Bid Publish Date
25-Aug-2026, 6:05 pm
Bid End Date
08-Sep-2026, 6:00 pm
Location
Progress
Directorate Of Purchase And Stores announces a tender for Spectrum Analyzers and Vector Network Analyzers (Q3) in MUMBAI, MAHARASHTRA. Quantity: 1. Submission Deadline: 08-09-2026 18: 00: 00. Last date to apply is approaching fast!
Quantity
1
Bid Type
Two Packet Bid
Bid Validity
130 (Days)
Bid Type
Service
Evaluation
Total value wise evaluation
Inspection Required
No
Tech Clarification Time
4 Days
EMD Required
No
MII Purchase Preference
No
MSE Purchase Preference
Yes
MSE Preference Band
L1+15%
MSE Exemption/Relaxation
Yes
Startup Exemption/Relaxation
Yes
Bid Splitting Applied
No
Warranty Period
5 years
Arbitration Clause
No
Mediation Clause
No
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
No
Item Category
Spectrum Analyzers and Vector Network Analyzers (Q3)
Advisory Bank
State Bank of India
ePBG Percentage
5%
ePBG Duration (Months)
66
Max Delivery Days
120
Delivery Locations
1
Delivery Cities
Mumbai Suburban
Delivery Pincodes
400085
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Dattatray Ashok Vispute | 400085,Material Receipt Cell (MRC), North Gate, Central Stores Unit, BARC, TROMBAY, MUMBAI | Mumbai Suburban | Maharashtra | 400085 | 1 | 120 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
OTHER
CATALOG Specification
OTHER
GEM_GENERAL_TERMS_AND_CONDITIONS
Discover companies most likely to bid on this tender
Certificate (Requested in ATC) *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
| Category | Specification | Requirement |
|---|---|---|
| FEATURES | Description of the itemuu | Spectrum analyzers |
| GENERAL FEATURES COMMON TO SPECTRUM AND VECTOR ANALYSER | Markers | Normal, delta |
| GENERAL FEATURES COMMON TO SPECTRUM AND VECTOR ANALYSER | Trigger | Free run, video, External |
| GENERAL FEATURES COMMON TO SPECTRUM AND VECTOR ANALYSER | Display | TFT Colour display with provision for three traces |
| GENERAL FEATURES COMMON TO SPECTRUM AND VECTOR ANALYSER | Interface | USB, LAN |
| GENERAL FEATURES COMMON TO SPECTRUM AND VECTOR ANALYSER | Software for transfer of screen shot/measurement data to PC | Yes |
| SPECTRUM ANALYZER FEATURES | Frequency Range | 9 kHz to 3 GHz |
| SPECTRUM ANALYZER FEATURES | Aging rate of spectrum analyzer frequency: Better than 1x10-6 /Year | Yes |
| SPECTRUM ANALYZER FEATURES | Frequency counter Resolution(minimum) (Hz) | 0.001Hz |
| SPECTRUM ANALYZER FEATURES | Frequency Span | 100Hz to 3GHz |
| SPECTRUM ANALYZER FEATURES | Spectral purity:SSB Phase Noise @10KHz offset shall be less than -90dBc/Hz at 1GHz | Yes |
| SPECTRUM ANALYZER FEATURES | Spectral purity:Harmonics @ -40dBm level <=60 dBc | yes |
| SPECTRUM ANALYZER FEATURES | Sweep time @span >or= 1 KHz | 0.1S to 1000 s |
| SPECTRUM ANALYZER FEATURES | Sweep time @span = 0 Hz | 10 micro S to 6000 S |
| SPECTRUM ANALYZER FEATURES | Resolution Bandwidth (-3dB) | 1 Hz to 8 MHz |
| SPECTRUM ANALYZER FEATURES | Video Bandwidth | 1 Hz to 8 MHz |
| SPECTRUM ANALYZER FEATURES | Amplitude : Max input level for protection (dBm) | +30 |
| SPECTRUM ANALYZER FEATURES | Amplitude : Max input level for measurement (dBm) | 20 |
| SPECTRUM ANALYZER FEATURES | Amplitude :level measurement uncertainity (dB) | <=1, <=±1 |
| SPECTRUM ANALYZER FEATURES | Average noise level displayed(@300HzRBW,10Hz VBW) | <=-130dBm for 10MHz to 3 GHz & |
| SPECTRUM ANALYZER FEATURES | Average noise level displayed(@100HzRBW,1Hz VBW) | <=-130dBm for 10MHz to 3 GHz & |
| SPECTRUM ANALYZER FEATURES | Average noise level displayed(@300HzRBW,1Hz VBW) | <=-130dBm for 10MHz to 3 GHz & |
| SPECTRUM ANALYZER FEATURES | Measurement features | C/N ratio,AM%,time domain power measurement,channel power measurement,Phase noise measurement,TOI(THIRD ORDER INTERCEPT),OBW(occupied bandwidth) |
| VECTOR NETWORK ANALYZER FEATURES | Frequency Range for 2port vector network analyzer | NA |
| VECTOR NETWORK ANALYZER FEATURES | Frequency Resolution (CW ON) (Hz) | NA |
| VECTOR NETWORK ANALYZER FEATURES | Return Loss (dB) | NA |
| VECTOR NETWORK ANALYZER FEATURES | VSWR (dB) | NA |
| VECTOR NETWORK ANALYZER FEATURES | Cable Loss (dB) | NA |
| VECTOR NETWORK ANALYZER FEATURES | Calibration with Precision calibration components-open, short, load should be included to achieve directivity >35 dB hint: NA for items designated as spectrum analyzers | NA |
| VECTOR NETWORK ANALYZER FEATURES | Abilty to locate Distance-to-fault in Co-axial cables | NA |
| VECTOR NETWORK ANALYZER FEATURES | Phase Range (degree) | NA |
| VECTOR NETWORK ANALYZER FEATURES | Phase Resolution (deg) | NA |
| VECTOR NETWORK ANALYZER FEATURES | RF output port | NA |
| VECTOR NETWORK ANALYZER FEATURES | Dynamic Range | NA |
| VECTOR NETWORK ANALYZER FEATURES | Nominal port power output: Hi Power/Lo-Power selectable | NA |
| VECTOR NETWORK ANALYZER FEATURES | Measurement of S11 and S21 and Smith Chart | NA |
| VECTOR NETWORK ANALYZER FEATURES | Vector Voltmeter for comparision and verification of the phase of the phase-matched cables | NA |
| OPERATING CONDITIONS | Operating Temperature | 0 to 50 degree C & 80% RH @ 40 degree C |
| OPERATING CONDITIONS | Operating Humidity | 80% RH @ 40 degree C |
| POWER REQUIREMENT | Power requirement | 230 +- 10 VAC, 50Hz |
| POWER REQUIREMENT | chemistry of the battery | NA |
| POWER REQUIREMENT | Suitable charger / adaptor provided | NA |
| WARRANTY | Warranty | 5 Or higher |
| CERTIFICATION/TEST REPORT DETAILS | The equipment supplied with calibration certificate | yes |
| CERTIFICATION/TEST REPORT DETAILS | conformity to Dry Heat for 16 hrs at a specified maximum temp in accordance with IS:9000/part-3/section-5/1977 latest | no |
| CERTIFICATION/TEST REPORT DETAILS | conformity to Cold Test for 4 hrs at a specified minimum temp in accordance with IS:9000/part-2/section-4/1977 latest | no |
| CERTIFICATION/TEST REPORT DETAILS | conformity to Damp Heat (Cyclic)Test for 2 Cycles of 24 h each at a temp of 40 degree C & 95% RH | no |
| CERTIFICATION/TEST REPORT DETAILS | Availability of Test Reports from Central Govt/NABL/ILAC approved accredited lab to prove conformity to the specification | No |