Bid Publish Date
20-Jul-2026, 12:01 pm
Bid End Date
14-Sep-2026, 7:00 pm
Location
Progress
N/a has released a public tender for Echo Cardiography Scanning Machine (V2) (Q2) in CENTRAL DELHI, DELHI. Quantity: 1. Submission Deadline: 14-09-2026 19: 00: 00. Download documents and apply online.
Quantity
1
Bid Type
Two Packet Bid
Bid Validity
180 (Days)
Bid Type
Service
Evaluation
Total value wise evaluation
Inspection Required
No
RA Qualification Rule
H1-Highest Priced Bid Elimination
Tech Clarification Time
2 Days
EMD Required
Yes
MII Purchase Preference
No
MSE Purchase Preference
Yes
MSE Preference Band
L1+15%
MSE Exemption/Relaxation
Yes
Startup Exemption/Relaxation
Yes
Bid Splitting Applied
No
Min. Avg. Annual Turnover
110
OEM Avg. Turnover
230
Past Performance
50 %
Warranty Period
7 days
Experience Required
3 Year (s)
Arbitration Clause
No
Mediation Clause
No
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
Yes
Item Category
Echo Cardiography Scanning Machine (V2) (Q2)
Advisory Bank
Bank Of Baroda
ePBG Percentage
5%
ePBG Duration (Months)
30
Pre-Bid Date
06-08-2026 14:00:00
Pre-Bid Venue
Office of Dr. Kaushal Kumar Tyagi, Addl Medical Superintendent, Nursing Home, Old Building, Dr. Ram Manohar Lohia Hospital, New Delhi
Max Delivery Days
90
Delivery Locations
1
Delivery Cities
New Delhi
Delivery Pincodes
110001
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Harikesh Meena | 110001,dr. ram manohar lohia hospital baba kharak singh marg new delhi | New Delhi | Delhi | 110001 | 1 | 90 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
OTHER
CATALOG Specification
OTHER
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS
Extended Deadline
14-Sep-2026, 1:30 pm
Opening Date
15-Sep-2026, 1:30 pm
Discover companies most likely to bid on this tender
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover
Additional Doc 1 (Requested in ATC)
Compliance of BoQ specification and supporting document *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
| Category | Specification | Requirement |
|---|---|---|
| Types of Probes & Features | Cardiac and Vascular Probes Configuration | TTE Phased Array, TTE Matrix Array, TEE Phased Array, TEE Matrix Array, Linear Probe |
| Types of Probes & Features | TTE(Transthoracic) Probe Type | Adult, Pediatric, Neonatal |
| Types of Probes & Features | TEE Adult Probe biplane and Multiplane frequency, in MHz | 2 to 7.5 MHz (±2), Not Applicable |
| Types of Probes & Features | TEE Pediatric Probe Required | Yes, No |
| Types of Probes & Features | TEE Pediatric Probe frequency, in MHz | 3 to 8 MHz (±2), Not Applicable |
| Types of Probes & Features | Linear Probe for linear/ vascular applications required | Yes |
| Types of Probes & Features | Adult/Pediatric Linear probe frequency in MHz | 5 to 14 MHz (±2) |
| Types of Probes & Features | No of Adult TTE Probes Required | 1 |
| Types of Probes & Features | No of Pediatric TTE Probes Required | 1 |
| Types of Probes & Features | No of Neonatal TTE Probes Required | 1 |
| Types of Probes & Features | No of Adult TEE Probes Required | 1, NA |
| Types of Probes & Features | No of Pediatric TEE Probes Required | 1, NA |
| Types of Probes & Features | No of Pediatric 3D/4D TEE Probes Required | 1, NA |
| Types of Probes & Features | No of Neonatal TEE Probes Required | 1, NA |
| Types of Probes & Features | No of Adult/pediatric Linear Probes supplied | 1, NA |
| Types of Probes & Features | Dedicated 3D/4D volume imaging for cardiology TTE ad TEE | Yes, No |
| Types of Probes & Features | Live 3 D imaging in color doppler system | Yes |
| Types of Probes & Features | Colour Doppler system with all application packages ,Quad loop for serial studies with high frame rate revies, harmonic imaging capability in all modes and integrated stress echo package digital storage and retrieval | Yes, No |
| Types of Probes & Features | Provision of Tissue Colorization (B-Colour) for improved contrast resolution | Yes |
| Types of Probes & Features | Application software | Adult, Pediatric, Neonatal, Vascular applications and Transesophagal applications |
| Types of Probes & Features | Cine Loop Features | Yes |
| Types of Probes & Features | Frame grabber facility for post analysis | Yes |
| Types of Probes & Features | Facility of Various maps for pre and post processing | Yes |
| Types of Probes & Features | ECG trigger facility | Yes |
| Types of Probes & Features | User defined Preset | Yes |
| Types of Probes & Features | Dedicated integrated dynamic stress echo package for flexible user defined protocols with stacked sub loops facility and contrast stress protocol usable for stress echocardiography | Yes |
| Types of Probes & Features | Number of transducer/probe ports | 3 |
| Scan Modes and Performance Parameters | Cardiac 3-D freehand Option | Yes, No |
| Scan Modes and Performance Parameters | Cardiac 3D/4D automatic Option | Yes |
| Scan Modes and Performance Parameters | Contrast Harmonic/Agent Imaging with angio mode | Yes |
| Scan Modes and Performance Parameters | Color Doppler 3-D/4-D option | Yes |
| Scan Modes and Performance Parameters | Power Dopple 3-D/4-D option | Yes |
| Scan Modes and Performance Parameters | Contrast-enhanced ultrasound (CEUS) option | Yes, No |
| Scan Modes and Performance Parameters | Tissue synchronization, tissue velocity | Yes |
| Scan Modes and Performance Parameters | Cardiac Shear wave elastography | Yes, No |
| Scan Modes and Performance Parameters | Cardiac compression elastography | Yes, No |
| Scan Modes and Performance Parameters | Provision for higher sensitivity for low frequency doppler in all probes | Yes |
| Scan Modes and Performance Parameters | Latest generation Electronic Phased/sector array Colour Doppler system | Yes |
| Scan Modes and Performance Parameters | Number of Channels in color doppler system | 500000 or more (if channels are being designated as digital channels) |
| Scan Modes and Performance Parameters | System should be capable of generating real time live 3- D images | Yes |
| Image display and processing | DICOM 3 Compliant | Yes |
| Image display and processing | Automated B-mode (2-D) image | Yes |
| Image display and processing | Automated CDI image optimization | Yes |
| Image display and processing | Automated PW Doppler image optimization | Yes |
| Image display and processing | Touch Screen | No |
| Image display and processing | Monitor display size, cm (in) | 21 Inch(±1) Or higher |
| Image display and processing | Split screen | Twin View and Live Quad Screen |
| Image display and processing | Single/dual monitors | Dual (image display on touch panel) |
| Image display and processing | Speckle-tracking strain and strain rate | Yes |
| Image display and processing | Analysis and calculation packages | 2-D wall motion tracking, 3-D wall motion tracking, Cardiac, Mitral valve analysis, User-programmable calculations, No |
| Image display and processing | Type of printer provided with the echo cardio machine | Themal Printer, B/W Laser Printer, Color Laser Printer |
| Image display and processing | Type of processor in PC | i7 or Latest |
| Image display and processing | Conectivity between Echo cardiogram and PC | Through IMS/Equivalent Software |
| Image display and processing | PC with all software inclusive interfaced with echocardiograph machine | Yes |
| Image display and processing | Inkjet printer provided with computer for reporting | No |
| Image display and processing | Glossy Colour Print Paper supplied | 500 sheets |
| Image display and processing | ECG Cable | 2 |
| Image display and processing | Echo cardiography sytem to be DICOM Ready | Yes |
| Image display and processing | Dicom Interfacability to HIS | Yes |
| Image display and processing | Dicom Interfacability Dicom to PACS | Yes |
| Image display and processing | Interfacability Dicom to RIS | Yes, No |