TenderDekho Logo
GEM

Tender Notice for Digital Radiography System (V2) in VARANASI, UTTAR PRADESH

Bid Publish Date

17-Jun-2026, 12:26 pm

Bid End Date

08-Jul-2026, 10:00 am

EMD

₹95,454

Progress

Issue17-Jun-2026, 12:26 pm
AwardPending
Explore all 6 tabs to view complete tender details

Quantity

1

Bid Type

Single Packet Bid

Categories 12

North Eastern Railway has released a public tender for Digital Radiography System (V2) (Q2) in VARANASI, UTTAR PRADESH. Quantity: 1. Submission Deadline: 08-07-2026 10: 00: 00. Download documents and apply online.

Additional Tender Data

Commercial Details

Tender Category

Goods

Bid To RA

No

Bid To RA Enabled

No

Item Category

Digital Radiography System (V2) (Q2)

Advisory Bank

State Bank of India

ePBG Percentage

5%

ePBG Duration (Months)

62

Delivery Details

Max Delivery Days

90

Delivery Locations

1

Delivery Cities

Varanasi

Delivery Pincodes

221002

Delivery Locations

ConsigneeAddressCityStatePincodeQuantityDelivery DaysAdditional Requirement
Anand Kumar Singh221002,Divisional Railway Hospital, N.E.Railway, VaranasiVaranasiUttar Pradesh221002190-

Authority Records

MINISTRY OF RAILWAYS

BID & GeM Expert Consultancy

End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.

Bid Preparation GeM Registration Document Filing

Free consultation · 24h response

Documents 4

GeM-Bidding-9471574.pdf

Main Document

GeM Category Specification

GEM_CATEGORY_SPECIFICATION

Other Documents

OTHER

GEM General Terms and Conditions Document

GEM_GENERAL_TERMS_AND_CONDITIONS

Technical Specifications 1 Item

Item #1 Details

View Catalog
Category Specification Requirement
General Product Digital Radiography System
General Description Digital Radiographic system with flat panel detector
Table and Table top parameters Type of table Floating Top Type
Table and Table top parameters Type of table movement Manual
Table and Table top parameters Number of movements of Table 6 Way
Table and Table top parameters Longitudinal Movement (with tolerance ± 10%) ±50
Table and Table top parameters Lateral movement (with tolerance ± 10%) ±15
Table and Table top parameters Length x Width (± 5%) in cm 210 x 70
Table and Table top parameters X Ray density < 1 mm A I
Table and Table top parameters Minimum and maximum adjustable heights (± 10%) 55-85
Table and Table top parameters Load bearing capacity of table in kg 200 or more
Table Bucky System Size in inches 17x17
Table Bucky System Grid Ratios 10:1 or higher
Table Bucky System Lines in mm >=92
Table Bucky System Longitudinal travel Manual
Upright Bucky Type Motorized
Upright Bucky Sizes in inches 17x17
Upright Bucky A E C 3 Field or higher
Upright Bucky Movable grid with Grid Ratios 10:1 or higher
Upright Bucky Line in inches >=92
Digital Detector Radiography Mode Digital
Digital Detector Number of detectors Dual
Digital Detector Detector Material Cesium Iodide
Digital Detector Detector Size in inches 17 X 17 and 17 x 17
Digital Detector Matrix size 3kx3k
Digital Detector Pixel size in µm <=175
Digital Detector Grey Scale resolution, bits/pixel 14 or better
Digital Detector DQE at 0 LP/mm 65% or more
Control panel for image acquisition Image preview time in seconds <=8
Control panel for image acquisition Control panel Microprocessor controlled
Control panel for image acquisition Automatic parameter selection Yes
Control panel for image acquisition Control panel entry Alphanumeric key board
Control panel for image acquisition Grid removable Yes
Control panel for image acquisition Anatomic specific post processing Yes
X Ray Generator and Tubes High Frequency X-ray Generator power in KW 65 KW
X Ray Generator and Tubes Maximum mA at 100 KVP 800
X Ray Generator and Tubes Focal spot 0.6 and 1.2
X Ray Generator and Tubes Heat Capacity, KHU 300-400
X Ray Generator and Tubes Control panel and X Ray tube Yes
X Ray Generator and Tubes Post Processing Features Zoom, contrast and brightness, adjustment, panning annotate, mark and reporting
X Ray Generator and Tubes Automatic tube positioning Yes
Tube suspension Tube mounted X Ray control Yes
Tube suspension Type of suspension Floor Mounted
Tube suspension Collimator Type Manual
Workstation Monitor size in inches 19
Workstation Processor i 5
Workstation RAM in GB 4
Workstation HDD 1 TB
Workstation DiCOM compatibility of work station Yes
Workstation Work station should be fully network ready and it should transfer images and patient data from and to hospital network using LAN connectivity or wireless LAN Yes
Workstation Work station integrated with system Yes
Workstation Work station with advance measurement tool should be provided Yes
Workstation DICOM Printing software to be provided Yes
Workstation Patient study on film Yes
Accessories Hand grips Yes
Accessories Foot switches for adjusting longitudinal/side to side movements, locking, and light adjustment Yes
Accessories Patient fixing belts and compression device (for performing excretory urography, etc) Yes
Accessories Integration DICOM 3.0 storage SOP Classes
Accessories Lead Aprons - Full body and Skirt type 2 no. each (0.5pb)
Accessories Lead glass window provided Yes
Accessories Footsteps for the table provided Yes
Accessories Power 440 V 50 HZ 3 Phase 16 Ampls socket
Accessories Voltage stablizer servo controlled in KVA >=50
Certifications Compliance to Medical Device Rules (MDR) 2017 as amended till date Yes
Certifications Availability of valid Medical Device license for the product issued from the competent authority defined under Drugs and Cosmetic Act 1940 and Rules made there under as amended till date Yes
Certifications Certification for manufacturing unit ISO:13485 (Latest)
Certifications Availability of Test Report for each supplied batch/product as per Medical Device Rule (MDR) 2017 as amended till date Yes
Certifications Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission or along with supplies as per buyer requirement Yes
Certifications Conformity to Safety Standards IEC 60601-2-54 or equivalent BIS
Certifications AERB Type Approved Equipment Yes
Certifications Conformity to Indian Standard IS 7620- Part 1, Part 2, Part 3 Yes
Warranty Warranty in Years 5 Or higher
Miscellaneous Parameters OEM/Reseller ( if supplied to reseller shall ensure uninterputed availabilty of all spares for 10 years) Yes
Miscellaneous Parameters Availability of toll free facility for technical support maintained by OEM or authorized agencies Yes
Miscellaneous Parameters Availability of toll free facility for technical support maintened by OEM or authorized agencies Yes
Miscellaneous Parameters User/Technical/Maintenance manuals to be supplied in English in hard and soft copy Yes
Miscellaneous Parameters Demonstration of equipment and training to be provided after completing supplies before acceptance Yes
Miscellaneous Parameters Copies of reports and certifications to be furnished to buyer on demand at time of supplies Yes
Miscellaneous Parameters QA, QC and other tests as per NABH shall be carried out by OEM or authorized agency of OEM Yes
Miscellaneous Parameters Copies of all calibration and test reports to be submitted at the time of Supply of Machine Yes

🤖 AI-Powered Bidder Prediction

Discover companies most likely to bid on this tender

Live AI
Historical Data

Required Documents

1

Certificate (Requested in ATC)

2

OEM Authorization Certificate *In case any bidder is seeking exemption from Experience / Turnover Criteria

3

the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer