Assam Medical Services Corporation Limited invites bids for Immunoassay Analyzer (V3) (Q2) in DIBRUGARH, ASSAM. Quantity: 9. Submission Deadline: 22-07-2026 17: 00: 00. Submit your proposal before the deadline.
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
No
Item Category
Immunoassay Analyzer (V3) (Q2)
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
GEM_CATEGORY_SPECIFICATION
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS
| Category | Specification | Requirement |
|---|---|---|
| PRODUCT INFORMATION | Type of Configuration | Benchtop, Floor mounted |
| PRODUCT INFORMATION | Type of system configuration | Standalone |
| PRODUCT INFORMATION | Type of system | Closed |
| PRODUCT INFORMATION | Processing modes | Continuous random access |
| PRODUCT INFORMATION | STAT | Yes Or higher |
| PRODUCT INFORMATION | Type of Automation | Fully automated |
| PRODUCT INFORMATION | Throughput of the system per hour | ≥ 200 |
| PRODUCT INFORMATION | Sample Loading Capacity at a time | ≥ 30 |
| PRODUCT INFORMATION | Type of samples | Serum, Plasma, Whole blood, Body fluids |
| PRODUCT INFORMATION | Sample volume per test in µL | ≥ 1 µL |
| PRODUCT INFORMATION | Type of method | Chemiluminescence (CLIA) |
| Test Menu | Cardiac Markers | Yes |
| Test Menu | Tests which can be performed by Cardiac Markers | CK-MB, Troponin I / Troponin-T, Myoglobin, hs-cTnl, Hs - Troponin -I, BNP |
| Test Menu | Infectious Bio Markers | Yes |
| Test Menu | Tests which can be performed by Bio Markers | Nephrocheck |
| Test Menu | Tumor Markers | Yes |
| Test Menu | Tests which can be performed by Tumor Markers | AFP, CEA, Total PSA, f-PSA, CA 125, CA 15.3, CA 19-9, CYFRA 21-1, IL-6, KL-6, HE4, ProGRP, PIVKA |
| Test Menu | Anemia profile | Yes |
| Test Menu | Tests which can be performed by Anemia profile | Ferritin |
| Test Menu | Bone Metabolism | Yes |
| Test Menu | Tests which can be performed by Bone metabolism | 25-OH Vitamin D |
| Test Menu | Thyroid profile | Yes |
| Test Menu | Tests which can be performed by thyroid profile | TSH, T4, T3, FT4, FT3, TGA (Anti - tg), Anti - TPO, Intact PTH |
| Test Menu | Infertility | Yes |
| Test Menu | Tests which can be performed for infertility | FSH, LH, HCG/ß-HCG, PRL, Estradiol, Progesterone, Testosterone, AMH |
| Test Menu | Infectious Test | Yes |
| Test Menu | Tests which can be performed for Infectious Test | Anti HCV, HBsAg, HIV Combi, Syphilis |
| Test Menu | Dual, Triple or Quad prenatal screening tests | Yes |
| Test Menu | Tests which can be performed for prenatal screening | AFP (prenatal screening), Free ß - HCG, PAPP - A, HCG/ß-HCG, Free Estriol, PLGF, SFLT |
| Test Menu | Liver Fibrosis Tests | Yes, No |
| Test Menu | Tests which can be performed for Liver Fibrosis | ELF, NA |
| Test Menu | Alzheimer Screening | Yes, No |
| Test Menu | Tests which can be performed for Alzheimer Screening | Beta Amyloyd, P.Tau, NA |
| Test Menu | Neuro Markers | Yes, No |
| Test Menu | Tests which can be performed for Neuro Markers | AB40, AB42, T Tau, P Tau, S100, NA |
| Test Menu | Immuno Suppressive drugs | Yes, No |
| Test Menu | Tests which can be performed for Immuno Suppressive drugs | Tacrolimus, Everolimus, Serolimus, Cyclosporin, NA |
| SYSTEM FEATURES | Closed-tube sampling | Yes, No Or higher |
| SYSTEM FEATURES | Direct tube sampling | Yes Or higher |
| SYSTEM FEATURES | Liquid-level sensing | Yes |
| SYSTEM FEATURES | Clot, Bubble, Viscosity detection | Yes |
| SYSTEM FEATURES | Auto dilution | Yes Or higher |
| SYSTEM FEATURES | Abnormal values flag | Yes Or higher |
| SYSTEM FEATURES | Autowash | Yes Or higher |
| SYSTEM FEATURES | Ultrasonic wash technique to prevent carryover | Yes, No Or higher |
| SYSTEM FEATURES | Disposable tips to prevent any carryover of sample | Yes, No Or higher |
| SYSTEM FEATURES | Both quantitative and qualitative analysis of results | Yes Or higher |
| SYSTEM FEATURES | Inbuilt facility for reagent mixing | Yes Or higher |
| SYSTEM FEATURES | Availability of inbuilt refrigeration system for long onboard stability of reagents | Yes Or higher |
| SYSTEM FEATURES | Module for reading reagents | RFID |
| SYSTEM FEATURES | Facility to collect both solid and liquid waste for disposal | Yes |
| SYSTEM FEATURES | Type of Calibration | Automatic and Manual |
| SYSTEM FEATURES | Calibration stability | ≥ 14 |
| SYSTEM FEATURES | LOT specific calibration | Yes Or higher |
| SYSTEM FEATURES | Should have inbuilt quality control system to monitor the quality results obtained and LJ charts must be visible | Yes Or higher |
| DATA MANAGEMENT | Type of data management | External PC |
| DATA MANAGEMENT | Type of display | LED/LCD/TFT |
| DATA MANAGEMENT | Size of Display in inches | ≥ 10 Inches |
| DATA MANAGEMENT | Touch Screen display | No |
| DATA MANAGEMENT | Handheld barcode reader | Yes |
| DATA MANAGEMENT | HIS/LIS Interface | RS 232 serial, Ethernet |
| DATA MANAGEMENT | Data Storage of sample | 1000 or more |
| DATA MANAGEMENT | Should be provided with inbuilt/external printer with necessary accessories | Yes |
| DATA MANAGEMENT | Type of External storage | USB |
| ACCESSORIES | Startup Reagent Kit to be provided | Yes |
| ACCESSORIES | Two set of all tubings | Yes Or higher |
| ACCESSORIES | Alerts for operator for level of reagents and to empty waste when indicated | Yes Or higher |
| POWER REQUIREMENTS | Type of power supply in VAC,Hz | 200-240 VAC, 50-60 Hz |
| POWER REQUIREMENTS | UPS Backup in minutes | 30 Minutes, 60 Minutes |
| CERTIFICATION AND REPORTS | Compliance to Drugs and Cosmetic Act 1940 and Medical Device Rules (MDR) 2017 as amended till date | Yes |
| CERTIFICATION AND REPORTS | Availability of valid Medical Device license for the product issued from the competent authority | Yes |
| CERTIFICATION AND REPORTS | License issuing authority as per MDR | Central Licensing Authority, State Licensing Authority |
| WARRANTY | Warranty in years (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) | 1 Year, 2 Year |
Discover companies most likely to bid on this tender
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover
Additional Doc 1 (Requested in ATC)
Additional Doc 2 (Requested in ATC)
Compliance of BoQ specification and supporting document *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
Extended Deadline
22-Jul-2026, 5:00 pm
Opening Date
22-Jul-2026, 5:30 pm
Extended Deadline
15-Jul-2026, 5:00 pm
Opening Date
15-Jul-2026, 5:30 pm
Extended Deadline
09-Jul-2026, 5:00 pm
Opening Date
09-Jul-2026, 5:30 pm
Extended Deadline
02-Jul-2026, 5:00 pm
Opening Date
02-Jul-2026, 5:30 pm
Extended Deadline
25-Jun-2026, 5:00 pm
Opening Date
25-Jun-2026, 5:30 pm
Extended Deadline
18-Jun-2026, 5:00 pm
Opening Date
18-Jun-2026, 5:30 pm
Sign up now to access all documents
Main Document
GEM_CATEGORY_SPECIFICATION
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS