Public procurement opportunity for State Council Of Educational Research And Training Automated HPLC System for Separation of complex mixtures (Q3) in North Goa, GOA. Quantity: 1 issued by. Submission Deadline: 12-09-2026 17: 00: 00. View full details and respond.
Quantity
1
Bid Type
Two Packet Bid
Bid Validity
90 (Days)
Bid Type
Service
Evaluation
Total value wise evaluation
Inspection Required
No
Tech Clarification Time
2 Days
EMD Required
Yes
MII Purchase Preference
No
MSE Purchase Preference
No
MSE Exemption/Relaxation
No
Startup Exemption/Relaxation
No
Bid Splitting Applied
No
Min. Avg. Annual Turnover
14
Past Performance
80 %
Experience Required
3 Year (s)
Arbitration Clause
No
Mediation Clause
No
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
No
Item Category
Automated HPLC System for Separation of complex mixtures (Q3)
Advisory Bank
State Bank of India
ePBG Percentage
1%
ePBG Duration (Months)
12
Max Delivery Days
30
Delivery Locations
1
Delivery Cities
North Goa
Delivery Pincodes
403521
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Suyash Bapu Dhuri | 403521,State Council of Educational Research & Training (SCERT), Near GCA Cricket Ground, Porvorim, Goa | North Goa | Goa | 403521 | 1 | 30 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS
Discover companies most likely to bid on this tender
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
Additional Doc 1 (Requested in ATC)
Additional Doc 2 (Requested in ATC)
Compliance of BoQ specification and supporting document *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
| Category | Specification | Requirement |
|---|---|---|
| Standard parameter | Instrument calibration standards | NA |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Components of unit | Quaternary Low-Pressure Gradient pump& Parallel Double Plunger |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Number of solvents delivery | 4 |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Flow rate setting range in ml/min | 0.001 to 10 |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Flow rate accuracy | ± 1 %, ± 1 % or better |
| UV-Vis Detector | UV-Vis Detector provided | Yes Or higher |
| UV-Vis Detector | Light source | Deuterium lamp |
| UV-Vis Detector | Spectrum Band width in nm | 4, 8 |
| UV-Vis Detector | Wavelength range in nm | 190 - 700 |
| UV-Vis Detector | Sampling Rate in Hz | 100 |
| UV-Vis Detector | The flow cell must be temperature Controlled | NA if not provided |
| UV-Vis Detector | Wavelength accuracy in nm | ± 1 |
| UV-Vis Detector | Wavelength reproducibility in nm | ± 0.1 |
| UV-Vis Detector | Drift in AU/Hour | less than 1 x 10-4 |
| UV-Vis Detector | Noise level in AU | ± 2.5 x 10-6 |
| UV-Vis Detector | Ability to monitor and quantitative 2 wavelengths simultaneously | Yes |
| UV-Vis Detector | Linearity in AU | 2.5 or more |
| UV-Vis Detector | Temperature control provided for Flow cell | NA if not provided |
| UV-Vis Detector | Temperature control range for flow cell in Degree Celsius | NA if not provided |
| UV-Vis Detector | flow cell path length in mm | 10 |
| UV-Vis Detector | Flow cell volume in ?l | 10 - 15 |
| Auto Sample Injector with cooler | Injection Method | Total volume sample injection |
| Auto Sample Injector with cooler | Carry over of sample | 0.025% or lower |
| Auto Sample Injector with cooler | Injection precision in RSD | less than 0.20 % RSD |
| Software | Operating Windows | Win 10 or latest |
| Photo Diode Array (PDA) Detector | Provision of Photo diode array | Yes Or higher |
| Photo Diode Array (PDA) Detector | PDA detector Wavelength range in nm | 190- 800 or better |
| Photo Diode Array (PDA) Detector | Number of Diodes | 512 |
| Photo Diode Array (PDA) Detector | PDA detector light source | Deuterium lamp and Tungsten lamp |
| Photo Diode Array (PDA) Detector | PDA detector Wavelength accuracy in nm | ± 1 |
| Photo Diode Array (PDA) Detector | Data Rate in Hz | 100, <100 |
| Photo Diode Array (PDA) Detector | Noise in AU | less than ± 0.6 x 10-5 or lower |
| Photo Diode Array (PDA) Detector | PDA detector Drift in AU/Hour | less than 1 x 10-3 |
| Photo Diode Array (PDA) Detector | Temperature controlled flow cell in C | NA if not provided |
| Photo Diode Array (PDA) Detector | Flow cell | Flow cell |
| Warranty & Reports | Warranty (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) | 2, 3, 4, 5, 1 Or higher |
| Warranty & Reports | Number of installations in Central /State/PSU Govt Hospitals | <3, >3 |
| Warranty & Reports | Availability of toll free facility for technical support maintained by OEM or authorized agencies | Yes Or higher |