| Category | Specification | Requirement |
|---|---|---|
| Standard parameter | Instrument calibration standards | NA |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Components of unit | Quaternary Low Pressure Gradient pump & Serial Double Plunger |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Number of solvents delivery | 4 |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Flow rate setting range in ml/min | 0.001 to 10 |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Flow rate accuracy | ± 0.1% |
| UV-Vis Detector | UV-Vis Detector provided | Yes, No Or higher |
| UV-Vis Detector | Light source | NA if not provided |
| UV-Vis Detector | Spectrum Band width in nm | NA if not provided |
| UV-Vis Detector | Wavelength range in nm | NA if not provided |
| UV-Vis Detector | Sampling Rate in Hz | NA if not provided |
| UV-Vis Detector | The flow cell must be temperature Controlled | NA if not provided |
| UV-Vis Detector | Wavelength accuracy in nm | NA if not provided |
| UV-Vis Detector | Wavelength reproducibility in nm | NA if not provided |
| UV-Vis Detector | Drift in AU/Hour | NA if not provided |
| UV-Vis Detector | Noise level in AU | NA if not provided |
| UV-Vis Detector | Ability to monitor and quantitative 2 wavelengths simultaneously | NA if not provided |
| UV-Vis Detector | Linearity in AU | NA if not provided |
| UV-Vis Detector | Temperature control provided for Flow cell | NA if not provided |
| UV-Vis Detector | Temperature control range for flow cell in Degree Celsius | NA if not provided |
| UV-Vis Detector | flow cell path length in mm | NA if not provided |
| UV-Vis Detector | Flow cell volume in ?l | NA if not provided |
| Auto Sample Injector with cooler | Injection Method | Total volume sample injection |
| Auto Sample Injector with cooler | Carry over of sample | 0.005% or lower |
| Auto Sample Injector with cooler | Injection precision in RSD | less than 0.50 % RSD |
| Software | Operating Windows | Win 10 or latest |
| Photo Diode Array (PDA) Detector | Provision of Photo diode array | Yes Or higher |
| Photo Diode Array (PDA) Detector | PDA detector Wavelength range in nm | 190- 800 or better |
| Photo Diode Array (PDA) Detector | Number of Diodes | 1024 |
| Photo Diode Array (PDA) Detector | PDA detector light source | Deuterium lamp and Tungsten lamp |
| Photo Diode Array (PDA) Detector | PDA detector Wavelength accuracy in nm | ± 1 |
| Photo Diode Array (PDA) Detector | Data Rate in Hz | 100, <100 |
| Photo Diode Array (PDA) Detector | Noise in AU | less than ± 0.6 x 10-5 or lower |
| Photo Diode Array (PDA) Detector | PDA detector Drift in AU/Hour | less than 1 x 10-3 |
| Photo Diode Array (PDA) Detector | Temperature controlled flow cell in C | NA if not provided |
| Photo Diode Array (PDA) Detector | Flow cell | Flow cell |
| Warranty & Reports | Warranty (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) | 2, 3, 4, 5 Or higher |
| Warranty & Reports | Number of installations in Central /State/PSU Govt Hospitals | <3, >3 |
| Warranty & Reports | Availability of toll free facility for technical support maintained by OEM or authorized agencies | Yes Or higher |
Discover companies most likely to bid on this tender
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover
Additional Doc 1 (Requested in ATC)
Compliance of BoQ specification and supporting document *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
Sign up now to access all documents
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS