LIQUID CHROMATOGRAPHY-MASS SPECTROMETER -TRIPLE QUADRUPOLE
National Institute Of Pharmaceutical Education And Research (niper) Kolkata
Bid Publish Date
27-Oct-2025, 12:57 pm
Bid End Date
10-Dec-2025, 8:00 pm
EMD
₹4,60,000
Location
Progress
Quantity
1
Bid Type
Two Packet Bid
Organization: West Bengal Medical Services Corporation Ltd. in Kolkata, under the Health And Family Welfare Department, seeks a fully integrated Gas Chromatography – Triple Quadrupole Mass Spectrometry (GC-MS-MS) system for regional laboratory use. Estimated scope includes multi-detector GC configuration with inclusive detectors and capillary columns, compatible liquid headspace, and desktop workstation. The bid includes on-site installation, commissioning, and a 3-year OEM warranty. The project requires a robust, scalable GC-MS-MS platform with dedicated service support and PESO-certified gas supply. The award favors systems with dual-detection capability, multiple injection ports, and integrated data handling, enabling concurrent results across detectors. This procurement targets the Siliguri regional facility and similar consignees, reflecting a high-complexity analytical instrument deployment in West Bengal. The tender emphasizes compliance, service readiness, and long-term spares availability, with strict eligibility for manufacturers and their Indian partners.
GC-MS-MS system with inclusive detectors (FID, TCD, ECD, NPD/FTD, FPD, PID, PFPD) and inlets
Capillary columns: number provided; length as per system; type of column specified
Detector MDL and sensitivity for full-scan and SRM/MRM modes
Headspace sampler compatible; liquid auto sampler with specified vial capacity
Pyrolyzer compatibility for solids/liquids; compatible desktop workstation with graphics
UL/CE certifications; EN 61010-1:2010 compliance; EN 61326-1:2013 EMI/EMC
NABL/ILAC test reports proving conformity; PESO-certified gas cylinders and regulators
EMD amount ₹460000 via RTGS/online transfer with bid submission
Delivery location: Regional Food Testing Laboratory, Siliguri; on-site installation
3-year OEM warranty and post-sale service with state-based center
EMD payment via RTGS/NEFT; post-delivery milestones linked payments per contract; proof required
Delivery and installation at Siliguri facility; commissioning included; proof of service center readiness
Penalties may apply for delayed installation or non-compliance with warranty obligations; terms to be specified in contract
In-state service center capability or guaranteed establishment within 30 days of award
Previous experience with GC-MS-MS systems or equivalent analytical platforms
Financial stability evidenced by audited statements and solvency indicators
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS
National Institute Of Pharmaceutical Education And Research (niper) Kolkata
Jawaharlal Institute Of Postgraduate Medical Education And Research (jipmer)
PONDICHERRY, PUDUCHERRY
Ap Forensic Science Laboratory
GUNTUR, ANDHRA PRADESH
N/a
SRINAGAR, JAMMU & KASHMIR
Indian Institute Of Technology (iit)
Tender Results
Loading results...
| Category | Specification | Requirement |
|---|---|---|
| GAS CHROMATOGRAPH (GC) CONFIGURATION | GC Detectors inclusive in the scope of supply | Flame Ionization Detector (FID), Thermal Conductivity Detector (TCD or Hot Wire Detector), Electron Capture Detector (ECD), Nitrogen–Phosphorus Detector (NPD) / Flame Thermionic Detector (FTD), Flame Photometric Detector (FPD), Photo Ionization Detector (PID), Pulsed Flame Photometric Detector (PFPD) |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of FID in the System - inclusive in the scope of Supply | 0, 1, 2 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of TCD in the System - inclusive in the scope of Supply | 0, 1, 2 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of ECD in the System - inclusive in the scope of Supply | 0, 1, 2 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of NPD/FTD in the System - inclusive in the scope of Supply | 0, 1, 2 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of FPD in the System - inclusive in the scope of Supply | 0, 1, 2 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of PID in the System - inclusive in the scope of Supply | 0, 1, 2 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "Number of PFPD in the System - inclusive in the scope of Supply | 0, 1, 2 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | Number of Inlets GC can Simultaneously Support | 1, 2, 3 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | Number of Detectors (Ports) GC can Simultaneously Support | 1, 2, 3, 4 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | Number of Capillary Columns provided in GC - inclusive in the scope of Supply | 0, 1, 2 |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | Type of Capillary Column | Polar, Non-Polar, Semi-Polar, NA |
| GAS CHROMATOGRAPH (GC) CONFIGURATION | "GC Capillary Column Length | 0, 30, 60, 100 |
| PERFORMANCE | "GC Detector Minimum Detection Level (MDL) | < 1.4 pg C/s for FID, < 1.8 pg C/s for FID, < 3.0 pg C/s for FID < 400 pg/ml for TCD, < = 1 microgram/ml for TCD, < = 10 microgram/ml for TCD, < 4.5 fg /sec for ECD, < 5.5 fg /sec for ECD, < 0.1pg N /sec for NPD/FTD, < 0.03 pg P /sec for NPD/FTD, < 100 fg P /sec for FPD, <2.5 pg S /sec for FPD, <5 pg S /sec for FPD, < 45 fg P /sec for FPD |
| COLUMN OVEN (GAS CHROMATOGRAPH - GC) | Maximum Operating Temperature (GC Oven) | upto 400 deg. C, upto 450 deg. C Or higher |
| COLUMN OVEN (GAS CHROMATOGRAPH - GC) | Number of Oven Ramps / Plateaus for Temperature programming | 6/7 or more, 9/10 or more, 20/21 or more |
| PNEUMATIC CONTROL | Number of Injection Ports | 1, 2 |
| PNEUMATIC CONTROL | Maximum Injection Port heating Temperature | upto 400 deg. C, upto 450 deg. C Or higher |
| PNEUMATIC CONTROL | Pressure Range | upto100 psi, upto 140 psi Or higher |
| PNEUMATIC CONTROL | Pressure set point Resolution (±)(psi) | 0.001, 0.01, 0.1 Or lower |
| PNEUMATIC CONTROL | Type of Sample Injection Port | Split, Splitless, split / Splitless, Programmable Temperature Vaporizer (PTV) |
| PNEUMATIC CONTROL | Split Ratio | 6000:1, 7000:1, 7500:1, NA |
| AUTO INJECTOR / AUTO SAMPLER | Compatible Liquid Auto Sampler - inclusive in the scope of supply | Yes, No |
| AUTO INJECTOR / AUTO SAMPLER | Capacity of the Liquid Auto Sampler (Number of Vials) | 0, 8, 100, 150 Or higher |
| AUTO INJECTOR / AUTO SAMPLER | Number of Injection Ports in the Liquid Auto Sampler | 0, 1, 2 |
| AUTO INJECTOR / AUTO SAMPLER | Number of Channels in the Liquid Auto Sampler | 0, 2 |
| AUTO INJECTOR / AUTO SAMPLER | "Liquid Auto Sampler Features (Select applicable features Only)" | Fully automated Liquid Auto Sampler, Injects with variable speed & varying syringe sizes, Operates through Software (Software inclusive in the scope of supply), NA |
| HEAD SPACE SAMPLER | Compatible Head Space Sampler- inclusive in the scope of supply | Yes, No |
| HEAD SPACE SAMPLER | Capacity of the Head Space Sampler Carousel (number of Vials) | 12 or better, NA |
