Progress
Quantity
1
Bid Type
Two Packet Bid
Directorate Of Medical Education & Research has released a public tender for Real Time PCR Machine (V2) (Q2) in KARNAL, HARYANA. Quantity: 1. Submission Deadline: 27-07-2026 17: 00: 00. Download documents and apply online.
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
Yes
Item Category
Real Time PCR Machine (V2) (Q2)
Advisory Bank
State Bank of India
ePBG Percentage
5%
ePBG Duration (Months)
26
Max Delivery Days
30
Delivery Locations
1
Delivery Cities
Karnal
Delivery Pincodes
132001
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Deepak Kumar | 132001,Kalpana Chawla Govt. Medical College, Karnal | Karnal | Haryana | 132001 | 1 | 30 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS
| Category | Specification | Requirement |
|---|---|---|
| GENERAL INFORMATION | Product Name | Real Time PCR Machine |
| PRODUCT INFORMATION | System Type | Open |
| PRODUCT INFORMATION | Capacity of Blocks | 96 well |
| PRODUCT INFORMATION | Number of channels (minimum) | 4 |
| PRODUCT INFORMATION | Maximum Heating ramp rate in degree per second | 4 |
| PRODUCT INFORMATION | Maximum Cooling ramp rate in degree per second | 4 |
| PRODUCT INFORMATION | Adjustable heating / cooling ramp rate | Yes Or higher |
| PRODUCT INFORMATION | Operating Temperature range (degree celcius) | 4 - 105 degree celcius |
| PRODUCT INFORMATION | Hot Lid Temperature (degree celcius) | 30 to 110 degree celcius |
| PRODUCT INFORMATION | Sample volume range in micro litre | 5 to 100 ml |
| PRODUCT INFORMATION | Programmable Steps and cycles | Yes |
| PRODUCT INFORMATION | Number of USB ports | 2 |
| PRODUCT INFORMATION | Pause / Start function | Yes |
| PRODUCT INFORMATION | Input power supply | Single phase (230 V, 50 Hz) |
| PRODUCT INFORMATION | Auto restart after power outages | Yes |
| PRODUCT INFORMATION | Boot up time | ≤1 minute |
| PRODUCT INFORMATION | Type of chemistries that can be run on system | Taqman, SYBR green, Molecular Beacon and all other fluorescent dye based |
| PRODUCT INFORMATION | Compatibility for well strips | 96 |
| PRODUCT INFORMATION | Compatibility for Individual PCR tubes | 0.1 to 0.2 ml tubes |
| PRODUCT INFORMATION | Source of Excitation | Tungsten / Xenon / LED / Halogen |
| PRODUCT INFORMATION | Number of excitation filters | ≥ 5, < 5 |
| PRODUCT INFORMATION | Number of emission filters | ≥ 5, < 5 |
| PRODUCT INFORMATION | Multiplexing ability dyes in a single run | ≥ 5, < 5 |
| PRODUCT INFORMATION | Pre calibration with at least 7 commonly used dyes | FAM / SYBR Green / VIC / HEX / NED / TAMRA / ROX / Texas Red / JOE / Cy5 / Quasar 670 / Cy5.5 / Quasar 705 / Cy3 |
| PRODUCT INFORMATION | Addition of new dyes should be possible without hardware change | Yes |
| PRODUCT INFORMATION | Analysis performed | Gene Expression, Plus/minus assay, SNP, Allelic Discrimination and Dissociation Curve Analysis, DNA Quantitation, Gene Expression Analysis, Comparative Analysis Curve, Standard Curve/ Relative Standard Curve, Capability for calculating dilution & reaction set up for HID kits, Sample Quality Assessment (Human male; female ratio inhibitions & degradation index) |
| PRODUCT INFORMATION | Instrument software should not restrict number of assay or target that can be run on a single 96 well plate in parallel | Yes |
| PRODUCT INFORMATION | Suitable software for Data Acquisition, analysis of run and also for Gene Expression analysis by relative quantity or normalized expression | Yes |
| PRODUCT INFORMATION | Software should have the capacity to analyze data of minimum 10 different runs at a time | Yes |
| PRODUCT INFORMATION | The instrument software must be capable of detecting and analyzing a different gene, SNP or pathogen target in every well of the 96-well plate | Yes |
| PRODUCT INFORMATION | Temperature setting accuracy (degree celcius) | ± 0.1 |
| PRODUCT INFORMATION | Well to well temperature uniformity (degree celcius) | ± 0.3 |
| DATA MANAGEMENT SYSTEM | Computer system with latest licensed operating system and antivirus provided | Yes |
| DATA MANAGEMENT SYSTEM | PC monitor type | LCD / LED |
| DATA MANAGEMENT SYSTEM | Minimum PC monitor size (inches) | 15 Inches or more |
| DATA MANAGEMENT SYSTEM | PC hard disk | 500 GB, 1 TB Or higher |
| DATA MANAGEMENT SYSTEM | UPS with power backup (Minutes) (minimum) | 60 minutes |
| CERTIFICATIONS | Compliance to Medical Device Rules (MDR) 2017 as amended till date | Yes |
| CERTIFICATIONS | Availability of valid Medical Device license for the product issued from the competent authority defined under Drugs and Cosmetic Act 1940 and Rules made there under as amended till date | Yes |
| CERTIFICATIONS | Certification for manufacturing unit | ISO:13485 (Latest) |
| CERTIFICATIONS | Availability of Test Report for each supplied batch/product as per Medical Device Rule (MDR) 2017 as amended till date | Yes |
| CERTIFICATIONS | Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission or along with supplies as per buyer requirement | Yes |
| CERTIFICATIONS | Electrical Safety Compliance Standard | IEC 60601 or Equivalent BIS |
| WARRANTY | Warranty in Years (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) | 2, 3, 4, 5 Or higher |
Discover companies most likely to bid on this tender
Experience Criteria
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
Sign up now to access all documents
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS