3 Part Automated Hematology Analyzer (V2)
N/a
VADODARA, GUJARAT
Progress
Quantity
2
Bid Type
Two Packet Bid
N/a announces a tender for 3 Part Automated Hematology Analyzer (V2) (Q2) in AMRELI, GUJARAT. Quantity: 2. Submission Deadline: 27-02-2026 09: 00: 00. Last date to apply is approaching fast!
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS
N/a
VADODARA, GUJARAT
N/a
JAMNAGAR, GUJARAT
N/a
NAVSARI, GUJARAT
Bhavnagar Municipal Corporation
BHAVNAGAR, GUJARAT
Indian Army
JHANSI, UTTAR PRADESH
Tender Results
Loading results...
| Category | Specification | Requirement |
|---|---|---|
| GENERAL | Product Description | 3 Part Automated Hematology Analyzer |
| PRODUCT INFORMATION | Type of Configuration | Bench top |
| PRODUCT INFORMATION | Type of system offered | Closed system |
| PRODUCT INFORMATION | Type of Automation | Fully Automatic |
| PRODUCT INFORMATION | Automatic Start Up,shut down and sample analysis | Yes |
| PRODUCT INFORMATION | Analysis principle | Based on principle of counting and sizing |
| PRODUCT INFORMATION | Multi channel analysis for better resolution | Yes |
| PRODUCT INFORMATION | Type of cell counting | 3 Part WBC Differential |
| PRODUCT INFORMATION | Testing mode selection available | CBC Mode |
| PRODUCT INFORMATION | Analysis available | WBC, Lymph#, Lymph%, Granulocytes #, Granulocytes%, Mixed #, Mixed%, RBC, HGB, HCT, MCV, PCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PDW-SD, PDW-CV, PCT, PLCR, P-LCC |
| PRODUCT INFORMATION | Analysis method for WBC | Electrical Impedance |
| PRODUCT INFORMATION | Method for platelet measurement | Electrical Impedance |
| PRODUCT INFORMATION | RBC Measurement method | Electrical Impedance |
| PRODUCT INFORMATION | Maximum sample aspiration volume needed in any of modes | Less than 50 µl |
| PRODUCT INFORMATION | Minimum sample volume required | 20 µl or less |
| PRODUCT INFORMATION | Throughput capacity of analyser in (samples/ hour) | 50-60 |
| PRODUCT INFORMATION | Types of reagents | 1 Hemolyzing Reagent,1 Diluent,1 Cleansing Solution |
| PRODUCT INFORMATION | Type of Calibration | Both (automatic and manual) |
| PRODUCT INFORMATION | Floating discriminator for platelets and RBC counting for reliable RBC and PLT data | Yes |
| PRODUCT INFORMATION | Automatic probe wipe | Yes |
| PRODUCT INFORMATION | Separate diluting nozzles for RBC and WBC | Yes |
| PRODUCT INFORMATION | Double bathing mechanism | Yes |
| PRODUCT INFORMATION | Automatic electric clog removal | Yes |
| DATA MANAGEMENT AND DISPLAY | Type of data management | Inbuilt system |
| DATA MANAGEMENT AND DISPLAY | Display | LCD, LED Or higher |
| DATA MANAGEMENT AND DISPLAY | Inbuilt monitor size in inches | More than 5 |
| DATA MANAGEMENT AND DISPLAY | Type of external storage | USB |
| DATA MANAGEMENT AND DISPLAY | Number of USB Port | 4 |
| DATA MANAGEMENT AND DISPLAY | Data management systems | Provide histograms in display and print |
| DATA MANAGEMENT AND DISPLAY | Facility for user defined flagging | Yes |
| DATA MANAGEMENT AND DISPLAY | Type of user Interface or data entry | All three (touchscreen, handheld barcode reader facility and manual) |
| DATA MANAGEMENT AND DISPLAY | Database capability of storing sets of results and graphics | ≥ 50000 |
| DATA MANAGEMENT AND DISPLAY | Facility for workload recording | Yes |
| DATA MANAGEMENT AND DISPLAY | Auto stop function in event of unacceptable control data | Yes |
| DATA MANAGEMENT AND DISPLAY | Ability to transmit results to host computer | Yes |
| DATA MANAGEMENT AND DISPLAY | Have auto cleaning function in the analyser’s software | Yes |
| DATA MANAGEMENT AND DISPLAY | Type of printer unit | Inbuilt |
| DATA MANAGEMENT AND DISPLAY | Printer type | Thermal Printer |
| POWER REQUIREMENTS | Type of power supply | 230-240 VAC,50-60 Hz |
| POWER REQUIREMENTS | Power Backup facility | Yes |
| POWER REQUIREMENTS | Type of UPS | Online |
| POWER REQUIREMENTS | Rating of UPS in KVA | 1 |
| Accessories, spare parts and consumables | Offered equipment unit to be supplied with sufficient consumables (with at least 2/3rd of total shelf life) required for, sufficient to carry out haematological testing of samples | 1000 |
| Accessories, spare parts and consumables | Two set of all tubings | Yes |
| Accessories, spare parts and consumables | Reagent expiry time should be minimum of 1 year | Yes |
| Accessories, spare parts and consumables | Alerts for operator for level of reagents and to empty waste when indicated | Yes |
| Accessories, spare parts and consumables | Availability of micro capillary adapter | Yes |
| Accessories, spare parts and consumables | Operating temperature and humidity | Capable of operating continuously in ambient temperature of 15 to 35 deg C and relative humidity of 15 to 85% in ideal circumstances |
| CERTIFICATION AND REPORTS | Availability of valid drug license for the product issued from the competent authority defined under Drugs and Cosmetic Act 1940 and Rules made there under as amended till date | Yes |
| CERTIFICATION AND REPORTS | Manufacturing unit certification | ISO:13485 (Latest) |
| CERTIFICATION AND REPORTS | Availability of Test Report for each supplied batch/product as per Medical Device Rules (MDR) 2017 as amended till date | Yes |
| CERTIFICATION AND REPORTS | Electrical Safety Standards | IEC/EN 60601-1 or equivalent BIS Standard |
| CERTIFICATION AND REPORTS | Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission and/or along with supplies as per buyer requirement | Yes |
| WARRANTY | Warranty in years (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) | 3, 5 Or higher |
| Miscellaneous Parameters | User/Technical/Maintenance manuals to be supplied in English in hard and soft copy | Yes |
| Miscellaneous Parameters | Details of equipments and procedures required for local calibration and routine maintenance to be supplied and advanced maintenance task documentation also to be furnished | Yes |
| Miscellaneous Parameters | List of important spares and accessories, with their part numbers to be supplied to the buyer at the time of supplying the equipment | Yes |
| Miscellaneous Parameters | OEM/Reseller shall ensure uninterrupted availability of all spares for 10 years | Yes |
| Miscellaneous Parameters | Availability of toll free facility for technical support maintened by OEM or authorized agencies | Yes |
| Miscellaneous Parameters | Installation and demonstration of equipment and training to be provided after completing supplies before acceptance | Yes |
| Miscellaneous Parameters | Calibration certificates as per NABH requirement | Yes |
| Miscellaneous Parameters | Time to attend breakdown calls | within 48 hrs |
Discover companies most likely to bid on this tender
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover
Additional Doc 1 (Requested in ATC)
Compliance of BoQ specification and supporting document *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
Extended Deadline
27-Feb-2026, 9:00 am
Opening Date
27-Feb-2026, 9:30 am
Extended Deadline
17-Feb-2026, 12:00 pm
Opening Date
17-Feb-2026, 12:30 pm
Extended Deadline
14-Feb-2026, 12:00 pm
Opening Date
14-Feb-2026, 12:30 pm
Sign up now to access all documents
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS