Bid Publish Date
27-Jul-2026, 3:16 pm
Bid End Date
07-Sep-2026, 9:00 am
Location
Progress
Jawaharlal Institute Of Postgraduate Medical Education And Research (jipmer) invites bids for Operating Table (V2) (Q2) in PONDICHERRY, PUDUCHERRY. Quantity: 2. Submission Deadline: 07-09-2026 09: 00: 00. Submit your proposal before the deadline.
Quantity
2
Bid Type
Two Packet Bid
Bid Validity
65 (Days)
Bid Type
Service
Evaluation
Total value wise evaluation
Inspection Required
No
Tech Clarification Time
2 Days
EMD Required
Yes
MII Purchase Preference
Yes
MII Preference Band
L1+20%
MSE Purchase Preference
Yes
MSE Preference Band
L1+15%
MSE Exemption/Relaxation
No
Startup Exemption/Relaxation
No
Bid Splitting Applied
No
Min. Avg. Annual Turnover
4
OEM Avg. Turnover
20
Past Performance
20 %
Warranty Period
15 years
Experience Required
3 Year (s)
Arbitration Clause
No
Mediation Clause
No
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
No
Item Category
Operating Table (V2) (Q2)
Advisory Bank
State Bank of India
ePBG Percentage
3%
ePBG Duration (Months)
38
Max Delivery Days
45
Delivery Locations
1
Delivery Cities
Pondicherry
Delivery Pincodes
605006
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Sreevathsa K S Prasad | 605006,Jawaharlal Institute of Postgraduate Medical Education and Research, Dhanvantari Nagar, Puducherry | Pondicherry | Puducherry | 605006 | 2 | 45 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
OTHER
OTHER
GEM_GENERAL_TERMS_AND_CONDITIONS
Extended Deadline
07-Sep-2026, 3:30 am
Opening Date
07-Sep-2026, 4:00 am
Discover companies most likely to bid on this tender
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover
Additional Doc 1 (Requested in ATC)
Additional Doc 2 (Requested in ATC)
Additional Doc 3 (Requested in ATC)
Additional Doc 4 (Requested in ATC)
Compliance of BoQ specification and supporting document *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
| Category | Specification | Requirement |
|---|---|---|
| General Features | Product Description | Operating Table |
| Table features | Type of OT table Actuator | Electro-hydraulic |
| Table features | Type of mechanism for functioning or controlling the table movements | Electro-hydraulic movement through remote and backup panel |
| Table features | Manual backup control Functionality | Yes |
| Table features | Remote controlled positions | Trendelenburg and reverse Trendelenburg, lateral tilt, longitudinal shift, height adjustment, antiflex and flex position, back plate up / down, 0 position |
| Table features | Number of section provided in Table top | 5, 6 |
| Table features | Headrests should be detachable and removable | Yes |
| Table features | Foot section can also be moved or removed for certain procedures | Yes |
| Table features | Head and foot sections are interchangeable so that the patient can be reversed for extended radiographic imaging or special procedures | Yes |
| Table features | C-Arm Compatible | Yes |
| Table features | Range of Table top length | 1900 mm - 2100 mm |
| Table features | Range of Table top width | 500 mm - 580 mm |
| Table features | Range of Table top Height (minimum) | 500 - 600 mm, 601 - 700 mm |
| Table features | Range of Table top Height (maximum) | 1000 mm - 1100 mm |
| Table features | Radiolucent table top | Yes |
| Table features | Mattress shall be in parts as per table top | Yes |
| Table features | Provision of X ray cassette channel | Yes |
| Table features | Table top shall be X-ray compatible and stain Resistant | Yes |
| Table Positions | Trendelenburg Angle, Range (maximum) | 25⁰ to 30⁰ |
| Table Positions | Reverse Trendelenburg Angle, Range (maximum) | 25⁰ to 30⁰ |
| Table Positions | Lateral Tilt Angle (Left), Range | 15⁰ to 25⁰ |
| Table Positions | Lateral Tilt Angle (Right), Range | 15⁰ to 25⁰ |
| Table Positions | Head Section UP/Down Angle | Yes |
| Table Positions | Back Section UP/Down Angle, Range | Upto +60⁰ to +90⁰, Upto - 30⁰ to -90⁰ |
| Table Positions | Leg UP/DOWN through remote control | Yes |
| Table Positions | Leg Section UP/Down, Range | Electrical movement 60 deg UP and upto 90 deg DOWN |
| Table Positions | Flex | Upto 230⁰ |
| Table Positions | Reflex | Upto 100⁰ |
| Table Positions | Table top longitudinal sliding, Range | 250 -300 mm, 301- 350 mm, 351- 400 mm, 401- 450 mm, No sliding |
| Table Positions | Patient Weight Capacity | Static (Upto 300 - 400 kg) and Dynamic (250 to 300 kg) |
| Table Positions | Detachable Head,Leg & Pelvic Section | Yes |
| Table Positions | Split leg section , detachable | 1 pair, NA |
| Table Positions | Quick Switch for Zero position | Yes |
| Table Positions | Battery backup provided | Yes |
| Table Positions | Minimum Battery backup of operating table in hrs | 3 hrs |
| Table Positions | Foot Pump | Yes |
| Table Positions | Both arm board should have up and down and rotation function | Yes |
| Table Positions | Operating table should have foot brake for ease of use | Yes |
| Table Positions | Reverse orientation | Yes |
| Accessories | Number of Anesthetic Screen with side clamp & lock | 1 |
| Accessories | Number of Arm rest with side clamp & lock | 1 pair |
| Accessories | Number of Shoulder support with side clamp & lock | 1 pair |
| Accessories | Number of Leg support with side clamp & lock | 1 pair |
| Accessories | Number of Bariatric table Extensions provided -Only for Bariatric Procedures | NA |
| Accessories | Radial setting clamp | NA |
| Accessories | Restraint body strap | 1 pair |
| Accessories | Number of Bolster Pillow | 1 |
| Accessories | IV pole and Bracket | No |
| Power requirements | Power input | Single phase 220-240 Volt, 50 Hz ( AC Supply ) |
| CERTIFICATIONS | Compliance to Drugs and Cosmetic Act 1940 and Medical Device Rules (MDR) 2017 as amended till date | Yes |
| CERTIFICATIONS | Availability of valid Medical Device license for the product issued from the competent authority | Yes |
| CERTIFICATIONS | License issuing authority as per MDR | Central Licensing Authority |
| WARRANTY | Warranty | 1 |