Progress
Bank Of India invites bids for Gold purity testing meter/ Karat meter (Q2) in DHARWARD, KARNATAKA. Quantity: 1. Submission Deadline: 18-10-2025 13: 00: 00. Submit your proposal before the deadline.
| S.No | Seller | Item | Date | Status |
|---|---|---|---|---|
| 1 | ACZET PRIVATE LIMITED Under PMA | Make : aczet Model : Stark | 16-10-2025 18:11:03 | |
| 2 | ASHLYN CHEMUNNOOR INSTRUMENTS PRIVATE LIMITED Under PMA | Make : redlands Model : EDX 600 | 17-10-2025 08:33:16 |
| Rank | Seller | Price | Item |
|---|---|---|---|
| L1 | ASHLYN CHEMUNNOOR INSTRUMENTS PRIVATE LIMITED(MSE,MII)( MSE Social Category:General ) Under PMA | Item Categories : Gold purity testing meter/ Karat meter |
Quantity
1
Bid Type
Two Packet Bid
Bid Validity
30 (Days)
Bid Type
Service
Evaluation
Total value wise evaluation
Inspection Required
No
Tech Clarification Time
2 Days
Min. Avg. Annual Turnover
3
OEM Avg. Turnover
26
Past Performance
40 %
Experience Required
3 Year (s)
Arbitration Clause
No
Mediation Clause
No
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
No
Item Category
Gold purity testing meter/ Karat meter (Q2)
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
GEM_GENERAL_TERMS_AND_CONDITIONS
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover
Additional Doc 1 (Requested in ATC)
Compliance of BoQ specification and supporting document *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
| Category | Specification | Requirement |
|---|---|---|
| Source/ Detector | X-ray source Voltage | 50 KV Max |
| Source/ Detector | Detector Type | Gas filled proportional counter |
| Type/ Application | Intended Application | Analysis of gold content in the ornaments/Jewelleries/bullions specimens / samples, and to detect various impurities/adulterants such as copper, silver etc. for assaying gold purity by financial instituions and banks etc. |
| Type/ Application | Construction | The equipment consist of aesthetically design and safe to use enclosure, x-ray generating tube/source , high voltage power supply module for x-ray source, detector, suitable processing and displaying unit, etc. |
| Performance/ Features | Major Element Analysis | Au, Ag, Cu |
| Performance/ Features | Accuracy | 0.1% or better for Au, 0.01% for Au Or higher |
| Performance/ Features | Other performace/ features | Auto-calibration |
| Generic | Dimension of Inner cavity/ chamber (LXWXH) | 300 mm X 300 mm X 250 mm or better |
| Generic | Testing time[User Selectable] (in sec) | 60 sec-180 sec or better |
| DISPLAY / DATA OUTPUT | Data Display unit(inclusive in the scope of Supply) | Suitable monitor (LCD/LED 14 inch or better ) |
| DISPLAY / DATA OUTPUT | Data Processing Unit (inclusive in the scope of Supply) | Suitable data processing unit (PC/all in one etc.) |
| DISPLAY / DATA OUTPUT | Software | Suitable software having processing and reporting features and exporting reports in Excel or CSV formats. |
| TEST REPORT & Certification | Compliance | AERB (Atomoic Energy Regulatory Board) Certified |
| TEST REPORT & Certification | Availability of Test Report from type of Lab, to prove the conformity of product to the specification | OEM |