A tender has been published for N/a High End Multipara Monitor (V3) (Q2) in MUMBAI, MAHARASHTRA. Quantity: 4 by. Submission Deadline: 17-09-2026 17: 00: 00. Check eligibility and apply.
Quantity
4
Bid Type
Two Packet Bid
Bid Validity
180 (Days)
Bid Type
Service
Evaluation
Total value wise evaluation
Inspection Required
No
RA Qualification Rule
H1-Highest Priced Bid Elimination
Tech Clarification Time
3 Days
EMD Required
Yes
MII Purchase Preference
Yes
MII Preference Band
L1+20%
MSE Purchase Preference
Yes
MSE Preference Band
L1+15%
MSE Exemption/Relaxation
Yes
Startup Exemption/Relaxation
Yes
Bid Splitting Applied
No
Min. Avg. Annual Turnover
12
Past Performance
80 %
Warranty Period
2 years
Experience Required
3 Year (s)
Arbitration Clause
No
Mediation Clause
No
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
Yes
Item Category
High End Multipara Monitor (V3) (Q2)
Advisory Bank
State Bank of India
ePBG Percentage
5%
ePBG Duration (Months)
28
Max Delivery Days
45
Delivery Locations
1
Delivery Cities
Raigad
Delivery Pincodes
410210
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Kanchana Gopalkrishnan | 410210,Tata Memorial Center - Advanced Centre for Treatment, Research and Education in Cancer (ACTREC), Plt no.1 &2, Sector 22, Kharghar, Navi Mumbai | Raigad | Maharashtra | 410210 | 4 | 45 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
OTHER
CATALOG Specification
OTHER
GEM_GENERAL_TERMS_AND_CONDITIONS
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
Additional Doc 1 (Requested in ATC)
Compliance of BoQ specification and supporting document *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
| Category | Specification | Requirement |
|---|---|---|
| Product Specifications | Type of Monitor | Modular |
| Product Specifications | Monitoring Parameters | ECG, HR, Respiration Rate, SpO2, Pulse Rate, NIBP, Temperature, IBP |
| Product Specifications | Cardiac output | Invasive (Thermodilution) |
| Product Specifications | S T Analysis | Yes |
| Product Specifications | PPV (derived value during IBP measurement) | Yes |
| Product Specifications | Length of time of full disclosure of patient data in hr | ≥ 72 Or higher |
| Product Specifications | Length of time of trending in hr | ≥ 72 Or higher |
| ECG PARAMETERS | Number of leads of ECG | 5 |
| ECG PARAMETERS | 12-Lead interpretation for ECG | Yes |
| ECG PARAMETERS | ECG Waveform number of leads display | 7 |
| ECG PARAMETERS | Arrhythmia detect | Yes |
| ECG PARAMETERS | Heart rate measurement range | Atleast 30 -250 bpm |
| ECG PARAMETERS | Accuracy | ± 2 bpm or better |
| RESPIRATION PARAMETERS | Respiration Method | Impedance pneumography |
| RESPIRATION PARAMETERS | Display waveform of respiration | Yes |
| RESPIRATION PARAMETERS | Respiration rate measurement range | 0 - 150 bpm or more |
| PULSE OXIMETERY PARAMETERS | SPO2 measurement range | 40 - 100% |
| PULSE OXIMETERY PARAMETERS | Accuracy of SPO2 measurement | ± 2% or better |
| NIBP PARAMETERS | Systolic Blood pressure monitoring range | 30 - 300 mmHg or more |
| NIBP PARAMETERS | Diastolic Blood pressure monitoring range | 10 - 200 mmHg or more |
| NIBP PARAMETERS | Accuracy of NIBP Measurement | ±3 mm/Hg or better |
| TEMPERATURE PARAMETERS | Number of channels for temperature | 2 |
| TEMPERATURE PARAMETERS | Type of probe | Skin |
| TEMPERATURE PARAMETERS | Temperature probe type | Reusable |
| TEMPERATURE PARAMETERS | Temperature probe range | Atleast 30 - 40 degree |
| TEMPERATURE PARAMETERS | Minimum gradation of temperature probe | 0.1 degree C or better |
| TEMPERATURE PARAMETERS | Accuracy in % | ±1% or better |
| IBP PARAMETERS | Number of channels for IBP | 2 |
| ETCO2 PARAMETERS | Sensor type | Sidestream |
| ALARMS | Availability of Audio and visual alarms with high , medium and low intensity adjustment | Yes |
| ALARMS | Operator can set audio visual alarm levels for low or high levels of each parameter independently | Yes |
| ALARMS | Systolic BP - low and high | Yes |
| ALARMS | Diastolic BP - low and high | Yes |
| ALARMS | Mean BP - low and high | Yes |
| ALARMS | Heart rate - low and high | Yes |
| ALARMS | SPO2 - low and high | Yes |
| ALARMS | Respiratory rate - low and high | Yes |
| ALARMS | Temperature - low and high | Yes |
| ALARMS | Alarm Volume Adjustable | Yes |
| ALARMS | Activating silence of alarm | Yes |
| ALARMS | Duration of alarm silence ( in seconds ) | 120 |
| ALARMS | Suspension of all alarms | Yes |
| ALARMS | Latching of alarm | Yes |
| ALARMS | Cuff leakage alarm | Yes |
| ALARMS | Cuff disconnect alarm | Yes |
| ALARMS | Alarm for inflation/deflation errors | Yes |
| ALARMS | Indicator for low-battery | Yes |
| ALARMS | EtCO2 - low and high | Yes |
| DISPLAY PARAMETERS | Display Size in inches | 12, 15, 17, 19 Or higher |
| DISPLAY PARAMETERS | Traces of display | >10 |
| DISPLAY PARAMETERS | User Interface of display | Touchscreen, Touchscreen and Rotary knob Or higher |
| POWER REQUIREMENTS | Power supply | Single Phase 220-240 VAC, 50Hz |
| POWER REQUIREMENTS | Mains cable | 3 meter or more with Indian plug |
| POWER REQUIREMENTS | Minimum Inbuilt Battery back up ( in minutes ) | 120 |
| POWER REQUIREMENTS | Recharging time of battery ( in hr ) | ≤ 6 |
| CERTIFICATIONS | Availability of valid Medical Device license for the product issued from the competent authority defined under Drugs and Cosmetic Act 1940 and Rules made there under as amended till date | Yes |
| CERTIFICATIONS | Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission or along with supplies as per buyer requirement | Yes |