A tender has been published for University Of Kashmir Automated HPLC System for Separation of complex mixtures (Q3) in SRINAGAR, JAMMU & KASHMIR. Quantity: 1 by. Submission Deadline: 11-08-2026 18: 00: 00. Check eligibility and apply.
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
Yes
Item Category
Automated HPLC System for Separation of complex mixtures (Q3)
Advisory Bank
HDFC Bank
Payment Timelines
Payments shall be made to the Seller within 45 days of issue of consignee receipt-cum-acceptance certificate (CRAC) and on-line submission of bills (This is in supersession of 10 days time as provided in clause 12 of GeM GTC)
Max Delivery Days
45
Delivery Locations
1
Delivery Cities
Srinagar
Delivery Pincodes
190006
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Gulzar Ahmad Bhat | 190006,University of Kashmir,Hazratbal,Srinagar | Srinagar | Jammu And Kashmir | 190006 | 1 | 45 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS
| Category | Specification | Requirement |
|---|---|---|
| Standard parameter | Instrument calibration standards | IEC 17025, NA |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Components of unit | Quaternary Low-Pressure Gradient pump& Parallel Double Plunger, Quaternary Low Pressure Gradient pump & Serial Double Plunger |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Number of solvents delivery | 4 |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Flow rate setting range in ml/min | 0.001 to 10 |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Flow rate accuracy | ± 1 %, ± 0.1%, ± 1 % or better |
| UV-Vis Detector | UV-Vis Detector provided | Yes, No Or higher |
| UV-Vis Detector | Light source | Deuterium lamp, NA if not provided, Tungsten Lamp |
| UV-Vis Detector | Spectrum Band width in nm | 4, 5, 6, 8, NA if not provided |
| UV-Vis Detector | Wavelength range in nm | 190 - 700, NA if not provided |
| UV-Vis Detector | Sampling Rate in Hz | 80, 90, 100, NA if not provided |
| UV-Vis Detector | The flow cell must be temperature Controlled | Yes, NA if not provided |
| UV-Vis Detector | Wavelength accuracy in nm | ± 1, NA if not provided |
| UV-Vis Detector | Wavelength reproducibility in nm | ± 0.1, NA if not provided |
| UV-Vis Detector | Drift in AU/Hour | less than 1 x 10-4, NA if not provided |
| UV-Vis Detector | Noise level in AU | ± 2.5 x 10-6, NA if not provided, ±5 x 10-6 |
| UV-Vis Detector | Ability to monitor and quantitative 2 wavelengths simultaneously | Yes, NA if not provided |
| UV-Vis Detector | Linearity in AU | 2.5 or more, NA if not provided |
| UV-Vis Detector | Temperature control provided for Flow cell | Yes, NA if not provided |
| UV-Vis Detector | Temperature control range for flow cell in Degree Celsius | 5 degree above room temp to 50 degree, NA if not provided |
| UV-Vis Detector | flow cell path length in mm | 10, NA if not provided |
| UV-Vis Detector | Flow cell volume in ?l | 10 - 15, NA if not provided, 16 - 20 |
| Auto Sample Injector with cooler | Injection Method | Total volume sample injection |
| Auto Sample Injector with cooler | Carry over of sample | 0.005% or lower, 0.025% or lower |
| Auto Sample Injector with cooler | Injection precision in RSD | less than 0.20 % RSD, less than 0.50 % RSD |
| Software | Operating Windows | Win 7 or latest, Win 10 or latest |
| Photo Diode Array (PDA) Detector | Provision of Photo diode array | Yes, No Or higher |
| Photo Diode Array (PDA) Detector | PDA detector Wavelength range in nm | 190- 800 or better, NA if not provided |
| Photo Diode Array (PDA) Detector | Number of Diodes | 512, 1024, NA if not provided |
| Photo Diode Array (PDA) Detector | PDA detector light source | Deuterium lamp and Tungsten lamp, NA if not provided, Deuterium lamp, Tungsten Lamp |
| Photo Diode Array (PDA) Detector | PDA detector Wavelength accuracy in nm | ± 1, NA if not provided |
| Photo Diode Array (PDA) Detector | Data Rate in Hz | 100, NA if not provided, <100 |
| Photo Diode Array (PDA) Detector | Noise in AU | less than ± 0.6 x 10-5 or lower, NA if not provided, less than ± 1 x 10-5 or lower |
| Photo Diode Array (PDA) Detector | PDA detector Drift in AU/Hour | less than 1 x 10-4, NA if not provided, less than 1 x 10-3 |
| Photo Diode Array (PDA) Detector | Temperature controlled flow cell in C | 5 degree above room temp to 50 degree, NA if not provided |
| Photo Diode Array (PDA) Detector | Flow cell | Flow cell, NA if not provided |
| Warranty & Reports | Warranty (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) | 2, 3, 4, 5, 1 Or higher |
| Warranty & Reports | Number of installations in Central /State/PSU Govt Hospitals | 0, <3, >3 |
| Warranty & Reports | Availability of toll free facility for technical support maintained by OEM or authorized agencies | Yes, No Or higher |
Discover companies most likely to bid on this tender
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover
Additional Doc 1 (Requested in ATC)
Additional Doc 2 (Requested in ATC) *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
Sign up now to access all documents
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS