TenderDekho Logo
Active GEM

Invitation to Bid - ICU Ventilators (V2) in EAST SINGHBHUM, JHARKHAND

Bid Publish Date

22-Aug-2026, 6:07 pm

Bid End Date

01-Sep-2026, 7:00 pm

EMD

₹75,000

Location

EAST SINGHBHUM , JHARKHAND

Progress

Issue22-Aug-2026, 6:07 pm
AwardPending

A tender has been published for Government Pharmacy Institute Bariatu Ranchi ICU Ventilators (V2) (Q2) in EAST SINGHBHUM, JHARKHAND. Quantity: 1 by. Submission Deadline: 01-09-2026 19: 00: 00. Check eligibility and apply.

Tender Data

Bid Details

Quantity

1

Bid Type

Two Packet Bid

Bid Validity

15 (Days)

Bid Type

Service

Evaluation

Total value wise evaluation

Inspection Required

No

Tech Clarification Time

2 Days

Preference & Exemptions

EMD Required

Yes

MII Purchase Preference

No

MSE Purchase Preference

No

MSE Exemption/Relaxation

No

Startup Exemption/Relaxation

No

Bid Splitting Applied

No

Requirements

OEM Avg. Turnover

60

Past Performance

50 %

Warranty Period

5 years

Experience Required

3 Year (s)

Arbitration Clause

No

Mediation Clause

No

Additional Tender Data

Commercial Details

Tender Category

Goods

Bid To RA

No

Bid To RA Enabled

No

Item Category

ICU Ventilators (V2) (Q2)

Advisory Bank

Bank Of Baroda

Payment Timelines

Payments shall be made to the Seller within 30 days of issue of consignee receipt-cum-acceptance certificate (CRAC) and on-line submission of bills (This is in supersession of 10 days time as provided in clause 12 of GeM GTC)

Delivery Details

Max Delivery Days

30

Delivery Locations

1

Delivery Cities

Saraikela Kharsawan

Delivery Pincodes

831002

Delivery Locations

ConsigneeAddressCityStatePincodeQuantityDelivery DaysAdditional Requirement
Mohammad Ishaque831002,CIVIL SURGEON OFFICE, SADAR HOSPITAL CAMPUS, KHASMAHAL, EAST SINGHBHUMSaraikela KharsawanJharkhand831002130-

Authority Records

HEALTH MEDICAL EDUCATION AND FAMILY WELFARE DEPARTMENT

BID & GeM Expert Consultancy

End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.

Bid Preparation GeM Registration Document Filing

Free consultation · 24h response

Documents 4

GeM-Bidding-9795238.pdf

Main Document

CATALOG-Specification-1

CATALOG Specification

Other Documents

OTHER

GEM General Terms and Conditions Document

GEM_GENERAL_TERMS_AND_CONDITIONS

Categories 1

🤖 AI-Powered Bidder Prediction

Discover companies most likely to bid on this tender

Live AI
Historical Data

Required Documents

1

Experience Criteria

2

Past Performance

3

Bidder Turnover

4

Certificate (Requested in ATC)

5

OEM Authorization Certificate

6

OEM Annual Turnover

7

Additional Doc 1 (Requested in ATC) *In case any bidder is seeking exemption from Experience / Turnover Criteria

8

the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer

Technical Specifications 1 Item

Item #1 Details

View Catalog
Category Specification Requirement
GENERAL Product Description ICU Ventilator
PRODUCT INFORMATION Patient Type Adult & pediatric, Adult, Pediatric and Neonatal
PRODUCT INFORMATION Type of technology Compressor
PRODUCT INFORMATION Tidal volume in ml (50 ml or less) to (1500 ml or more), (2 ml or less) to (1500 ml or more), (20 ml or less) to (1500 ml or more)
PRODUCT INFORMATION Respiration rate, breaths/min upto 150 or more
PRODUCT INFORMATION Trigger mechanism Both pressure and flow
PRODUCT INFORMATION Inspiratory flow rate, L/min up to 150 or more
PRODUCT INFORMATION Inspiratory pressure, cm H2O upto 80 or more
PRODUCT INFORMATION IE ratio 1:10 to 4:1
PRODUCT INFORMATION Sigh breath function No
PRODUCT INFORMATION Control panel lock Yes
PRODUCT INFORMATION Facility for double lung ventilation No
PATIENT ASSESSMENT TOOLS Number of loops 2 loops (P-V, F-V) with facility of saving of 1 Loops for reference
PATIENT ASSESSMENT TOOLS Lung recruitment tools (PV loops) / OLT (Open lung tools) Yes
PATIENT ASSESSMENT TOOLS Lung mechanics visualization tool No
PATIENT ASSESSMENT TOOLS Capnography/CO2 monitoring No
PATIENT ASSESSMENT TOOLS Esophageal / transpulmonary pressure monitoring No
PATIENT ASSESSMENT TOOLS APRV(Airway Pressure Release Ventilation) /Bivent/Bi level Yes
PATIENT ASSESSMENT TOOLS MMV(Mandatory Minute Ventilation) + PSV (Pressure Support Ventilation) /ASV( Adaptive Support Ventilation ) or ALPV(Adaptive Lung Protection Ventilation) or Automode No
PATIENT ASSESSMENT TOOLS PRVC(Pressure-Regulated Volume Control ventilation)/Auto flow Yes
PATIENT ASSESSMENT TOOLS High flow oxygen therapy Yes
PATIENT ASSESSMENT TOOLS Automatic weaning system provided No
PATIENT ASSESSMENT TOOLS Volume Support Yes
PATIENT ASSESSMENT TOOLS RSBI (Rapid Shallow Breathing Index) Yes
PATIENT ASSESSMENT TOOLS Expiratory Time constant ( Tcexp) Yes
PATIENT ASSESSMENT TOOLS Compliance static and/or dynamic Yes
PATIENT ASSESSMENT TOOLS Inspired and/or expired resistance Yes
PATIENT ASSESSMENT TOOLS Occlusion pressure No
PATIENT ASSESSMENT TOOLS Inspiratory and expiratory hold Yes
PATIENT ASSESSMENT TOOLS Spontaneous frequency Yes
PATIENT ASSESSMENT TOOLS Total peep, intrinsic peep, extrinsic peep Yes
PATIENT ASSESSMENT TOOLS Expiratory block is autoclavable and no routine calibration required Yes
PATIENT ASSESSMENT TOOLS Auto compensation for ET Tube No
PATIENT ASSESSMENT TOOLS Facility for automatic compliance & Leakage compensation for circuit available Yes
PATIENT ASSESSMENT TOOLS RS 232 out put port Yes
PATIENT ASSESSMENT TOOLS USB/Memory card No
PATIENT ASSESSMENT TOOLS Availability of Remote alarm/display port No
PATIENT ASSESSMENT TOOLS Facility to Report (vent alarms and patient status) No
PATIENT ASSESSMENT TOOLS HL/7 Compliant out put Yes
PATIENT ASSESSMENT TOOLS Type of coloured display LED/LCD/TFT
PATIENT ASSESSMENT TOOLS Display should be touch screen Yes
PATIENT ASSESSMENT TOOLS NIV (Non Invasive Ventilation) to be possible in all modes of ventilation available Yes
PATIENT ASSESSMENT TOOLS Minimum Size of display (±2%) 12.0
PATIENT ASSESSMENT TOOLS Graphic display have automatic scaling facility for waves Yes
HARDWARE AND ELECTRICAL REQUIREMENTS Provision of UPS No
HARDWARE AND ELECTRICAL REQUIREMENTS Backup time in hrs NA if not provided
HARDWARE AND ELECTRICAL REQUIREMENTS Built in air source Air compressor (external)
HARDWARE AND ELECTRICAL REQUIREMENTS Availability of stand alone compressor Yes
HARDWARE AND ELECTRICAL REQUIREMENTS Compressor shall provide an oil free Medical air, with Peak output flow should be minimum 150 LPM and Air quality complying with ISO compressed air purity class Yes
HARDWARE AND ELECTRICAL REQUIREMENTS Seller shall ensure compatibility of compressor with ventilator Yes
HARDWARE AND ELECTRICAL REQUIREMENTS Nebulizer with particle size less than 3 micron Yes
ADDITIONAL ACCESSORIES Humidifier: Servo controlled with digital monitoring of inspired gas temperature with heating wire NA
WARRANTY Warranty (years) 3, 5 Or higher