Progress
Eastern Railway invites bids for Digital Radiography System (V2) (Q2) in MUNGER, BIHAR. Quantity: 1. Submission Deadline: 03-09-2026 19: 00: 00. Submit your proposal before the deadline.
Quantity
1
Bid Type
Single Packet Bid
Bid Validity
90 (Days)
Bid Type
Service
Evaluation
Total value wise evaluation
Inspection Required
Yes
Tech Clarification Time
2 Days
EMD Required
Yes
MII Purchase Preference
Yes
MII Preference Band
L1+20%
MSE Purchase Preference
Yes
MSE Preference Band
L1+15%
MSE Exemption/Relaxation
No
Startup Exemption/Relaxation
No
Bid Splitting Applied
No
Warranty Period
2 years
Arbitration Clause
No
Mediation Clause
No
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
No
Item Category
Digital Radiography System (V2) (Q2)
Advisory Bank
State Bank of India
ePBG Percentage
5%
ePBG Duration (Months)
62
Max Delivery Days
60
Delivery Locations
1
Delivery Cities
Munger
Delivery Pincodes
811214
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Jagat Kishore Prasad | 811214,EASTERN RAILWAY MAIN HOSPITAL,JAMALPUR, DISTT- MUNGER (BIHAR) PIN 811214 | Munger | Bihar | 811214 | 1 | 60 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
OTHER
GEM_GENERAL_TERMS_AND_CONDITIONS
Discover companies most likely to bid on this tender
Certificate (Requested in ATC)
OEM Authorization Certificate
Additional Doc 1 (Requested in ATC)
Additional Doc 2 (Requested in ATC)
Additional Doc 3 (Requested in ATC)
Additional Doc 4 (Requested in ATC) *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
| Category | Specification | Requirement |
|---|---|---|
| Table and Table top parameters | Type of table | Elevating Type, Floating Top Type |
| Table and Table top parameters | Type of table movement | Motorized, Manual |
| Table and Table top parameters | Number of movements of Table | 4 Way, 6 Way |
| Table and Table top parameters | Portability of table | No, Yes |
| Table and Table top parameters | Length x Width (± 5%) in cm | 210 x 70, 240 x 93 |
| Table and Table top parameters | Minimum and maximum adjustable heights (± 10%) | 55-85, Not applicable |
| Table and Table top parameters | Load bearing capacity of table in kg | 200 or more, 300 or more |
| Table Bucky System | Type of Table bucky movement | Auto tracking and auto synchronization, Auto tracking, Auto tracking and auto positioning, Not Available |
| Table Bucky System | Bucky rotatable | No, Yes |
| Table Bucky System | Size in inches | 17x17, 14x17 |
| Table Bucky System | AEC | 3 Field or higher, NA |
| Table Bucky System | Grid Ratios | 10:1 or higher |
| Table Bucky System | Longitudinal travel | Manual, Motorized |
| Upright Bucky | Type | Manual vertical Bucky Stand, Motorized, Motorized tilting, Not Available |
| Upright Bucky | Sizes in inches | 17x17, 14x17 |
| Upright Bucky | A E C | 3 Field or higher |
| Upright Bucky | Tilt | No, Yes |
| Digital Detector | Number of detectors | Single, Dual |
| Digital Detector | Detector Material | Cesium Iodide |
| Digital Detector | Detector Size in inches | 17 x 17, 14 x 17, 17 X 17 and 17 x 17, 14 x 17 and 14 x 17, 17 x 17 and 14 x 17 |
| Digital Detector | Matrix size | 2kx2k, 3kx3k, 2.3kx2.3k, 2.8k x 2.8k, 3.5k x 4.2k |
| Control panel for image acquisition | Control panel entry | Touch screen, Alphanumeric key board |
| Control panel for image acquisition | Virtual grid software | No, Yes |
| Control panel for image acquisition | Grid removable | No, Yes |
| X Ray Generator and Tubes | High Frequency X-ray Generator power in KW | 80 KW, 40 KW, 50 KW, 65 KW, 32 KW |
| X Ray Generator and Tubes | Maximum mA at 100 KVP | 500, 600, 800, 300 |
| X Ray Generator and Tubes | Focal spot | 0.6 and 1.2, 0.6 and 1, 1 and 2 |
| X Ray Generator and Tubes | Heat Capacity, KHU | 400-600, 300-400, 500-800, 140-200 |
| X Ray Generator and Tubes | Automatic tube positioning | No, Yes |
| Tube suspension | Tube mounted X Ray control | No, Yes |
| Tube suspension | Type of suspension | Ceiling, Floor Mounted |
| Tube suspension | Collimator Type | Manual, Automatic |
| Workstation | Monitor size in inches | 19, 21 |
| Workstation | Processor | i 3, i 5, i 7, Intel Xenon |
| Workstation | RAM in GB | 2, 4, 8, 16 |
| Workstation | HDD | 1 TB, 1 TB RAID |
| Accessories | Hand grips | Yes, Not provided |
| Accessories | Foot switches for adjusting longitudinal/side to side movements, locking, and light adjustment | Yes, Not provided |
| Accessories | Patient fixing belts and compression device (for performing excretory urography, etc) | No, Yes |
| Accessories | Stitching Hardware and Software | No, Yes |
| Accessories | Integration | DICOM 3.0 storage SOP Classes, NA |
| Accessories | DAP Meter provided | No, Yes |
| Accessories | Lead Apron | No, Yes |
| Accessories | Light weight ‘zero lead’ radiation protection aprons provided | No, Yes |
| Accessories | Lead Aprons - Full body and Skirt type | 2 no. each (0.5pb), Not provided |
| Accessories | Thyroid shields provided | Yes, Not provided |
| Accessories | Gonadsheet Male provided | Yes, Not provided |
| Accessories | Gonad sheet Female provided | Yes, Not provided |
| Accessories | Wall mounted stand for radiation protection aprons supplied | Yes, Not provided |
| Accessories | Lead glass window provided | Yes, Not provided |
| Accessories | Footsteps for the table provided | Yes, Not provided |
| Accessories | Power | 230 V 50 HZ Single phase -16 Ampls socket e, 440 V 50 HZ 3 Phase 16 Ampls socket |
| Accessories | Voltage stablizer servo controlled in KVA | >=50, Not provided |
| Certifications | Compliance to Medical Device Rules (MDR) 2017 as amended till date | Yes |
| Certifications | AERB Type Approved Equipment | Yes |
| Warranty | Warranty in Years | 2, 3, 4, 5 Or higher |