Contract Bidding Open For Gas Chromatography Mass Spectrometry (GC - MS) in NAGPUR, MAHARASHTRA

Bid Publish Date

12-Dec-2025, 3:46 pm

Bid End Date

25-Dec-2025, 1:00 pm

EMD

₹2,00,000

Progress

Issue12-Dec-2025, 3:46 pm
AwardPending
Explore all 4 tabs to view complete tender details

Quantity

1

Bid Type

Two Packet Bid

Categories 10

A tender has been published for National Forensic Sciences University Gas Chromatography Mass Spectrometry (GC - MS) (Q3) in NAGPUR, MAHARASHTRA. Quantity: 1 by. Submission Deadline: 25-12-2025 13: 00: 00. Check eligibility and apply.

Documents 5

GeM-Bidding-8694002.pdf

Main Document

Other Documents

OTHER

CATALOG-Specification-1

CATALOG Specification

Buyer uploaded ATC document

ATC

GEM General Terms and Conditions Document

GEM_GENERAL_TERMS_AND_CONDITIONS

Past Similar Tenders (Historical Results)

5 found

Gas Chromatography Mass Spectrometry (GC - MS)

State Forensic Science Laboratory Chhattisgarh

DURG, CHHATTISGARH

Posted: 28 July 2025
Closed: 7 August 2025
GEM

Gas Chromatography Mass Spectrometry (GC - MS)

Bihar State Police - A Voyage Of Excellence

PATNA, BIHAR

Posted: 30 July 2025
Closed: 20 August 2025
GEM

Gas Chromatography Mass Spectrometry (GC - MS)

Gauhati University

KAMRUP, ASSAM

Posted: 24 December 2024
Closed: 14 January 2025
GEM

Gas Chromatography Mass Spectrometry (GC - MS),Microscopes - Pathological and Research as per IS 43

Chhatrapati Shahu Ji Maharaj University Kanpur

KANPUR NAGAR, UTTAR PRADESH

Posted: 25 February 2025
Closed: 17 March 2025
GEM

Gas Chromatography Mass Spectrometry (GC-MS)

Sss National Institute Of Bio Energy

KAPURTHALA, PUNJAB

Posted: 13 January 2025
Closed: 10 February 2025
GEM

Technical Specifications 1 Item

Item #1 Details

View Catalog
Category Specification Requirement
GAS CHROMATOGRAPH (GC) CONFIGURATION "GC Detectors inclusive in the scope of supply Flame Ionization Detector (FID), Thermal Conductivity Detector (TCD or Hot Wire Detector), Electron Capture Detector (ECD), Nitrogen–Phosphorus Detector (NPD) / Flame Thermionic Detector (FTD), Flame Photometric Detector (FPD), Photo Ionization Detector (PID), Pulsed Flame Photometric Detector (PFPD)
GAS CHROMATOGRAPH (GC) CONFIGURATION "Number of FID in the System - inclusive in the scope of Supply 0, 1, 2
GAS CHROMATOGRAPH (GC) CONFIGURATION "Number of TCD in the System - inclusive in the scope of Supply 0, 1, 2
GAS CHROMATOGRAPH (GC) CONFIGURATION "Number of ECD in the System - inclusive in the scope of Supply 0, 1, 2
GAS CHROMATOGRAPH (GC) CONFIGURATION "Number of NPD/FTD in the System - inclusive in the scope of Supply 0, 1, 2
GAS CHROMATOGRAPH (GC) CONFIGURATION "Number of FPD in the System - inclusive in the scope of Supply 0, 1, 2
GAS CHROMATOGRAPH (GC) CONFIGURATION "Number of PID in the System - inclusive in the scope of Supply 0, 1, 2
GAS CHROMATOGRAPH (GC) CONFIGURATION "Number of PFPD in the System - inclusive in the scope of Supply 0, 1, 2
GAS CHROMATOGRAPH (GC) CONFIGURATION Number of Inlets GC can Simultaneously Support 1, 2, 3
GAS CHROMATOGRAPH (GC) CONFIGURATION Number of Detectors (Ports) GC can Simultaneously Support 1, 2, 3, 4
GAS CHROMATOGRAPH (GC) CONFIGURATION Number of Capillary Columns provided in GC - inclusive in the scope of Supply 0, 1, 2
GAS CHROMATOGRAPH (GC) CONFIGURATION Type of Capillary Column Polar, Non-Polar, Semi-Polar
GAS CHROMATOGRAPH (GC) CONFIGURATION "GC Capillary Column Length (Hint:-Select '0' if Not Applicable)" 0, 30, 60, 100
GAS CHROMATOGRAPH (GC) CONFIGURATION GC Packed Column Length 0, 3, 6, 9, 12
PERFORMANCE GC Detector Minimum Detection Level (MDL) < 1.4 pg C/s for FID, < 1.8 pg C/s for FID, < 3.0 pg C/s for FID < 400 pg/ml for TCD, < = 1 microgram/ml for TCD, < = 10 microgram/ml for TCD, < 4.5 fg /sec for ECD, < 5.5 fg /sec for ECD, < 0.1pg N /sec for NPD/FTD, < 0.03 pg P /sec for NPD/FTD, < 45 fg P /sec for FPD, < 100 fg P /sec for FPD, <2.5 pg S /sec for FPD, <5 pg S /sec for FPD
COLUMN OVEN (GAS CHROMATOGRAPH - GC) Maximum Operating Temperature (GC Oven) upto 400 deg. C, upto 450 deg. C (2) Or higher
COLUMN OVEN (GAS CHROMATOGRAPH - GC) Number of Oven Ramps / Plateaus for Temperature programming 6/7 or more, 9/10 or more, 20/21 or more
PNEUMATIC CONTROL Number of Injection Ports 1, 2
PNEUMATIC CONTROL Maximum Injection Port heating Temperature upto 400 deg. C (1), upto 450 deg. C (2) Or higher
PNEUMATIC CONTROL Pressure Range upto100 psi (1), upto 120 psi (2) Or higher
PNEUMATIC CONTROL Pressure set point Resolution (±)(psi) 0.001, 0.01, 0.1 Or lower
PNEUMATIC CONTROL Type of Sample Injection Port Split, Splitless, split / Splitless, Purged packed Inlet
PNEUMATIC CONTROL Split Ratio 6000:1, 7000:1, 7500:1
AUTO INJECTOR / AUTO SAMPLER Compatible Liquid Auto Sampler - inclusive in the scope of supply Yes, No
AUTO INJECTOR / AUTO SAMPLER Capacity of the Liquid Auto Sampler (Number of Vials) 0, 100, 150 Or higher
AUTO INJECTOR / AUTO SAMPLER Number of Injection Ports in the Liquid Auto Sampler 0, 1, 2
AUTO INJECTOR / AUTO SAMPLER Number of Channels in the Liquid Auto Sampler 0, 2
AUTO INJECTOR / AUTO SAMPLER Liquid Auto Sampler Features Fully automated Liquid Auto Sampler, Injects with variable speed & varying syringe sizes, Operates through Software (Software inclusive in the scope of supply), NA
HEAD SPACE SAMPLER Compatible Head Space Sampler- inclusive in the scope of supply Yes, No
HEAD SPACE SAMPLER Capacity of the Head Space Sampler Carousel (number of Vials) 12 or better, NA
HEAD SPACE SAMPLER Head Space Sampler Temperature Range Upto 200 deg. C or better, NA
PYROLYZER SYSTEM Compatible Pyrolyzer System for pyrolysis of solid and liquid samples- inclusive in the scope of supply Yes, No
MASS SPECTROPHOTOMETER (MS) SYSTEM Ionization modes Electron Ionization (EI), Chemical Ionization (CI), EI/PCI/NCI
MASS SPECTROPHOTOMETER (MS) SYSTEM Detector Type Electron Multiplier, Photo Multiplier, Triple-Axis, HED-EM
MASS SPECTROPHOTOMETER (MS) SYSTEM Operating modes Full Scan Mode, Selective Ion Monitoring (SIM) Mode, Both Full Scan & SIM Modes
PERFORMANCE OF THE MASS SPECTROPHOTOMETER (MS) SYSTEM Sensitivity (S/N Ratio) in full scan 1500:1 with 1 pg/microliter OFN, 800:1 (EI) or better OFN, 700:1 (EI) or better OFN, 250:1 (EI) or better OFN, 300:1 (PCI) or better BZP, 2000:1 (NCI) or better OFN
SALIENT FEATURES OF GC-MS "Salient Featutes of GC-MS (Select applicable features Only)" Automatic leak testing/detection of Pneumatics, Direct setting of split flow rates and ratios, Flow sensor for control & storage of split ratio, Split and Septum Purge gases, Able to direct inject onto packed and wide-bore capillary columns, Oven power turns-off automatically when the lid/door is opened, Over Heat Protection, Display Unit with built-in Key Board, Controlled by External PC System, Programmable ion source temperature, System includes variety of auto tune algorithms to tune the instrument for maximum sensitivity or for specific target compounds, EI source is inert to active compounds and programmable, EI Source have a dual filament design with automatic selection of other filament if one fails during analysis;, NA
LIBERARIES Libraries for GC-MS NIST Library for MS - Licensed version - 2017 or latest, Wiley Library, NA
SOFTWARE "Software Features (Select Applicable Features only)" Compatible with Windows 7 or higher version operating system, Seamless integration & control of all the GC parameters, Interface with Multiple GCs through the comprehensive real-time monitoring and diagnostic software, Real time notification via advisories and indicators, Counters, Chromatographic attributes, Instrument diagnostics, Access to maintenance information, Access to maintenance, run and event logs, Retention Time - Locking module for analyzing target compounds in complex matrices
DESKTOP WORKSTATION Desktop workstation Suitable to GC-MS - inclusive in the scope of supply Yes, No
DESKTOP WORKSTATION Graphics Card Memory Size 0, 1, 2, 4
ONLINE UPS UPS Provided in the Scope of Supply Yes, No
ONLINE UPS Rating of online UPS(KVA) 5.0 KVA, 7.5 KVA, 10 KVA, NA
ONLINE UPS Backup time of online UPS 0, 60, 90, 120
GENERIC Built-in Display Yes, No
GENERIC Connectivity interface USB, WIFI, ETHERNET, Bluetooth, RS 232, RJ-45
GENERIC Provision to split the sample from single injection to two different detectors and get results for two detectors simultaneously Yes, No
GENERIC Back Flash - inclusive in the scope of supply Post Column Back Flash, Mid Column Back Flash, NA
ACESSORIES Gas Supply inclusive in the scope of supply Nitrogen, Hydrogen, Argon, Helium, Zero Air, Methane, Isobutane, Not offered
ACESSORIES Number of Cyliners of each type supplied - inclusive in the scope of supply 0, 1, 2
ACESSORIES Gas cylinders certified by PESO Yes, NA
ACESSORIES Regulator - inclusive in the scope of supply Dual Stage Regulator, NA
ACESSORIES "Other Accessories inclusive in the scope of supply (Select only those Accessories which are part of supply)" Gas purification panel for all Gases with moisture trap, Hydrocarbon trap & Oxygen trap, Syringe for manual injection x 05 nos, Auto Sampler Vials x 1000 vials with screw cap, Graphite ferrules x 10 nos, Vessel ferrules x 10 nos, Septa for injector x 100 nos, Inlet liner for splitles x 4 nos, Inlet liner for split x 4 nos, O-rings x 10 nos. of each type, Column Cutter x 2 nos, Gas tube Cutter x 2 nos, Tool Kit x 1 no., User / Technical / Maintenance Manual, Standard Samples - NIST / IPL Traceable, Direct Injection Port, Splitter, Multishot Pyrolyser, Singleshot Pyrolyser, Syringe for Auto Sampler injection x 10 nos, EI filaments x 5 nos, NA
ENVIRONMENTAL Operating Humidity (RH) (%) at 40 degree C Up to 95
INSTALLATION & COMMISSIONING / WARRANTY / AVAILABILITY OF SPARES / TRAINING Warranty on equipment 1, 2, 3, 5
INSTALLATION & COMMISSIONING / WARRANTY / AVAILABILITY OF SPARES / TRAINING Number of days onsite training provided 3, 5, 7, 10
CERTIFICATION / TEST REPORT "Availability of UL/CE Certificate as per EN 61010- 1:2010 Safety requirements for electrical equipment for measurement,control, and laboratory use General requirements" Yes, No
CERTIFICATION / TEST REPORT Conformance to EMC/EMI as per EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use Yes, No
CERTIFICATION / TEST REPORT Availability of Test Reports from Central Govt / NABL approved / ILAC accredited Lab to prove conformity to the specification Yes, No

🤖 AI-Powered Bidder Prediction

Discover companies most likely to bid on this tender

Live AI
Historical Data

Required Documents

1

Experience Criteria

2

Past Performance

3

Bidder Turnover

4

Certificate (Requested in ATC)

5

OEM Authorization Certificate

6

OEM Annual Turnover

7

Compliance of BoQ specification and supporting document *In case any bidder is seeking exemption from Experience / Turnover Criteria

8

the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer