Telangana State Dairy Development Cooperative Fed invites bids for Automated HPLC System for Separation of complex mixtures (Q3) in HYDERABAD, TELANGANA. Quantity: 1. Submission Deadline: 22-08-2026 17: 00: 00. Submit your proposal before the deadline.
Quantity
1
Bid Type
Two Packet Bid
Bid Validity
30 (Days)
Bid Type
Service
Evaluation
Total value wise evaluation
Inspection Required
No
Tech Clarification Time
2 Days
EMD Required
Yes
MII Purchase Preference
No
MSE Purchase Preference
Yes
MSE Preference Band
L1+15%
MSE Exemption/Relaxation
No
Startup Exemption/Relaxation
No
Bid Splitting Applied
No
Min. Avg. Annual Turnover
20
OEM Avg. Turnover
20
Past Performance
80 %
Warranty Period
15 years
Experience Required
3 Year (s)
Arbitration Clause
No
Mediation Clause
No
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
No
Item Category
Automated HPLC System for Separation of complex mixtures (Q3)
Advisory Bank
State Bank of India
Payment Timelines
Payments shall be made to the Seller within 45 days of issue of consignee receipt-cum-acceptance certificate (CRAC) and on-line submission of bills (This is in supersession of 10 days time as provided in clause 12 of GeM GTC)
Max Delivery Days
15
Delivery Locations
1
Delivery Cities
Hyderabad
Delivery Pincodes
500017
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Kovuru Devendra Kumar | 500017,Vijaya Bhavan, Lalapet,Hyderabad | Hyderabad | Telangana | 500017 | 1 | 15 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
OTHER
GEM_GENERAL_TERMS_AND_CONDITIONS
Extended Deadline
22-Aug-2026, 11:30 am
Opening Date
22-Aug-2026, 12:00 pm
Extended Deadline
10-Aug-2026, 7:30 am
Opening Date
10-Aug-2026, 8:00 am
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover
Additional Doc 1 (Requested in ATC)
Additional Doc 2 (Requested in ATC)
Additional Doc 3 (Requested in ATC)
Additional Doc 4 (Requested in ATC)
Compliance of BoQ specification and supporting document *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
| Category | Specification | Requirement |
|---|---|---|
| Standard parameter | Instrument calibration standards | IEC 17025 |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Components of unit | Quaternary Low-Pressure Gradient pump& Parallel Double Plunger |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Number of solvents delivery | 4 |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Flow rate setting range in ml/min | 0.001 to 10 |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Flow rate accuracy | ± 1 % |
| UV-Vis Detector | UV-Vis Detector provided | Yes, No Or higher |
| UV-Vis Detector | Light source | Deuterium lamp |
| UV-Vis Detector | Spectrum Band width in nm | 4, 5, 6, 8, NA if not provided |
| UV-Vis Detector | Wavelength range in nm | 190 - 700 |
| UV-Vis Detector | Sampling Rate in Hz | 100 |
| UV-Vis Detector | The flow cell must be temperature Controlled | Yes |
| UV-Vis Detector | Wavelength accuracy in nm | ± 1 |
| UV-Vis Detector | Wavelength reproducibility in nm | ± 0.1, NA if not provided |
| UV-Vis Detector | Drift in AU/Hour | less than 1 x 10-4 |
| UV-Vis Detector | Noise level in AU | ±5 x 10-6 |
| UV-Vis Detector | Ability to monitor and quantitative 2 wavelengths simultaneously | Yes |
| UV-Vis Detector | Linearity in AU | 2.5 or more |
| UV-Vis Detector | Temperature control provided for Flow cell | Yes, NA if not provided |
| UV-Vis Detector | Temperature control range for flow cell in Degree Celsius | 5 degree above room temp to 50 degree |
| UV-Vis Detector | flow cell path length in mm | 10 |
| UV-Vis Detector | Flow cell volume in ?l | 10 - 15 |
| Auto Sample Injector with cooler | Injection Method | Total volume sample injection |
| Auto Sample Injector with cooler | Carry over of sample | 0.005% or lower |
| Auto Sample Injector with cooler | Injection precision in RSD | less than 0.50 % RSD |
| Software | Operating Windows | Win 7 or latest, Win 10 or latest |
| Photo Diode Array (PDA) Detector | Provision of Photo diode array | Yes, No Or higher |
| Photo Diode Array (PDA) Detector | PDA detector Wavelength range in nm | 190- 800 or better |
| Photo Diode Array (PDA) Detector | Number of Diodes | 1024 |
| Photo Diode Array (PDA) Detector | PDA detector light source | Deuterium lamp and Tungsten lamp |
| Photo Diode Array (PDA) Detector | PDA detector Wavelength accuracy in nm | ± 1 |
| Photo Diode Array (PDA) Detector | Data Rate in Hz | 100, NA if not provided, <100 |
| Photo Diode Array (PDA) Detector | Noise in AU | less than ± 0.6 x 10-5 or lower, NA if not provided, less than ± 1 x 10-5 or lower |
| Photo Diode Array (PDA) Detector | PDA detector Drift in AU/Hour | less than 1 x 10-3 |
| Photo Diode Array (PDA) Detector | Temperature controlled flow cell in C | 5 degree above room temp to 50 degree, NA if not provided |
| Photo Diode Array (PDA) Detector | Flow cell | Flow cell, NA if not provided |
| Warranty & Reports | Warranty (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) | 3, 4, 5 Or higher |
| Warranty & Reports | Number of installations in Central /State/PSU Govt Hospitals | <3, >3 |
| Warranty & Reports | Availability of toll free facility for technical support maintained by OEM or authorized agencies | Yes, No Or higher |