Infant Warmer (V2)
All India Institute Of Medical Sciences (aiims)
RAIPUR, CHHATTISGARH
Bid Publish Date
10-Jul-2026, 11:09 am
Bid End Date
23-Jul-2026, 4:00 pm
Location
Progress
Quantity
2
Bid Type
Two Packet Bid
The procurement is for an Infant Warmer (V2) at JIPMER Karaiakal, Puducherry, under the Department of Health and Family Welfare. The bidder must supply a baby warmer featuring a user‑friendly touch panel, rotatable warmer head for X‑ray access, height adjustment, and a comprehensive display with hygienic, liquid‑proof control panel. Certifications include MDR 2017 compliance, valid medical device license under the Drugs and Cosmetics Act, and manufacturer/testing certifications. The contract requires data sheets, OEM authorization, and a potential comprehensive warranty option. The scope emphasizes safe, uniform heat distribution and X‑ray cassette compatibility. Bidding may require evidence of prior government/PSU supply and Make in India certification. This is a regulated medical device procurement with strict documentation and testing expectations as part of JIPMER’s equipment modernization in KaraiKal. The unique aspects include explicit X‑ray readiness markings, cassette holder provisions, and dedicated alarms with 15‑minute cycle reminders for manual mode.
Product name: Infant Warmer (V2) with MDR 2017 compliance
Key features: touch panel control, rotatable warmer head, audiovisual alarms, height adjustable cradle
Safety: liquid‑proof control panel, large digital display, hygiene‑friendly disinfection
X‑ray readiness: provision for X‑ray cassette holder and markings on bassinet
Accessories: reusable and disposable skin temperature probes, storage drawer, X‑ray tray
Standards: MDR 2017, Electrical Safety Compliance; Medical Device license under Drugs and Cosmetic Act 1940
Warranty: option for comprehensive warranty if bid includes it
EMD viability depends on MSME/NSIC exemption; otherwise payment security is required
Delivery and installation at JIPMER KaraiKal, Puducherry UT following contract
Mandatory data sheet, OEM authorization, and MDR‑related test reports with bid
Not specified in data; bidders should prepare for standard govt terms including milestone payments and possible advance per contract
Delivery and installation to JIPMER KaraiKal, Puducherry UT; ensure on‑time performance and commissioning
Not detailed; bidders should anticipate LDs for late delivery or non‑conformance as per buyer terms
3+ years in supplying neonatal/infant warmer or comparable medical devices to government/PSU
MDR 2017 compliance, valid Medical Device license under Drugs and Cosmetic Act, and electrical safety certification
OEM authorization and Make in India certification attached
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
No
Item Category
Infant Warmer (V2) (Q2)
Max Delivery Days
15
Delivery Locations
1
Delivery Cities
Karaikal
Delivery Pincodes
609602
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Leelapalli Anusha | 609602,WRPQ+3J3, Kamarajar Salai Ext, Karaikal, Puducherry 609602 | Karaikal | Puducherry | 609602 | 2 | 15 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS
All India Institute Of Medical Sciences (aiims)
RAIPUR, CHHATTISGARH
All India Institute Of Medical Sciences (aiims)
PATNA, BIHAR
South Central Railway
N/a
BAREILLY, UTTAR PRADESH
N/a
PONDICHERRY, PUDUCHERRY
Tender Results
Loading results...
| Category | Specification | Requirement |
|---|---|---|
| PRODUCT INFORMATION | User friendly touch panel control | Yes |
| PRODUCT INFORMATION | Type of heater | Ceramic / High Alloy Steel / Quartz infrared / Calrod heater |
| PRODUCT INFORMATION | Audiovisual alarm facility | Overheating beyond set temperature range, Spurt in Heater output i.e. more than 60% heater output for 10 minutes system should raise alarm and cutoff heater, Power failure, Heater failure, Patient temperature less/Greater than required/set temperature, Probe failure, Temperature Probe disconnect |
| PRODUCT INFORMATION | Warmer head should be rotatable in different direction, so as to allow taking X-ray | Yes |
| PRODUCT INFORMATION | Colour temperature | 3700K to 5100K |
| PRODUCT INFORMATION | Facility to lock the keyboard to avoid unwanted user modification of the set parameters | Yes |
| PRODUCT INFORMATION | Should have a Feather Touch operation with large digital display and comprehensive alarms, Control Panel should be liquid proof and allow easy and hygienic disinfection | Yes |
| PRODUCT INFORMATION | In Manual Mode an auditory and visual alarm shall be given at least every 15 minutes | Yes |
| PRODUCT INFORMATION | Height Adjustable feature | Manual |
| PRODUCT INFORMATION | Provision for X-Ray cassette holder | Yes |
| PRODUCT INFORMATION | Markings on the bassinet and X-Ray cassette holder to enable proper positioning of the baby while doing the X-Ray | Yes |
| Settings | Settings with facility to clearly display the selected mode | Manual mode |
| Settings | User's interface | Manual and Servo controlled temperature regulation |
| Settings | Software and/or standard of communication(where ever required) | Inbuilt software; Interruption and restoration of the power supply does not change the preset values |
| PHYSICAL CHARACTERISTICS | Should maintain desired set temperature and uniform heat distribution in baby cradle | Yes |
| ACCESSORIES | Number of Skin temperature probe (Reusable) | 10 |
| ACCESSORIES | Number of Skin temperature probe (Disposable) | 50, Not Provided |
| ACCESSORIES | Number of Storage drawer | 2, Not Provided |
| ACCESSORIES | Number of X- ray tray | 1 |
| CERTIFICATIONS | Compliance to Medical Device Rules (MDR) 2017 as amended till date | Yes |
| CERTIFICATIONS | Availability of valid Medical Device license for the product issued from the competent authority defined under Drugs and Cosmetic Act 1940 and Rules made there under as amended till date | Yes |
| CERTIFICATIONS | Certification for manufacturing unit | ISO:13485 (Latest) |
| CERTIFICATIONS | Availability of Test Report for each supplied batch/product as per Medical Device Rule (MDR) 2017 as amended till date | Yes |
| CERTIFICATIONS | Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission or along with supplies as per buyer requirement | Yes |
| CERTIFICATIONS | Electrical Safety Compliance Standard | IEC 60601-1-2 or Equivalent BIS |
| WARRANTY | Warranty (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) | 5 Or higher |
Discover companies most likely to bid on this tender
GST registration certificate
PAN card
Experience certificates for similar government/PSU supply (3+ years) with contract copies
Financial statements and turnover proof
EMD/Security deposit (or MSME/NSIC exemption certificate)
OEM authorization / Manufacturer Authorization
Data Sheet of offered infant warmer with reference parameters
Make in India authorization certificate
Electrical Safety Compliance certificate
MDR compliance certificates and Medical Device license
Test reports for each batch/product as per MDR 2017
Malicious Code certificate
Submission of all required certifications, licenses, and test reports at bid submission
Bidders must meet MDR 2017 compliance, provide Medical Device license under the Drugs and Cosmetic Act, and submit OEM authorization, data sheets, and test reports. Include 3+ years of government/PSU supply experience, Make in India certificate, and EMD documentation (if applicable). Delivery to JIPMER KaraiKal is required.
Submit GST, PAN, 3+ years experience proof with contracts, OEM authorization, MDR certificates, electrical safety certificate, Make in India authorization, data sheets, and test reports for every batch. Include Malicious Code certificate and vendor licenses as per buyer terms.
The warmer must have a user friendly touch panel, rotatable head for X‑ray access, height adjustability, audible/visual alarms every 15 minutes in manual mode, liquid‑proof control panel, and X‑ray cassette holder with positioning markings. Includes reusable and disposable skin temperature probes.
Bidders must provide MDR 2017 compliance documentation, valid Medical Device license under the Drugs and Cosmetic Act, plus batch test reports as per MDR. Ensure all certifications are current and verifiable at bid submission or delivery.
Delivery and installation should be completed at JIPMER KaraiKal, Department of Obstetrics and Gynecology, Third Floor Academic Building, 609602, Puducherry UT, per purchaser terms. Ensure coordination for on‑site commissioning and handover with hospital Biomedical teams.
Submit a valid Make in India authorization certificate from the OEM or authorized distributor, with complete entity details, product model, and scope of supply. This must accompany bid submission to demonstrate local manufacturing or authorized capability for the infant warmer.
EMD exemptions apply if the bidder holds MSME/NSIC registration with an exemption certificate. If not, provide standard EMD as per tender terms. Ensure certificate validity and submission alongside other bid documents.
A comprehensive warranty option may be offered in bidding, which would supersede the standard catalogue warranty if selected. Verify coverage period, components included, and AMC terms during bid submission.
Sanjay Gandhi Post Graduate Institute Of Medical Sciences Lucknow
📍 LUCKNOW, UTTAR PRADESH
Jawaharlal Institute Of Postgraduate Medical Education And Research (jipmer)
📍 KARAIKAL, PUDUCHERRY
Southern Railway
📍 CHENNAI, TAMIL NADU
All India Institute Of Medical Sciences (aiims)
📍 JODHPUR, RAJASTHAN
All India Institute Of Medical Sciences (aiims)
📍 UDHAM SINGH NAGAR, UTTARAKHAND
Sign up now to access all documents
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS