Mobile Digital Radiography System (V2)
N/a
Northern Railway has released a public tender for Mobile Digital Radiography System (V2) (Q2) in AMBALA, HARYANA. Quantity: 1. Submission Deadline: 09-03-2026 11: 00: 00. Download documents and apply online.
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS
N/a
N/a
UMARIA, MADHYA PRADESH
Chacha Nehru Bal Chikitsalaya
EAST DELHI, DELHI
Fakhruddin Ali Ahmed Medical College And Hospital
BARPETA, ASSAM
Secretariat
VIRUDHUNAGAR, TAMIL NADU
Tender Results
Loading results...
| Category | Specification | Requirement |
|---|---|---|
| General | Product Description | Mobile Digital Radiography System |
| General | Scope of supply | Compact, easily transportable, digital mobile radiographic unit with articulated/telescopic arm |
| Digital Detector | Radiography Mode | Digital |
| Digital Detector | Number of detector | 1 |
| Digital Detector | Detector Material | Cesium Iodide |
| Digital Detector | Detector Size | 14 inches x 17 inches(wireless ) |
| Digital Detector | Matrix size | 2kx2k |
| Digital Detector | Pixel size in µm | <=150 |
| Digital Detector | Grey Scale resolution, bits/pixel | 14 or better |
| Digital Detector | DQE at 0 LP/mm | 65% or more |
| Digital Detector | Machine should have a detector storage compartment | Yes |
| Digital Detector | Weight of detector in kg | <=5 |
| Digital Detector | Detector should be designed and calibrated for General Radiography Purposes and must be fully integrated with the mobile unit including the controls | Yes |
| Digital Detector | Detector storage capacity | 100 images or more |
| Control panel for image acquisition | Image preview time in seconds | <=6 |
| Control panel for image acquisition | Control panel | Microprocessor controlled |
| Control panel for image acquisition | Automatic parameter selection | Yes |
| Control panel for image acquisition | Control panel entry | Touch screen |
| Control panel for image acquisition | Anatomic specific post processing | Yes |
| Control panel for image acquisition | Display | Digital display of mAs and KV and an electronic timer |
| Control panel for image acquisition | Inbuilt console should have touch TFT screen and control | Yes |
| Control panel for image acquisition | Console should enable to view the image, and provide post processing features, using touch screen | Yes |
| Control panel for image acquisition | Post processing features | Zoom, contrast and brightness, adjustment, panning annotate, mark and reporting |
| X Ray Generator and Tubes | High Frequency X-ray Generator Power in KW | 3.5 KW |
| X Ray Generator and Tubes | KV range | 40-110 |
| X Ray Generator and Tubes | Focal spot, mm | 0.8 - 2 |
| X Ray Generator and Tubes | Heat Capacity, KHU | 40-100 |
| X Ray Generator and Tubes | Anode Type | Stationary |
| X Ray Generator and Tubes | Exposure time in ms | 0.1 to 3 Sec |
| X Ray Generator and Tubes | Control Panel and X-Ray Tube | Yes |
| Tube Suspension | X Ray tube output should match with generator | Yes |
| Tube Suspension | Type of suspension | Articulated arm |
| Tube Suspension | Collimator rotation | ± 90 degree with on off timer |
| Ergonomics | Unit should have small foot print | Yes |
| Ergonomics | Height of the column in cm (Should facilitate easy transportation in the lift and areas with small height doors etc) | <=150 |
| Ergonomics | Weight of mobile X Ray unit in kg | <=600 |
| Ergonomics | Cables should be concealed in the arm system | Yes |
| Ergonomics | Exposure release switch | Detachable with a cord of at least 5 meters. |
| Workstation | Inbuilt work station | Yes |
| Workstation | Monitor size in inches | 14 |
| Workstation | Processor | i 5 |
| Workstation | RAM in GB | 4 |
| Workstation | HDD | 500 GB |
| Workstation | Dicom compatibility of work station | Yes |
| Workstation | Inbuilt console should have touch screen control | Yes |
| Workstation | Number of images which can be stored | 10000 or more |
| Workstation | Machine should be fully network ready and it should be possible to transfer images and patient data from and to hospital network using LAN connectivity or wireless LAN | Yes |
| Accessories | Stitching hardware and software | No |
| Accessories | Integration | DICOM 3.0 storage SOP Classes |
| Accessories | Lead Apron | Yes |
| Accessories | Light weight ‘zero lead’ radiation protection aprons provided | Yes |
| Accessories | Lead Aprons - Full body and Skirt type | 2 no. each (0.5pb) |
| Accessories | Thyroid shields provided | Yes |
| Accessories | Gonadsheet Male provided | Yes |
| Accessories | Gonad sheet Female provided | Yes |
| Accessories | Extractable measurable tape should be available | Yes |
| Accessories | Motorized movement for complete machine | No |
| Accessories | Castor wheel | Yes |
| Accessories | Unit should have effective breaking system for parking | Yes |
| Accessories | Power Requirements | 230 V 50 HZ Single phase -16 Ampls socket |
| Accessories | Machine should run on mains as well as battery supply | No |
| Certifications | Compliance to Medical Device Rules (MDR) 2017 as amended till date | Yes |
| Certifications | Availability of valid Medical Device license for the product issued from the competent authority defined under Drugs and Cosmetic Act 1940 and Rules made there under as amended till date | Yes |
| Certifications | Certification for manufacturing unit | ISO:13485 (Latest) |
| Certifications | Availability of Test Report for each supplied batch/product as per Medical Device Rule (MDR) 2017 as amended till date | Yes |
| Certifications | Conformity to Safety Standards | IEC 60601-2-54 or equivalent BIS |
| Certifications | Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission or along with supplies as per buyer requirement | Yes |
| Certifications | Conformity to Indian Standard IS 7620- Part 1, Part 2, Part 3 | Yes |
| Miscellaneous Parameters | OEM/Reseller (if supplied to reseller shall ensure uninterputed availabilty of all spares for 10 years) | Yes |
| Miscellaneous Parameters | QA, QC and other tests as per NABH shall be carried out by OEM or authorized agency of OEM | Yes |
| Warranty | Warranty in Years | 5 Or higher |
Discover companies most likely to bid on this tender
Past Performance
Certificate (Requested in ATC)
OEM Authorization Certificate *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
N/a
📍 NAINITAL, UTTARAKHAND
Gujarat Medical Education And Research Society (gmers)
📍 AHMEDABAD, GUJARAT
Health Department
📍 PALAKKAD, KERALA
N/a
📍 NAINITAL, UTTARAKHAND
Dg Armed Forces Medical Service
Sign up now to access all documents
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS