TenderDekho Logo
GEM

Open Tender for Holter Monitor (V2) in AIZAWL, MIZORAM

Bid Publish Date

23-Feb-2026, 7:47 pm

Bid End Date

06-Mar-2026, 8:00 pm

Location

AIZAWL , MIZORAM

Progress

Issue23-Feb-2026, 7:47 pm
AwardPending
Explore all 4 tabs to view complete tender details

Quantity

1

Bid Type

Single Packet Bid

Categories 3

Public procurement opportunity for Hospital Medical Education Holter Monitor (V2) (Q2) in AIZAWL, MIZORAM. Quantity: 1 issued by. Submission Deadline: 06-03-2026 20: 00: 00. View full details and respond.

Documents 3

GeM-Bidding-9033247.pdf

Main Document

CATALOG-Specification-1

CATALOG Specification

GEM General Terms and Conditions Document

GEM_GENERAL_TERMS_AND_CONDITIONS

Past Similar Tenders (Historical Results)

5 found

Holter with Monitor for analysis

Indian Navy

Posted: 6 December 2024
Closed: 28 December 2024
GEM

Holter with Monitor for analysis

N/a

Posted: 19 December 2024
Closed: 1 January 2025
GEM

Holter with Monitor for analysis

Indian Army

LUCKNOW, UTTAR PRADESH

Posted: 20 December 2024
Closed: 10 January 2025
GEM

Holter with Monitor for analysis

Directorate Of Purchase And Stores

Posted: 4 November 2024
Closed: 28 November 2024
GEM

Computer Monitor (V2)

Mecon Limited

BANGALORE, KARNATAKA

Posted: 12 November 2025
Closed: 24 November 2025
GEM

Technical Specifications 1 Item

Item #1 Details

View Catalog
Category Specification Requirement
GENERAL Product Name Holter Monitor
PRODUCT INFORMATION Type Worn in a case on a belt
PRODUCT INFORMATION Number of Channels 3
PRODUCT INFORMATION Number of Leads 5
PRODUCT INFORMATION Analyzing Various Arrhythmias Ventricular Ectopics, Supraventricular Ectopics, Ventricular Tachycardia, Ventricular Fibrillation, Supraventricular Tachycardia, Atrial Fibrillation, Sinus Pause
PRODUCT INFORMATION Full Disclosure of ECG with Different Color Coding for VE, SVE and Pauses
PRODUCT INFORMATION Hourly Count Menu (Minimum) Two Scan
PRODUCT INFORMATION Minimum Templates Editing to Analyse VE, SVE, V-RUN,SV-RUN, PAUSE, ST, etc 20 Templates
PRODUCT INFORMATION Atrial Fibrillation/Flutter analysis Yes
PRODUCT INFORMATION Heart rate variability (HRV) analysis Yes
PRODUCT INFORMATION SAECG-Late potential analysis No
PRODUCT INFORMATION Pace Maker auto detection and analysis Yes
PRODUCT INFORMATION 12 Lead ST Scan Yes
PRODUCT INFORMATION QT analysis with validation Yes
PRODUCT INFORMATION To print ECG strips from any part of Holter recording Yes
PRODUCT INFORMATION T Wave altemans Yes
PRODUCT INFORMATION Facility for electronic compact technique for Flash card/USB Cable Yes
PRODUCT INFORMATION Should be able to edit and reclassify beats and arrhythmias Yes
PRODUCT INFORMATION Indication for Data Storage and Device Status LED
PRODUCT INFORMATION Sleep Apnea Analysis Yes
PRODUCT INFORMATION Real time ST analysis Yes
PRODUCT INFORMATION ST profile, ST elevation, ST depression, Heart rate, Mets, NIBP, ST index should be able to be clearly analyze Yes
PRODUCT INFORMATION System should have automatic arythmia detection Yes
PRODUCT INFORMATION System should present comprehensive final report on minute by minute record of ST segment trend Yes
PRODUCT INFORMATION System should have facility to do re-analysis of stored ECG by changing line measurement point title ST and J Yes
PRODUCT INFORMATION Real time average complexes Yes
PRODUCT INFORMATION Standard Leads Cables Set provided Yes
PRODUCT INFORMATION ST segment analysis should be available for all selected channels Yes
RECORDER CONFIGURATION Recorder should be minimum 3 channel ECG for recording data duration 72 Hrs
RECORDER CONFIGURATION Digital Recorder with removable storage media Yes
RECORDER CONFIGURATION Memory storage capacity of Card 32GB
RECORDER CONFIGURATION Number of Hours Store ECG into Flash Memory Card/Inbuilt Memory 24 or 48 Hours
RECORDER CONFIGURATION Saved data should be accessed using an integrated reading device Yes
RECORDER CONFIGURATION Compatibility with Alkaline Battery Yes
RECORDER CONFIGURATION Pacemaker spike detection Minimum 512 samples/sec/ Channel
RECORDER CONFIGURATION Recording and Storage Minimum 128 samples/sec/Channel
RECORDER CONFIGURATION Venticular Late Potential (VLP) Minimum 1000 samples/sec/Channel
DATA MANAGEMENT SYSTEM Computer system with latest licensed operating system and antivirus provided Yes
DATA MANAGEMENT SYSTEM PC monitor type LCD/LED
DATA MANAGEMENT SYSTEM Software Holter analysis, Trends, Strips, Page, QT Analysis, Shape, Event Review, Episode Review , Final Report, TWA Alterns Analysis
DATA MANAGEMENT SYSTEM Minimum size for color PC monitor (inches) 17 inch or more
DATA MANAGEMENT SYSTEM Printer Ink-Jet, Colour Laser Printer, Black and White Laser Printer
DATA MANAGEMENT SYSTEM UPS with power backup (Minutes) 30, 60, NA if not provided
REPORT FORMAT Format for reporting Patient information Indications, medications, symptoms, time, diagnosis, notes and conclusions with space for personnel and physician electronic signature.
REPORT FORMAT Format for Examination summery Quality of recording, baseline rhythm, min and max heart rate, VE, SVE, VT, SVT, Blocks and Pauses
REPORT FORMAT Report format Covering page, Arrhythmia analysis report, ST segment analysis, Automatically and manually selected ECG strips
WARRANTY Warranty in Years (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) 1, 2, 3, 4, 5
CERTIFICATIONS Compliance to Medical Device Rules (MDR) 2017 as amended till date YES
CERTIFICATIONS Availability of valid drug license for the product issued from the competent authority defined under Drugs and Cosmetic Act 1940 and Rules made there under as amended till date Yes
CERTIFICATIONS Manufacturing unit certification ISO:13485 (Latest)
CERTIFICATIONS Availability of Test Report for product as per Medical Device Rules (MDR) 2017 as amended till date Yes
CERTIFICATIONS Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission and/or along with supplies as per buyer requirement Yes

🤖 AI-Powered Bidder Prediction

Discover companies most likely to bid on this tender

Live AI
Historical Data

Required Documents

1

Experience Criteria

2

Past Performance

3

Bidder Turnover

4

Certificate (Requested in ATC)

5

OEM Authorization Certificate

6

OEM Annual Turnover

7

Additional Doc 1 (Requested in ATC)

8

Compliance of BoQ specification and supporting document *In case any bidder is seeking exemption from Experience / Turnover Criteria

9

the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer

Similar Tenders

5 found
Urgent

Computer Monitor (V2)

Intelligence Bureau

📍 PATNA, BIHAR

⏰ Deadline: 1 day left
🛒 Type: Goods
View GEM

Computer Monitor (V2)

Director Of Public Instructions

📍 THIRUVANANTHAPURAM, KERALA

⏰ Deadline: 2 weeks left
🛒 Type: Goods
View GEM
Urgent

Computer Monitor (V2)

Bharat Electronics Limited (bel)

📍 GHAZIABAD, UTTAR PRADESH

⏰ Deadline: 5 days left
🛒 Type: Goods
View GEM

Computer Monitor (V2)

N/a

📍 NORTH 24 PARGANAS, WEST BENGAL

⏰ Deadline: 2 weeks left
🛒 Type: Goods
View GEM
Urgent

Computer Monitor (V2)

N/a

📍 CHANDIGARH, CHANDIGARH

⏰ Deadline: 7 days left
🛒 Type: Goods
View GEM