Bid Publish Date
04-Aug-2026, 5:23 pm
Bid End Date
01-Sep-2026, 4:00 pm
Progress
Public procurement opportunity for The issuing authority Electrocardiography (ECG) Machine (V2). Quantity: 16 issued by. Submission Deadline: 01-09-2026 16: 00: 00. View full details and respond.
Quantity
16
Bid Type
Service
EMD Required
Yes
MII Purchase Preference
No
MSE Purchase Preference
No
MSE Exemption/Relaxation
No
Startup Exemption/Relaxation
No
Bid Splitting Applied
No
Warranty Period
5 years
Tender Category
Goods
Bid To RA
No
ePBG Percentage
2%
ePBG Duration (Months)
38
Delivery Locations
4
Delivery Cities
Kakinada, Karaikal, Mahe, Pondicherry
Delivery Pincodes
533464, 605006, 609602, 673310
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Sanjay C K | 673310,Pharmacy Stores, Office of the Deputy Director, Government General Hospital, Hospital Road, Near to Church, Mahe. (UT of Puducherry). | Mahe | Puducherry | 673310 | - | - | - |
| Nrusimhadevara Durga Bala Subrahmanyam | 533464,Pharmacy Stores, Office of the Deputy Director, Government General Hospital, Jail Street, Yanam. UT of Puducherry. | Kakinada | Andhra Pradesh | 533464 | - | - | - |
| K Gnanaprakash | 605006,Office of the Chief of Government Pharmacy, Indira Nagar, Gorimedu, Opp To Esi Hospital, Puducherry. | Pondicherry | Puducherry | 605006 | - | - | - |
| Albert Salomon | 609602,Pharmacy Stores, Office of the Medical Superintendent, Government General Hospital, No.63, Kamaraj Street, Karaikal. | Karaikal | Puducherry | 609602 | - | - | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
GEM_GENERAL_TERMS_AND_CONDITIONS
Extended Deadline
01-Sep-2026, 10:30 am
Opening Date
01-Sep-2026, 11:00 am
Extended Deadline
28-Aug-2026, 10:30 am
Opening Date
28-Aug-2026, 11:00 am
Discover companies most likely to bid on this tender
| Category | Specification | Requirement |
|---|---|---|
| GENERAL FEATURES | Product Description | ECG Machine |
| PRODUCT INFORMATION | Operating modes of ECG Machine | Automatic, Manual and Rhythm |
| PRODUCT INFORMATION | ECG machine should have ECG lead annotation facility | Yes |
| PRODUCT INFORMATION | Leads which is in ECG machine should be able to acquire simultaneously and interpret them | 12 |
| PRODUCT INFORMATION | Number of channels | 12 Channel |
| PRODUCT INFORMATION | ECG machine should acquire lead ECG for both adult and pediatric patients | Yes |
| PRODUCT INFORMATION | The ECG machine should have facility to show lead fail indication | Yes |
| PRODUCT INFORMATION | The ECG machine should have facility to show lead reversal indication | Yes |
| PRODUCT INFORMATION | The ECG machine should have facility to show the impedance to quality check of connection | Yes |
| PRODUCT INFORMATION | Acquisition time for ECG Machine in sec | 10 sec |
| PRODUCT INFORMATION | Digital sampling rate for Pacemaker spike detection | 8000 s/sec/channel |
| PRODUCT INFORMATION | Recording of digital sampling for pacemaker | 1000 s/sec/channel |
| PRODUCT INFORMATION | ECG machine should have real time colour backlit display of ECG waveforms with signal qualify indication for each lead | Yes |
| PRODUCT INFORMATION | ECG machine should have frequency filters | Artifact, AC and low and high pass frequency filters |
| PRODUCT INFORMATION | Number of ECGs which can be store in ECG Machine | 151 to 200 |
| PRODUCT INFORMATION | ECG machine should have full screen preview of ECG report for quality assessment checks prior to print | Yes |
| PRODUCT INFORMATION | Type of inbuilt screen | LCD, LED |
| PRODUCT INFORMATION | Size of screen in inches | 7 |
| PRODUCT INFORMATION | Display resolution of ECG machine in pixels | 640 x 480 |
| PRODUCT INFORMATION | ECG machine should have interpretation facility of the amplitudes, duration and morphologies of ECG waveforms and associated rhythm for adult and pediatric patient | Yes |
| PRODUCT INFORMATION | ECG machine should have alphanumeric keyboard for patient data entry | Hard keys |
| PRODUCT INFORMATION | Availability of latest interpretation software | Yes |
| PRODUCT INFORMATION | Wireless acquisition module with RF technology | No |
| PRODUCT INFORMATION | System should have the dedicated software to download the ECG form machine in PDF format | Yes |
| PRINTER | Printer Type | Thermal Printer |
| PRINTER | Recorder Paper Size | A4 Size |
| PRINTER | Recorder Speed | 50 mm/sec |
| PRINTER | Resolution of digital array printer | 200 dpi x 500 dpi |
| PRINTER | Number of Thermal paper | 500, 1000, 0 Or higher |
| PRINTER | ECG machine report format | Report formats of 3x4, 6x2, Rhythm for up to selected leads, 12 lead extended measurement, 1 minute of continuous waveform data for 1 selected lead |
| BATTERY | Provision of Battery | Yes |
| BATTERY | Battery Type | Built in Rechargeable Battery |
| BATTERY | Battery capacity of continuous rhythm recording on single charge (minutes) | 60 |
| BATTERY | Battery capacity | 50 ECG or 1 hour of continous rhythm recording on single charge |
| BATTERY | Connectivity to ECG Machine | LAN |
| BATTERY | The individual patient lead should be change without replacing the whole patient cable assembly | Yes |
| ENVIRONNMENTS PARAMETERS | Operating temperature and humidity | Temperature of 10 to 40 degree Celsius and Relative Humidity of 15 to 90% |
| POWER SUPPLY | Power input | 220-240 V AC, 50 Hz fitted with Indian plug |
| ACCESSORIES | ECG Machine 12 leads with interpretation | 1 |
| ACCESSORIES | Patient Cable | Not provided |
| ACCESSORIES | Chest Electrodes Adult (set of Six) | 1 |
| ACCESSORIES | Chest Electrodes Paediatric (set of Six) | Not provided |
| ACCESSORIES | Limb Electrodes | Not provided |
| ACCESSORIES | Power cable for charging | Yes |
| ACCESSORIES | Supplied with Clip electrode | Yes |
| ACCESSORIES | Compatible trolley provided | Yes |
| CERTIFICATION & REPORTS | Compliance to Medical Device Rules (MDR) 2017 as amended till date | Yes |
| CERTIFICATION & REPORTS | Availability of valid Medical Device license from competent authority | Yes |
| CERTIFICATION & REPORTS | Manufacturing unit certification | ISO:13485 (Latest) |
| CERTIFICATION & REPORTS | Availability of Test Report for each supplied batch/product as per Medical Device Rules (MDR) 2017 as amended till date | Yes |
| CERTIFICATION & REPORTS | Compliance to Safety Standards | IS 13450:Part 2:Sec 25 / IEC 60601-2-25 |
| CERTIFICATION & REPORTS | Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission and/or along with supplies as per buyer requirement | Yes |
| WARRANTY | Warranty (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) | 5 Or higher |
| Miscellaneous Parameters | User/Technical/Maintenance manuals to be supplied in English in hard and soft copy | Yes |
| Miscellaneous Parameters | The Principal Manufacturer must have direct Presence/approved service center In India | Yes |
| Miscellaneous Parameters | Installation and demonstration of equipment and training to be provided after completing supplies before acceptance | Yes |