary by ±25%
Option for comprehensive warranty in bidding; supersedes catalogue warranty if chosen
OEM authorizations and project-specific certificates required for probes and components
Final payment terms to be defined in contract; ensure alignment with government procurement norms and EMD compliance
Delivery quantity may vary up to 25%; extended delivery time based on quantity increase with minimum 30 days
Penalty clauses to be specified in contract; ensure performance bonds if required
Evidence of prior supply of ultrasound systems to government or large health facilities
Proven capability to interface with PACS/RIS and to deliver comprehensive probe sets
Valid GST and financial stability demonstrating execution capacity
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
No
Item Category
Ultrasound Machine (V2) (Q2)
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS
N/a
LUCKNOW, UTTAR PRADESH
N/a
LUCKNOW, UTTAR PRADESH
N/a
RAJKOT, GUJARAT
N/a
NORTH EAST DELHI, DELHI
N/a
CENTRAL DELHI, DELHI
Tender Results
Loading results...
| Category | Specification | Requirement |
|---|---|---|
| Types of Probes & Features | Convex Probe | 2.5 to 5 MHz (±2), Not Provided |
| Types of Probes & Features | Microconvex probe | 4 to 8 MHz (±2), Not Provided |
| Types of Probes & Features | Linear array probe | 5 to 10 MHz (±2), 3 to 12 MHz (±2), 6 to 15 MHz (±2), 6 to 18 MHz (±2), Not provided |
| Types of Probes & Features | TVS Probe | 4 to 9 MHz (±2), Not Provided |
| Types of Probes & Features | Cardiac Probe | 1 to 5 MHz (±2), 5 to 9 MHz (±2), Not Provided |
| Types of Probes & Features | Hockey Stick Probe | 6 to 17 MHz (±2), 6 to 13 MHz (±2), 8 to 20 MHz (±2), Not Provided |
| Types of Probes & Features | Volume Convex Probe | 2.5 to 5 MHz (±2), Not Provided |
| Types of Probes & Features | Volume Linear Probe | 3 to 12 MHz (±2), 6 to 15 MHz (±2), 6 to 18 MHz (±2), 5 to 10 MHz (±2), Not Provided |
| Types of Probes & Features | Volume TVS Probe | 4 to 9 MHz (±2), Not Provided |
| Types of Probes & Features | Endorectal Probe | 5 to 9 MHz (±2), Not Provided |
| Types of Probes & Features | E-Breast Elastography technique | Yes, No |
| Types of Probes & Features | E-Thyroid Elastography technique | Yes, No |
| Types of Probes & Features | Liver Elastography technique | Yes, No |
| Types of Probes & Features | Gel Warmer | Yes, No |
| Types of Probes & Features | Dedicated 3-D/4-D | Yes, No |
| Scan Modes | Contrast harmonic imaging with angio mode | Yes, No |
| Scan Modes | Provision of Harmonic imaging in Linear Probes | Yes, No |
| Scan Modes | Color Doppler imaging (CDI) | Yes, No |
| Scan Modes | Color Doppler 3-D/4-D option | Yes, No |
| Scan Modes | Power Doppler imaging (PDI) | Yes, No |
| Scan Modes | Power Doppler 3-D/4-D option | Yes, No |
| Scan Modes | Continuous wave | Yes, No |
| Scan Modes | Triplex mode | Yes, No |
| Scan Modes | Tissue Doppler imaging | Yes, No |
| Scan Modes | Contrast-enhanced ultrasound (CEUS) option | Yes, No |
| Scan Modes | Tissue synchronization, tissue velocity | Yes, No |
| Scan Modes | Provision for higher sensitivity for low frequency doppler in all probes | Yes, No |
| Scan Modes | Number of channels in color doppler system | 50000 or more (if channels are being designated as digital channels), 300000 or more (if channels are being designated as digital channels), 1000000 or more (if channels are being designated as digital channels), 2500000 or more (if channels are being designated as digital channels), 4000000 or more (if channels are being designated as digital channels), 128 or more (if channels are being designated as physical channels) |
| Scan Modes | Live 3 D imaging in color doppler system | Yes, No |
| Scan Modes | Colour Doppler system with all application packages ,Quad loop for serial studies with high frame rate revies, harmonic imaging capability in all modes and integrated stress echo package digital storage and retrieval | Yes, No |
| Scan Modes | Provision of Tissue Colorization (B-Colour) for improved contrast resolution | Yes, No |
| Scan Modes | Cine Loop Features | Yes, No |
| Scan Modes | Frame grabber facility for post analysis | Yes, No |
| Scan Modes | Number of transducer/probe ports | 2, 3, 4 or More |
| Scan Modes | System should be capable of generating real time live 3- D images | Yes, No |
| Scan Modes | Contrast agent imaging | Yes, No |
| Image Display and Processing | DICOM 3 Compliant | Yes, No |
| Image Display and Processing | Automated B-mode (2-D) image | Yes, No |
| Image Display and Processing | Automated CDI(Color Doppler Imaging) image optimization | Yes, No |
| Image Display and Processing | Automated PW Doppler image optimization | Yes, No |
| Image Display and Processing | Touch Screen | Yes, No |
| Image Display and Processing | Monitor display size (inch) | 15 or more, 17 or more, 19 or more, 21 or more Or higher |
| Image Display and Processing | Split screen | Twin View, Twin View and Live Quad Screen, No |
| Image Display and Processing | Single/dual monitors | Single, Single with optional second console, Dual (image display and touch panel) |
| Image Display and Processing | Max number frames | >=1000, >=500 |
| Image Display and Processing | Digital storage hard drive, GB | >=500, >=200 or more |
| Image Display and Processing | DVD/CD Writer/ USB | Yes, No |
| Image Display and Processing | Type of processor (in PC) | i7 or latest, NA |
| Image Display and Processing | RAM capacity of PC in GB (hint: In case PC not provided put zero) | ≥ 4, NA |
| Image Display and Processing | Processor speed in GHz (PC) | ≥ 3, NA |
| Image Display and Processing | Frame Grabber incorporated(PC) | Yes, No |
| Image Display and Processing | Connectivity between USG and PC | Yes, No |
| Image Display and Processing | PC with all software inclusive interfaced with USG machine | Yes, No |
| Image Display and Processing | Laser Color printer provided | Yes, No |
| Image Display and Processing | Inkjet/Thermal printer provided | Yes, No |
| Image Display and Processing | CD/DVD/USB Produced should be playable on any system | Yes, No |
| Image Display and Processing | Glossy Colour Print Paper supplied | 500 sheets, 1000 sheets, NA |
| Image Display and Processing | Dicom Interfacability Dicom to PACS | Yes, No |
| Image Display and Processing | Interfacability Dicom to RIS | Yes, No |
| Power Requirements | Resettable over current breaker shall be fitted for protection | Yes, No |
| Power Requirements | Power backup | UPS, Inbuilt battery |
| Power Requirements | Back up time in minutes (minimum) | 30, 60, 90, 120 |
| Power Requirements | Suitable Servo Controlled Stabilizer / CVT should be provided | Yes, No |
| Warranty and Maintenance | Warranty in Years (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) | 3, 5 Or higher |
Discover companies most likely to bid on this tender
GST registration certificate
Permanent Account Number (PAN) card
Experience certificates for similar ultrasound system supply
Financial statements (last 3 years) or equivalent financials
EMD payment receipt/document
Technical bid documents and compliance certificates
OEM authorization/brand authorization for ultrasound probes and components
Any additional documents requested in Buyer Added Terms
Key insights about CHHATTISGARH tender market
Bidders should meet eligibility criteria, submit GST and PAN, provide financials and experience certificates, and attach OEM authorizations for all probes. Ensure EMD of ₹31,480 is paid and confirm DICOM 3 compliance, elastography options, and 3D/4D imaging capabilities.
Submit GST certificate, PAN, 3-year financial statements, experience certificates for similar supplies, OEM authorization for probes, EMD receipt, technical bid, and any specific CGMSCL terms; ensure DICOM interoperability and IS/CE compliance as applicable.
System must support Convex, Microconvex, Linear, TVS, Cardiac, Endorectal probes; elastography for breast, thyroid, and liver; CEUS, harmonic imaging, 3D/4D, and color/Power Doppler with triplex mode; DICOM 3 compliant with PACS/RIS interfaces.
EMD is ₹31,480; estimated contract value is ₹31,48,000. The quantity may vary ±25% under option clause, with delivery timelines adjusted accordingly and a minimum 30 days extended period.
A comprehensive warranty option is available through bidding and, if selected, supersedes the standard catalogue warranty. Ensure terms in the contract specify coverage duration and service response times.
Purchaser may increase/decrease quantity up to 25% at contract placement and during currency; delivery time scales with quantity change and minimum 30 days, beginning from the last date of the original/extended delivery order.
System must be DICOM 3 compliant, with PACS/RIS interoperability, IS/CE compliant as specified, and compatible with hospital workflows, offline storage, and PC interfacing as per CGMSCL terms.
All India Institute Of Medical Sciences (aiims)
📍 RAIPUR, CHHATTISGARH
Western Railway
📍 MUMBAI, MAHARASHTRA
Institute Of Kidney Diseases And Research Centre
📍 AHMEDABAD, GUJARAT
N/a
📍 CHAMBA, HIMACHAL PRADESH
All India Institute Of Medical Sciences (aiims)
📍 NADIA, WEST BENGAL