Bid Publish Date
21-Aug-2026, 5:19 pm
Bid End Date
05-Sep-2026, 10:00 am
Value
₹2,50,000
Location
Progress
Kamdhenu University invites bids for SPECTROPHOTOMETER (Q2) in SABARKANTHA, GUJARAT. Quantity: 2. Submission Deadline: 05-09-2026 10: 00: 00. Submit your proposal before the deadline.
Quantity
2
Bid Type
Two Packet Bid
Bid Validity
180 (Days)
Bid Type
Service
Evaluation
Total value wise evaluation
Inspection Required
No
RA Qualification Rule
H1-Highest Priced Bid Elimination
Tech Clarification Time
2 Days
EMD Required
No
MII Purchase Preference
Yes
MII Preference Band
L1+20%
MSE Purchase Preference
Yes
MSE Preference Band
L1+20%
MSE Exemption/Relaxation
No
Startup Exemption/Relaxation
Yes
Bid Splitting Applied
No
OEM Avg. Turnover
5
Past Performance
80 %
Experience Required
3 Year (s)
Arbitration Clause
No
Mediation Clause
No
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
Yes
Item Category
SPECTROPHOTOMETER (Q2)
Advisory Bank
HDFC Bank
ePBG Percentage
5%
ePBG Duration (Months)
14
Max Delivery Days
15
Delivery Locations
1
Delivery Cities
Sabar Kantha
Delivery Pincodes
383010
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Patel Madhuribahen Ratishkumar | 383010,College of Veterinary Science and AH Himatnagar, Rajpur(Nava), Ta. Himatnagar | Sabar Kantha | Gujarat | 383010 | 2 | 15 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS
Discover companies most likely to bid on this tender
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover
Additional Doc 1 (Requested in ATC)
Additional Doc 2 (Requested in ATC) *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
| Category | Specification | Requirement |
|---|---|---|
| LIGHT SOURCE | Light Source | Deuterium Lamp, Tungsten Halogen Lamp |
| LIGHT SOURCE | Diffraction Grating (lines/mm) | 1200 |
| LIGHT SOURCE | Switching of Light Source | Automatic |
| LIGHT SOURCE | Wavelength Setting And Scanning | Automatic |
| TYPE | Type of Spectrophotometer | UV - VISIBLE |
| TYPE | Type of Optical System | Single Beam |
| TYPE | Monochromator Type | Holographic Grating |
| TYPE | Detector Type | Photodiode |
| TYPE | Spectral Bandwidth Type | Fixed Bandwidth |
| OPTICAL PERFORMANCE | Source Wavelength - Minimum | 200.0 |
| OPTICAL PERFORMANCE | Source Wavelength - Maximum | 1100, 1500, 2000, 2500, 2700, 3000, 3200, 3300, 1650 Or higher |
| OPTICAL PERFORMANCE | Resolution | 0.002, 0.3, 0.5, 1, 1.25, 1.5, 1.7, 2, 0.2, 0.01 |
| OPTICAL PERFORMANCE | Wavelength Accuracy (+/-) | 0.5 |
| OPTICAL PERFORMANCE | Bandwidth (nm) | 0.1, 0.2, 0.5, 1, 1.5, 1.8, 2.0, 4.0, 5.0, 0.1 to 5.0, 7, 8.5 |
| OPTICAL PERFORMANCE | If variable band width, Specify the incremental step value(mm) | 0.1, 1.0, NA |
| OPTICAL PERFORMANCE | Focal length (mm) | 0 to 49, 50 to 99, 100 to 149, 150 to 199, 200 to 249, 250 to 299, 300 and above |
| OPTICAL PERFORMANCE | Photometric measurement modes | Transmittance, Absorbance, Concentration (Wavelength, time) |
| OPTICAL PERFORMANCE | Photometric Absorbance (Max) (Abs) (+/-) | 0.2 - 8 |
| OPTICAL PERFORMANCE | Photometric Absorbance Accuracy (Max) (+/-) | 0.001 - 0.5 |
| OPTICAL PERFORMANCE | Photometric Transmittance (Max) (+/-) | 100.0 |
| OPTICAL PERFORMANCE | Photometric Transmittance Accuracy (Max) (+/-) | 0 - 0.5 |
| OPTICAL PERFORMANCE | Photometric Reflectance (Max) (+/-) | 0 - 100 |
| OPTICAL PERFORMANCE | Stray Light Correction | Automatic |
| OPTICAL PERFORMANCE | Baseline Correction | Automatic |
| GENERIC | Display | In-built display |
| GENERIC | Display type | LCD with Backlight |
| GENERIC | Display Size | 0 - 30 |
| GENERIC | Test result printing | Through external printer |
| GENERIC | If inbuilt printer, Type of printer | N/A |
| GENERIC | Suitable software for connecting PC & Printer | Yes |
| GENERIC | Connectivity | USB |
| GENERIC | Internal storage(GB) | 0 - 80 |
| GENERIC | Sample switching/selection | Automatic |
| GENERIC | Capacity of Sample holder / Sample Carousel | 2, 4 |
| GENERIC | Mximum Path Length of sample | 10.0 |
| GENERIC | Number of Cuvettes supplied-inclusive in the scope of supply | 2 |
| GENERIC | Power Supply | 230 V ± 10, 50 Hz |
| GENERIC | Spares and Consumables inclusive in the scope of supply | No, Yes |
| OPERATING CONDITIONS | Minimum Operating Temperature | 5.0 |
| OPERATING CONDITIONS | Maximum Operating Temperature | 45.0 |
| OPERATING CONDITIONS | Operating Humidity (RH) (%) at 40 degree C | 80.0 |
| WARRANTY | Onsite Warranty | 1.0 |
| WARRANTY | Availability of Spares (at extra cost) | 2, 3, 4, 5, 6, 7, 8, 10 |
| Certification | Availability of UL/CE Certification as per EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control, and laboratory use General requirements | Yes, No |
| Certification | Conformance to EMC/EMI as per EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use | Yes, No |
| Certification | Availability of Test Reports from Central Govt / NABL approved / ILAC acredited lab to prove conformity to the specification | Yes, No |
| Certification | If yes test report to furnished,to the buyer on demand | N/A |