Progress
Bihar Agricultural University invites bids for SPECTROPHOTOMETER (Q2) in BHAGALPUR, BIHAR. Quantity: 1. Submission Deadline: 21-08-2026 13: 00: 00. Submit your proposal before the deadline.
Quantity
1
Bid Type
Two Packet Bid
Bid Validity
60 (Days)
Bid Type
Service
Evaluation
Total value wise evaluation
Inspection Required
No
Tech Clarification Time
2 Days
EMD Required
No
MII Purchase Preference
No
MSE Purchase Preference
Yes
MSE Preference Band
L1+15%
MSE Exemption/Relaxation
No
Startup Exemption/Relaxation
No
Bid Splitting Applied
No
Experience Required
3 Year (s)
Arbitration Clause
No
Mediation Clause
No
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
No
Item Category
SPECTROPHOTOMETER (Q2)
Max Delivery Days
15
Delivery Locations
1
Delivery Cities
Bhagalpur
Delivery Pincodes
813210
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Anshuman Kohli | 813210,NH - 80, RD BUILDING, BIHAR AGRICULTURAL COLLEGE, SABOUR, BHAGALPUR (NEAR SABOUR RAILWAY STATION) | Bhagalpur | Bihar | 813210 | 1 | 15 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
GEM_GENERAL_TERMS_AND_CONDITIONS
Experience Criteria
OEM Authorization Certificate *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
| Category | Specification | Requirement |
|---|---|---|
| LIGHT SOURCE | Light Source | Deuterium Lamp, Tungsten Halogen Lamp |
| LIGHT SOURCE | Diffraction Grating (lines/mm) | 1200 |
| LIGHT SOURCE | Switching of Light Source | Automatic |
| LIGHT SOURCE | Wavelength Setting And Scanning | Automatic |
| TYPE | Type of Spectrophotometer | UV - VISIBLE |
| TYPE | Type of Optical System | Double Beam |
| TYPE | Monochromator Type | Czerny-turner |
| TYPE | Detector Type | Photo Multiplier Tube (PMT) |
| TYPE | Spectral Bandwidth Type | Variable Bandwidth |
| OPTICAL PERFORMANCE | Source Wavelength - Minimum | 190.0 |
| OPTICAL PERFORMANCE | Source Wavelength - Maximum | 1100, 1500, 2000, 2500, 2700, 3000, 3200, 3300, 1650 Or higher |
| OPTICAL PERFORMANCE | Resolution | 0.01 |
| OPTICAL PERFORMANCE | Wavelength Accuracy (+/-) | 0.1 |
| OPTICAL PERFORMANCE | Bandwidth (nm) | 0.1 |
| OPTICAL PERFORMANCE | Focal length (mm) | 250 to 299 |
| OPTICAL PERFORMANCE | Photometric measurement modes | Transmittance, Absorbance, Concentration (Wavelength, time), Reflectance, Energy |
| OPTICAL PERFORMANCE | Photometric Absorbance (Max) (Abs) (+/-) | 4 |
| OPTICAL PERFORMANCE | Photometric Absorbance Accuracy (Max) (+/-) | 0.001 - 0.001 Or lower |
| OPTICAL PERFORMANCE | Photometric Transmittance (Max) (+/-) | 100.0 |
| OPTICAL PERFORMANCE | Photometric Transmittance Accuracy (Max) (+/-) | 0.0 - 0.1 Or lower |
| OPTICAL PERFORMANCE | Photometric Reflectance (Max) (+/-) | 100.0 |
| GENERIC | Display | Through PC |
| GENERIC | Display type | TFT |
| GENERIC | Display Size | 19.0 - 30.0 Or higher |
| GENERIC | Test result printing | Through external printer |
| GENERIC | If inbuilt printer, Type of printer | N/A |
| GENERIC | Internal storage(GB) | 80.0 - 80.0 Or higher |
| GENERIC | Mximum Path Length of sample | 100.0 |
| GENERIC | Number of Cuvettes supplied-inclusive in the scope of supply | 10 |
| GENERIC | Spares and Consumables inclusive in the scope of supply | No |
| OPERATING CONDITIONS | Minimum Operating Temperature | 10.0 |
| OPERATING CONDITIONS | Maximum Operating Temperature | 35.0 |
| OPERATING CONDITIONS | Operating Humidity (RH) (%) at 40 degree C | 80.0 |
| WARRANTY | Onsite Warranty | 2.0 |
| WARRANTY | Availability of Spares (at extra cost) | 7.0 |
| Certification | Availability of UL/CE Certification as per EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control, and laboratory use General requirements | Yes |
| Certification | Conformance to EMC/EMI as per EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use | No |
| Certification | Availability of Test Reports from Central Govt / NABL approved / ILAC acredited lab to prove conformity to the specification | No |