Directorate Of Medical Education CO2 Incubator Tender Tamil Nadu 2025 – CO2 Incubator Equipment Procurement
Directorate Of Medical Education
CHENNAI, TAMIL NADU
Bid Publish Date
09-Feb-2026, 12:09 pm
Bid End Date
02-Mar-2026, 2:00 pm
Location
Progress
Quantity
1
Bid Type
Two Packet Bid
Directorate Of Purchase And Stores invites bids for CO2 Incubator (Q3) in MUMBAI, MAHARASHTRA. Quantity: 1. Submission Deadline: 02-03-2026 14: 00: 00. Submit your proposal before the deadline.
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS
Directorate Of Medical Education
CHENNAI, TAMIL NADU
Sree Chitra Tirunal Institute For Medical Sciences And Technology (sctimst)
THIRUVANANTHAPURAM, KERALA
Dr Bhubanesar Borooah Cancer Institute Bbci
KAMRUP, ASSAM
Rajiv Gandhi Centre For Biotechnology (rgcb)
THIRUVANANTHAPURAM, KERALA
Indian Army
PUNE, MAHARASHTRA
Tender Results
Loading results...
| Category | Specification | Requirement |
|---|---|---|
| Performance Parameters | Temperature management range in degree Celsius | 5 to 55 ºC |
| Performance Parameters | Temperature accuracy | ±0.1°C at 37 ºC, ±0.1°C |
| Performance Parameters | Temperature stability | ±0.1°C at 37°C |
| Performance Parameters | Gas uniformity | ±0.1% at set 5.0% CO2 concentration at 37 ºC set temperature |
| Performance Parameters | CO2 recovery rate | Under 6 min after door opening and closing event to attain 5% CO2. |
| Performance Parameters | Dry Heat Sterilization temperature in degree Celsius | 140 |
| Performance Parameters | Shelves thickness in mm | 1 or more |
| Performance Parameters | Minimum Time taken for CO2 recovery rate after door opening | Under 6 min after door opening and closing event to attain 5% CO2. |
| Performance Parameters | Availability of Fan less design to reduce chance of contamination, reduce noise | Yes, Fan with in chamber HEPA for better recovery and ISO class 5 clean air Or higher |
| Performance Parameters | Sensor used | Infra-Red (IR) and NDIR type CO2 Sensor |
| Performance Parameters | Maximum temperature for Sensor functionality | 140 ºC or more during dry heat sterilizations |
| Reports | Warranty in years | 3, 4, 5 Or higher |
Discover companies most likely to bid on this tender
Certificate (Requested in ATC) *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
Indian Council Of Medical Research (icmr)
📍 CHENNAI, TAMIL NADU
WEST BENGAL MEDICAL SERVICES CORPORATION LIMITED
📍 24 PARAGANAS NORTH, WEST BENGAL
Sign up now to access all documents
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS