Bid Publish Date
18-Aug-2026, 11:39 am
Bid End Date
09-Sep-2026, 3:00 pm
EMD
₹50,000
Location
Progress
Public procurement opportunity for All India Institute Of Medical Sciences (aiims) Fourier Transform Infra Red (FTIR) Spectrometer (Q2) in JODHPUR, RAJASTHAN. Quantity: 1 issued by. Submission Deadline: 09-09-2026 15: 00: 00. View full details and respond.
Quantity
1
Bid Type
Two Packet Bid
Bid Validity
180 (Days)
Bid Type
Service
Evaluation
Total value wise evaluation
Inspection Required
No
RA Qualification Rule
50% Lowest Priced Technically Qualified Bidders
Tech Clarification Time
7 Days
EMD Required
Yes
MII Purchase Preference
Yes
MII Preference Band
L1+20%
MSE Purchase Preference
Yes
MSE Preference Band
L1+15%
MSE Exemption/Relaxation
Yes
Startup Exemption/Relaxation
Yes
Bid Splitting Applied
No
Warranty Period
5 years
Arbitration Clause
No
Mediation Clause
No
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
Yes
Item Category
Fourier Transform Infra Red (FTIR) Spectrometer (Q2)
Advisory Bank
Bank Of Baroda
ePBG Percentage
3%
ePBG Duration (Months)
62
Pre-Bid Date
27-08-2026 16:00:00
Pre-Bid Venue
Medical College, AIIMS Jodhpur
Max Delivery Days
60
Delivery Locations
1
Delivery Cities
Jodhpur
Delivery Pincodes
342005
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Abhishek Singh | 342005,AIIMS Basani Industrial Area, Phase II Jodhpur | Jodhpur | Rajasthan | 342005 | 1 | 60 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
OTHER
CATALOG Specification
OTHER
OTHER
GEM_GENERAL_TERMS_AND_CONDITIONS
Discover companies most likely to bid on this tender
Experience Criteria
Certificate (Requested in ATC)
OEM Authorization Certificate
Additional Doc 1 (Requested in ATC) *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
| Category | Specification | Requirement |
|---|---|---|
| OPTICAL PERFORMANCE | Wavenumber- Maximum (Per cm) | 6000 |
| OPTICAL PERFORMANCE | Wavenumber - Minimum (Per cm) | 350 |
| OPTICAL PERFORMANCE | Spectral (Wavenumber) Resolution (Per cm) | 0.5 |
| OPTICAL PERFORMANCE | Spectral (Wave Number) Accuracy (Per cm) (+/-) | 0.05 |
| OPTICAL PERFORMANCE | Beam Splitter | KBr |
| OPTICAL PERFORMANCE | Automatic beam splitter changeover for MIR to NIR or MIR to FIR without manual intervention | Yes |
| OPTICAL PERFORMANCE | Mirror Coating | Gold |
| OPTICAL PERFORMANCE | Peak to Peak Signal to Noise Ratio for 1 minute scan | 30000:1 or more |
| DETECTOR | Detector Type | Deuterated Triglycine Sulphate (DTGS) |
| DETECTOR | Detector having temperature control | Yes |
| LIGHT SOURCE | Light Source | Mid IR Source |
| SAMPLE TYPE / ANALYSIS FACILITY | Analyze Liquid samples | Yes |
| SAMPLE TYPE / ANALYSIS FACILITY | Analyze Solid samples | Yes |
| SAMPLE TYPE / ANALYSIS FACILITY | Analyze Powder samples | Yes |
| SAMPLE TYPE / ANALYSIS FACILITY | Analyze thin films on metals samples | Yes |
| SAMPLE TYPE / ANALYSIS FACILITY | Analyze Gas samples | Yes, No |
| SOFTWARE | Customizable Report Template | Yes, No |
| FTIR LIBRARIES | FTIR Georgian Crime Library - inclusive in the scope of supply | No |
| FTIR LIBRARIES | FTIR Canadian Forensic Library - inclusive in the scope of supply | No |
| FTIR LIBRARIES | FTIR Sadler Drug Library - inclusive in the scope of supply | No |
| FTIR LIBRARIES | FTIR Polymers & Polymer additives Library, nclusive in the scope of supply | Yes, No |
| FTIR LIBRARIES | FTIR Surfactants Library, inclusive in the scope of supply | No |
| FTIR LIBRARIES | FTIR Minerals & inorganic CompoundsLibrary, inclusive in the scope of supply | Yes, No |
| FTIR LIBRARIES | FTIR Organic Compounds Library, inclusive in the scope of supply | Yes, No |
| FTIR LIBRARIES | FTIR Drugs Library, inclusive in the scope of supply | No |
| FTIR LIBRARIES | FTIR Adhesives & Sealants Library, inclusive in the scope of supply | No |
| FTIR LIBRARIES | FTIR Cosmetics & Personal products Library, inclusive in the scope of supply | Yes, No |
| ACCESSORIES | Suitable Hydraulic Press 15 Ton for pellet/sample preparation - inclusive in the scope of supply | No |
| ACCESSORIES | Suitable Mortar and pestle - inclusive in the scope of supply | Yes |
| ACCESSORIES | Suitable Dies - inclusive in the scope of supply | Yes, No |
| ACCESSORIES | Suitable KBr set for Pellets - inclusive in the scope of supply | Yes, No |
| ACCESSORIES | KBr Powder - inclusive in the scope of supply | Yes, No |
| ACCESSORIES | Suitable Pellet holder - inclusive in the scope of supply | Yes, No |
| ACCESSORIES | Types of Suitable ATR - inclusive in the scope of supply | Yes |
| ACCESSORIES | ATR crystal - inclusive in the scope of supply | Diamond |
| ACCESSORIES | Suitable Universal liquid cell holder - inclusive in the scope of supply | Yes, No |
| ACCESSORIES | Suitable Dry Box - inclusive in the scope of supply | Yes, No |
| ACCESSORIES | Suitable Replacement dessicant (rechargeable) - inclusive in the scope of supply | Yes, No |
| ACCESSORIES | Suitable Liquid Cell Window - inclusive in the scope of supply | No |
| ACCESSORIES | Liquid Cell Window material | KBr |
| ACCESSORIES | Shape of the Cell Window | Round |
| ACCESSORIES | Suitable Gas cell - inclusive in the scope of supply | Yes, No |
| GENERIC | FTIR and Sample compartment having provision for purging with gas | No |
| GENERIC | FTIR provided with in-built air drier for humidity and vapour protection | No |
| GENERIC | Connectivity | USB, WIRELESS, WIFI, ETHERNET, Blue tooth, RS 232 |
| OPERATING CONDITIONS | Minimum Operating Temperature | 10.0 |
| OPERATING CONDITIONS | Maximum Operating Temperature | 50.0 |
| OPERATING CONDITIONS | Operating Humidity (RH) (%) at 40 degree C | 85.0 |
| WARRANTY / AVAILABILITY OF SPARES / TRAINING | Warranty on equipment | 1, 2, 3, 5 |
| WARRANTY / AVAILABILITY OF SPARES / TRAINING | Warranty on Source | 2.0 |
| WARRANTY / AVAILABILITY OF SPARES / TRAINING | Warranty on interferometer | 10.0 |
| WARRANTY / AVAILABILITY OF SPARES / TRAINING | Warranty on Laser | 10.0 |
| WARRANTY / AVAILABILITY OF SPARES / TRAINING | Onsite training | 5.0 |
| CERTIFICATION | Availability of UL/CE Certification as per EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control, and laboratory use General requirements | Yes |
| CERTIFICATION | Conformance to EMC/EMI as per EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use | No |
| CERTIFICATION | Availability of Test Reports from Central Govt / NABL approved / ILAC accredited Lab to prove conformity to the specification | No |