| HEAD SPACE SAMPLER | Head Space Sampler Temperature Range | Upto 200 deg. C or better, NA |
| PYROLYZER SYSTEM | Compatible Pyrolyzer System for pyrolysis of solid and liquid samples- inclusive in the scope of supply | Yes, No |
| MASS SPECTROPHOTOMETER (MS-MS) SYSTEM | Ionization modes | Electron Ionization (EI), EI/PCI/NCI |
| MASS SPECTROPHOTOMETER (MS-MS) SYSTEM | Detector Type | Electron Multiplier, Triple-Axis, HED-EM |
| MASS SPECTROPHOTOMETER (MS-MS) SYSTEM | Operating modes | Full Scan Mode, Selective Ion Monitoring (SIM) Mode, Both Full Scan & SIM Modes, Selective /Multiple Reaction Monitoring (SRM/MRM) |
| PERFORMANCE OF THE MASS SPECTROPHOTOMETER (MS-MS) SYSTEM | Sensitivity (S/N Ratio) in full scan | 300:1, 1500:1 Or higher |
| PERFORMANCE OF THE MASS SPECTROPHOTOMETER (MS-MS) SYSTEM | Sensitivity (S/N Ratio) in SRM/MRM | 6000:1, 15000:1, 30000:1 Or higher |
| GENERIC | Power Supply | 230 V ± 10%, 415 V ± 10% |
| GENERIC | Built-in Display | Yes, No |
| GENERIC | Provision to split the sample from single injection to two different detectors and get results for two detectors simultaneously | Yes, No |
| GENERIC | Back Flash - inclusive in the scope of supply | Pre Column Back Flush, Post Column Back Flash, Mid Column Back Flash, NA |
| SALIENT FEATURES OF GC-MS-MS | "Salient Featutes of GC-MS (Select applicable features Only)" | Automatic leak testing/detection of Pneumatics, Direct setting of split flow rates and ratios, Flow sensor for control & storage of split ratio, Split and Septum Purge gases, Able to direct inject onto packed and wide-bore capillary columns, Oven power turns-off automatically when the lid/door is opened, Over Heat Protection, Display Unit with built-in Key Board, Controlled by External PC System, Programmable ion source temperature, System includes variety of auto tune algorithms to tune the instrument for maximum sensitivity or for specific target compounds, EI source is inert to active compounds and programmable, EI Source have a dual filament design with automatic selection of other filament if one fails during analysis;, NA, Off-axis detection system for better removal of neutrals, Electron Impact ion source with dual filaments temperature up to 350 °C, Emission current should be able to set up to 300 µA or better, MRM/SRM Transitions should be 800 transitions/sec, Buyer specific columns - DB-fame /equivalent column 60 m x 0.25 mm id x 0.25 micron- 2 nos , DB-1 /equivalent column 60 m x 0.25 mm id x 0.25 micron- 1 nos, Instrument detection limit sensitivity of < 2 fg during installation using 5 fg of OFN with 1 ul injection with Auto sampler |
| LIBERARIES | Libraries for GC-MS | NIST Library for MS - Licensed version - 2017 or latest, MRM data base Library, Library for Pesticides, Library for Environmental, Library for Forensic, Wiley Library, NA |
| SOFTWARE | "Software Features (Select Applicable Features only)" | Compatible with Windows 7 or higher version operating system, Seamless integration & control of all the GC parameters, Interface with Multiple GCs through the comprehensive real-time monitoring and diagnostic software, Real time notification via advisories and indicators, Counters, Chromatographic attributes, Instrument diagnostics, Access to maintenance information, Access to maintenance, run and event logs, Retention Time - Locking module for analyzing target compounds in complex matrices |
| ACCESSORIES | Gas Supply inclusive in the scope of supply | Nitrogen, Hydrogen, Argon, Helium, Zero Air, Not offered |
| ACCESSORIES | Number of Cyliners of each type supplied - inclusive in the scope of supply | 0, 1, 2 |
| ACCESSORIES | Gas cylinders certified by PESO | Yes, NA for Gas Cylinder not offered |
| ACCESSORIES | Regulator - inclusive in the scope of supply | Dual Stage Regulator, NA |
| ACCESSORIES | "Other Accessories inclusive in the scope of supply (Select only those Accessories which are part of supply)" | Gas purification panel for all Gases with moisture trap, Hydrocarbon trap & Oxygen trap, Syringe for manual injection x 05 nos, Auto Sampler Vials x 1000 vials with screw cap, Graphite ferrules x 10 nos, Vessel ferrules x 10 nos, Septa for injector x 100 nos, Inlet liner for splitles x 4 nos, Inlet liner for split x 4 nos, O-rings x 10 nos. of each type, Column Cutter x 2 nos, Gas tube Cutter x 2 nos, Tool Kit x 1 no., User / Technical / Maintenance Manual, Standard Samples - NIST / IPL Traceable, Direct Injection Port, Splitter, Multishot Pyrolyser, Singleshot Pyrolyser, Syringe for Auto Sampler injection x 10 nos, EI filaments x 5 nos, NA |
| DESKTOP WORKSTATION | Desktop workstation Suitable to GC-MS - inclusive in the scope of supply | Yes, No |
| DESKTOP WORKSTATION | Graphics Card Memory Size | 0, 1, 2, 4 |
| ONLINE UPS | UPS Provided in the Scope of Supply | Yes, No |
| ONLINE UPS | Rating of online UPS(KVA) | 5.0 KVA, 7.5 KVA, 10 KVA, NA |
| ONLINE UPS | Backup time of online UPS | 0, 60, 90, 120 |
| INSTALLATION & COMMISSIONING / WARRANTY / AVAILABILITY OF SPARES / TRAINING | Warranty on equipment | 1, 2, 3, 5 |
| INSTALLATION & COMMISSIONING / WARRANTY / AVAILABILITY OF SPARES / TRAINING | Number of days onsite training provided | 3, 5, 7, 10 |
| CERTIFICATION / TEST REPORT | "Availability of UL/CE Certificate as per EN 61010- 1:2010 Safety requirements for electrical equipment for measurement,control, and laboratory use General requirements" | Yes, No |
| CERTIFICATION / TEST REPORT | Conformance to EMC/EMI as per EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use | Yes, No |
| CERTIFICATION / TEST REPORT | Availability of Test Reports from Central Govt / NABL approved / ILAC accredited Lab to prove conformity to the specification | Yes, No |
Discover companies most likely to bid on this tender
GST registration certificate
Permanent Account Number (PAN) card
Experience certificates for GC-MS/MS or equivalent analytical systems
Audited financial statements for the last 2 years
EMD submission proof (online RTGS/NEFT receipt)
OEM authorization/authorization letter for the proposed GC-MS-MS system
Technical compliance certificates and NABL/ILAC lab test reports (where applicable)
Installation, commissioning, and maintenance service capability proof
End-user certificate or declaration (if requested)
Extended Deadline
10-Dec-2025, 8:00 pm
Opening Date
10-Dec-2025, 8:30 pm
Extended Deadline
29-Nov-2025, 8:00 pm
Opening Date
29-Nov-2025, 8:30 pm
Key insights about WEST BENGAL tender market
Bidders should submit the EMD ₹460000 via RTGS/NEFT to WBMSCL with bid documents including GST, PAN, OEM authorization, NABL/ILAC reports, and technical compliance. Include installation plans, service center readiness within 30 days, and 3-year OEM warranty details.
Required documents include GST certificate, PAN, audited financial statements, experience certificates for GC-MS/MS, OEM authorization, EMD proof, technical compliance certificates, and NABL/ILAC test reports. End-user certificates may be requested per ATC.
System must comply with EN 61010-1:2010 safety requirements and EN 61326-1:2013 EMI/EMC. Availability of UL/CE certificates, NABL/ILAC conformity reports, and PESO-certified gas cylinders is required for acceptance.
Delivery is to Regional Food Testing Laboratory, Siliguri. Installation and commissioning are included, with on-site support and a guaranteed 3-year OEM warranty; post-installation training is part of service.
EMD is ₹460,000 payable via RTGS/NEFT to West Bengal Medical Services Corporation Ltd. Include the transaction proof with the bid submission and reference the bid name in the transfer details.
Bidder must have a functional service center in-state or establish one within 30 days of award, provide prior GC-MS/MS experience, and demonstrate financial stability via audited statements and OEM support capability.