Echo Cardiography Scanning Machine (V2)
Jawaharlal Institute Of Postgraduate Medical Education And Research (jipmer)
PONDICHERRY, PUDUCHERRY
Progress
Quantity
4
Bid Type
Two Packet Bid
Organization: Hll Lifecare Limited, under the Department of Health And Family Welfare, issues a procurement for an Echo Cardiography Scanning Machine (V2) (Q2). Estimated value not disclosed; EMD: ₹280,000. Location: India. The scope includes an advanced, prosthetically capable cardiac ultrasound platform with 2D/3D imaging, full Doppler suite, TEE/TEE pediatric probes, and an extensive range of adult/pediatric linear probes. Key differentiators include real-time 3D imaging, dynamic stress echo, comprehensive DICOM integration, and broad probe-port capability. This tender covers high-end cardiovascular imaging requirements and extensive post-processing features, with a focus on turnkey integration and software-enabled performance.
Echo system with 2D, M-Mode, PW, TDI, CW, Power Doppler, directional angio; TEE probe support
Minimum 8,00,000 digital processing channels
Dynamic range ≥ 200 dB
Display: 21 inch (or larger) high-resolution; 12 inch+ touch panel
Four universal probe ports with electronic switching
Live 3D imaging; 3D/4D volume imaging for TTE/TEE; contrast imaging; color Doppler 3D/4D
DICOM Ready; interfacing to HIS/PACS/RIS; automated B-mode, CDI, PW Doppler optimization
EMD amount and payment terms to be clearly stated in the bid
Delivery term variants up to 50% quantity change; extended delivery details
Mandatory OEM authorizations and product brochures with technical compliance
Detailed government payment terms to be followed; ensure mobilization advance if allowed and milestone-based payments with acceptance criteria.
Option clause permits quantity variation up to 50%; delivery period linked to last delivery order date; extended time calculated by formula.
LD terms to be specified in final contract; ensure penalties for late delivery and non-compliance with technical specs.
Experience in supplying high-end echocardiography systems with 3D/4D capabilities
Evidence of OEM authorization for critical components
Financial statements demonstrating stability and ability to support large procurement
Main Document
BOQ
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS
Jawaharlal Institute Of Postgraduate Medical Education And Research (jipmer)
PONDICHERRY, PUDUCHERRY
All India Institute Of Medical Sciences (aiims)
Agartala Government Medical College
WEST TRIPURA, TRIPURA
Directorate Of Medical Education And Training
MIRZAPUR, UTTAR PRADESH
Directorate Of Medical Education And Training
Tender Results
Loading results...
| Category | Specification | Requirement |
|---|---|---|
| Types of Probes & Features | Cardiac and Vascular Probes Configuration | TTE Phased Array, TTE Matrix Array, TEE Phased Array, TEE Matrix Array, Linear Probe |
| Types of Probes & Features | TTE(Transthoracic) Probe Type | Adult, Pediatric, Neonatal |
| Types of Probes & Features | TEE Adult Probe biplane and Multiplane frequency, in MHz | 2 to 7.5 MHz (±2), Not Applicable |
| Types of Probes & Features | TEE Pediatric Probe Required | Yes, No |
| Types of Probes & Features | TEE Pediatric Probe frequency, in MHz | 3 to 8 MHz (±2), Not Applicable |
| Types of Probes & Features | Linear Probe for linear/ vascular applications required | Yes, No |
| Types of Probes & Features | Adult/Pediatric Linear probe frequency in MHz | 5 to 14 MHz (±2), 5 to 10MHz (±2), Not Applicable |
| Types of Probes & Features | No of Adult TTE Probes Required | 1, NA |
| Types of Probes & Features | No of Pediatric TTE Probes Required | 1, NA |
| Types of Probes & Features | No of Neonatal TTE Probes Required | 1, NA |
| Types of Probes & Features | No of Adult TEE Probes Required | 1, NA |
| Types of Probes & Features | No of Pediatric TEE Probes Required | 1, NA |
| Types of Probes & Features | No of Pediatric 3D/4D TEE Probes Required | 1, NA |
| Types of Probes & Features | No of Neonatal TEE Probes Required | 1, NA |
| Types of Probes & Features | No of Adult/pediatric Linear Probes supplied | 1, NA |
| Types of Probes & Features | Dedicated 3D/4D volume imaging for cardiology TTE ad TEE | Yes, No |
| Types of Probes & Features | Live 3 D imaging in color doppler system | Yes, No |
| Types of Probes & Features | Colour Doppler system with all application packages ,Quad loop for serial studies with high frame rate revies, harmonic imaging capability in all modes and integrated stress echo package digital storage and retrieval | Yes, No |
| Types of Probes & Features | Provision of Tissue Colorization (B-Colour) for improved contrast resolution | Yes, No |
| Types of Probes & Features | Application software | Adult, Pediatric, Neonatal, Vascular applications and Transesophagal applications |
| Types of Probes & Features | Cine Loop Features | Yes, No |
| Types of Probes & Features | Frame grabber facility for post analysis | Yes, No |
| Types of Probes & Features | Facility of Various maps for pre and post processing | Yes, No |
| Types of Probes & Features | ECG trigger facility | Yes, No |
| Types of Probes & Features | User defined Preset | Yes, No |
| Types of Probes & Features | Dedicated integrated dynamic stress echo package for flexible user defined protocols with stacked sub loops facility and contrast stress protocol usable for stress echocardiography | Yes, No |
| Types of Probes & Features | Number of transducer/probe ports | 1, 2, 3, 4 |
| Scan Modes and Performance Parameters | Cardiac 3-D freehand Option | Yes, No |
| Scan Modes and Performance Parameters | Cardiac 3D/4D automatic Option | Yes, No |
| Scan Modes and Performance Parameters | Contrast Harmonic/Agent Imaging with angio mode | Yes, No |
| Scan Modes and Performance Parameters | Color Doppler 3-D/4-D option | Yes, No |
| Scan Modes and Performance Parameters | Power Dopple 3-D/4-D option | Yes, No |
| Scan Modes and Performance Parameters | Contrast-enhanced ultrasound (CEUS) option | Yes, No |
| Scan Modes and Performance Parameters | Tissue synchronization, tissue velocity | Yes, No |
| Scan Modes and Performance Parameters | Cardiac Shear wave elastography | Yes, No |
| Scan Modes and Performance Parameters | Cardiac compression elastography | Yes, No |
| Scan Modes and Performance Parameters | Provision for higher sensitivity for low frequency doppler in all probes | Yes, No |
| Scan Modes and Performance Parameters | Latest generation Electronic Phased/sector array Colour Doppler system | Yes |
| Scan Modes and Performance Parameters | Number of Channels in color doppler system | 50000 or more (if channels are being designated as digital channels), 500000 or more (if channels are being designated as digital channels), 5000000 or more (if channels are being designated as digital channels), 128 or more (if channels are being designated as physical channels) |
| Scan Modes and Performance Parameters | System should be capable of generating real time live 3- D images | Yes, No, Yes with MV No Analysis |
| Image display and processing | DICOM 3 Compliant | Yes, No |
| Image display and processing | Automated B-mode (2-D) image | Yes, No |
| Image display and processing | Automated CDI image optimization | Yes, No |
| Image display and processing | Automated PW Doppler image optimization | Yes, No |
| Image display and processing | Touch Screen | Yes, No |
| Image display and processing | Monitor display size, cm (in) | 12 Inch(±1), 15 Inch(±1), 19 Inch(±1), 21 Inch(±1) Or higher |
| Image display and processing | Split screen | Twin View, Twin View and Live Quad Screen |
| Image display and processing | Single/dual monitors | Single, Single with optional second console, Dual (image display on touch panel) |
| Image display and processing | Speckle-tracking strain and strain rate | Yes, No |
| Image display and processing | Analysis and calculation packages | 2-D wall motion tracking, 3-D wall motion tracking, Cardiac, Mitral valve analysis, User-programmable calculations, No |
| Image display and processing | Type of printer provided with the echo cardio machine | Themal Printer, B/W Laser Printer, Color Laser Printer |
| Image display and processing | Type of processor in PC | i7 or Latest, NA |
| Image display and processing | Conectivity between Echo cardiogram and PC | Through IMS/Equivalent Software, NA |
| Image display and processing | PC with all software inclusive interfaced with echocardiograph machine | Yes, NA |
| Image display and processing | Inkjet printer provided with computer for reporting | Yes, No |
| Image display and processing | Glossy Colour Print Paper supplied | 500 sheets, 1000 sheets, Not provided |
| Image display and processing | ECG Cable | 2, NA |
| Image display and processing | Echo cardiography sytem to be DICOM Ready | Yes, No |
| Image display and processing | Dicom Interfacability to HIS | Yes, No |
| Image display and processing | Dicom Interfacability Dicom to PACS | Yes, No |
| Image display and processing | Interfacability Dicom to RIS | Yes, No |
Discover companies most likely to bid on this tender
GST registration certificate
PAN card
Experience certificates for similar echocardiography systems
Financial statements for last 3 years
EMD submission proof (DD/online payment receipt)
Technical bid documents and compliance statements
OEM authorizations for critical components
Product brochures and system configurations
DICOM/Warehouse/Interfacing capability documentation
Extended Deadline
27-Feb-2026, 12:00 pm
Opening Date
27-Feb-2026, 12:30 pm
Extended Deadline
24-Feb-2026, 12:00 pm
Opening Date
24-Feb-2026, 12:30 pm
Key insights about KERALA tender market
Bidders must meet eligibility criteria, submit EMD of ₹280,000, provide OEM authorizations, product brochures, and technical compliance for 2D/3D/4D imaging, DICOM Ready interfaces, and multiple probe configurations. Ensure compliance with option clause for quantity changes up to 50% and provide installation plan.
Required documents include GST certificate, PAN, last 3 years financials, experience certificates for similar systems, EMD receipt, OEM authorizations, technical bid and brochures, DICOM interfacing proof, and post‑purchase service commitments.
System must support dedicated 3D/4D volume imaging for TTE/TEE, real-time live 3D imaging, color Doppler 3D/4D, and speckle-tracking strain; DICOM Ready with interoperable HIS/PACS/RIS interfaces and cine memory ≥ 350 MB.
EMD is ₹280,000; payment terms to be confirmed in final tender but typically via online transfer or demand draft; ensure receipt proof and alignment with government payment norms.
Delivery period can be adjusted by the option clause up to 50% quantity variation; base delivery timeline is defined in the contract, with minimum delivery extensions in line with the formula provided.
System must be DICOM Ready with HIS/PACS/RIS interfacability; HD display ≥21 inches; ECG trigger; tissue colorization; 2D/3D, TEEs, and workstation integration with PC software.
Probes include TTE and TEE adult/pediatric, pediatric 3D/4D TEE probes, linear vascular probes, and multiple ports with electronic switching; dedicated 3D/4D volume imaging and high frame rate reviews are required.
Obtain official OEM authorization letters, provide full product brochures, and attach detailed compliance matrices mapping each tender specification to the offered hardware and software features.
Sign up now to access all documents
Main Document
BOQ
CATALOG Specification
ATC