Progress
Kerala Livestock Development Board (kldb) announces a tender for Electronic Milk Analyser with adulteration testing equipment (Q3) in ERNAKULAM, KERALA. Quantity: 1. Submission Deadline: 14-09-2026 16: 00: 00. Last date to apply is approaching fast!
Quantity
1
Bid Type
Two Packet Bid
Bid Validity
60 (Days)
Bid Type
Service
Evaluation
Total value wise evaluation
Inspection Required
No
Tech Clarification Time
2 Days
EMD Required
Yes
MII Purchase Preference
No
MSE Purchase Preference
No
MSE Exemption/Relaxation
No
Startup Exemption/Relaxation
No
Bid Splitting Applied
No
Min. Avg. Annual Turnover
2
OEM Avg. Turnover
22
Past Performance
30 %
Experience Required
2 Year (s)
Arbitration Clause
No
Mediation Clause
No
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
No
Item Category
Electronic Milk Analyser with adulteration testing equipment (Q3)
Advisory Bank
State Bank of India
ePBG Percentage
5%
ePBG Duration (Months)
14
Payment Timelines
Payments shall be made to the Seller within 30 days of issue of consignee receipt-cum-acceptance certificate (CRAC) and on-line submission of bills (This is in supersession of 10 days time as provided in clause 12 of GeM GTC)
Max Delivery Days
30
Delivery Locations
1
Delivery Cities
Ernakulam
Delivery Pincodes
686669
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Archana Ayyappan | 686669,Mudavoor P O Muvattupuzha Ernakulam Dist Kerala | Ernakulam | Kerala | 686669 | 1 | 30 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
GEM_GENERAL_TERMS_AND_CONDITIONS
Extended Deadline
14-Sep-2026, 10:30 am
Opening Date
15-Sep-2026, 10:30 am
Discover companies most likely to bid on this tender
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover
Additional Doc 1 (Requested in ATC) *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
| Category | Specification | Requirement |
|---|---|---|
| Construction | Body Material | Polycarbonate |
| Functional Parameters | Sample Volume, ml | =<5 |
| Functional Parameters | Minimum temperature of Sample raw milk | 5.0 |
| Functional Parameters | Maximum temperature of Sample raw milk | 40.0 |
| Functional Parameters | Measured Parameters | Fat, SNF, Protein |
| Performance Parameters | Detection Limit for Hydrogen Peroxide | NA |
| Performance Parameters | Detection Limit for Ammonium Sulphate | =<0.1% |
| Performance Parameters | Accuracy of Density Measurement | NA |
| Performance Parameters | Maximum Measuring limit for Added Water | 70%-75% |
| Performance Parameters | Minimum Measuring limit for Added Water (%) (Hint- Mention minimum detection limit) | 20%-25% |
| Performance Parameters | Accuracy of Lactose Measurement | NA |
| Performance Parameters | Accuracy of Protein Measurement | 0.05%-0.1% |
| Performance Parameters | Maximum Measuring limit for Protein | 8% |
| Performance Parameters | Maximum Measuring limit for Density(Kg/m3) | NA |
| Performance Parameters | Detection Limit for Added Water | >=20% |
| Performance Parameters | Detection Limit for salt(Sodium Chloride) | NA |
| Performance Parameters | Detection Limit for Detergent | NA |
| Performance Parameters | Minimum Measuring limit for Lactose | NA |
| Performance Parameters | Minimum Measuring limit for Protein | =<1% |
| Performance Parameters | Maximum Measuring limit for SNF | 10%-15% |
| Performance Parameters | Detection Limit for Sucrose | >=0.4% |
| Performance Parameters | Detection Limit for Sodium Bi-Carbonate | NA |
| Performance Parameters | Detection Limit for Urea | >=0.1% |
| Performance Parameters | Maximum Measuring limit for Lactose | NA |
| Performance Parameters | Accuracy of SNF Measurement | 0.1%-0.15% |
| Performance Parameters | Maximum Measuring limit for Fat | 12%-15% |
| Performance Parameters | Accuracy of Added Water Measurement | NA |
| Performance Parameters | Minimum Measuring limit for Density(Kg/m3) | NA |
| Performance Parameters | Minimum Measuring limit for SNF | =<1% |
| Performance Parameters | Accuracy of Fat Measurement | 0.10%-0.15% |
| Performance Parameters | Minimum Measuring limit for Fat | =<0.01% |
| Performance Parameters | Detection Limit for Maltodextrin | >=0.6% |
| Performance Parameters | Detection Limit for Caustic Soda | NA |
| Performance Parameters | Detection limit of melamine | NA |
| Generic Parameters | Count of Samples for which reagents,consumables,spares & acccessories are provided with Instrument | 10000 |
| Generic Parameters | Supplier capable of supply of samples,reagent,consumables & accessories as & when required | Yes |
| Generic Parameters | Built In Diagnostics to ensure accurate & error free operations | Yes |
| Generic Parameters | DC Power Supply | 12V |
| Generic Parameters | Rated Power, Watt | 20-30 |
| Generic Parameters | Communication Interface data transfer modes | USB, RS-232, SD Card, Cloud, Wifi |
| Generic Parameters | Display Type | LCD, LCD Graphical Colur TFT Display(Touch Screen) |
| Generic Parameters | POS Printer Provided with Instrument | No |
| Generic Parameters | PC Based software provided for Data analysis | Yes |
| Generic Parameters | Caliberation feature Available | Reference Sample |
| Generic Parameters | User Manual Provided | SoftCopy, Hard Copy |
| Installation & training | No. of days of training program | 2 |
| Installation & training | No of people to be trained in training program | 10-15 |
| Installation & training | No of training sessions executed | 10 |
| Warranty & services | No. of On Site Preventive Maintainance visit during warranty Period | 2 |
| Warranty & services | Warranty Extension in case of Breakdown for more than 30 days due to Manufacturing Defects/Faulty Spares | Yes |
| Warranty & services | Warranty from date of installation(System & all other parts) in years | 1.0 |