TenderDekho Logo
Active GEM

Western Railway Cardiac TMT Machine Tender Vadodara Gujarat 2026 - ISI/ MDR Compliant, Windows OS, 3KVA Stabilizer

Bid Publish Date

12-Aug-2026, 7:37 pm

Bid End Date

31-Aug-2026, 10:00 am

Progress

Issue12-Aug-2026, 7:37 pm
AwardPending

Tender Overview

Organization: Western Railway (Indian Railways) in Vadodara, Gujarat. Procurement: Cardiac TMT Machine (V2) with extensive software and hardware integration for exercise stress testing, ECG data capture, and advanced analysis. Target scope includes a full system with treadmill, computer system, data transmission, and accessories. Estimated value and EMD details are not disclosed in the provided data; emphasis is on turnkey installation, demonstration, and training at the consignee location. A unique differentiator is the broad suite of system software features, including ST analysis, beat-by-beat review, and comprehensive reporting, along with MDR/medical device compliance and long-term spares availability. This procurement targets a medically certified, vendor-supported, in-hospital cardio lab solution tailored to Indian Railways facilities.

Technical Specifications & Requirements

  • Treadmill specifications cover emergency stop, zero start, fold-down handrails, remote keyboard control, auto-protocol operation, and a power-backup compatible with a 3 KVA voltage stabilizer. Maximum patient load, speed range, incline, belt quality, and self-calibration features are specified to ensure diagnostic reliability.
  • Computer system features demand a touch-screen monitor, RAM, CPU, OS licensing, CD-ROM, serial/USB ports, keyboard, mouse, embedded UPS (3 minutes), and laser/thermal printers with A4 printing capability and duplex options.
  • Data transmission and signal processing emphasize advanced noise suppression, source consistency filtering, artifact elimination, and real-time ST analysis with beat-by-beat review, supporting multiple protocols (Bruce, Balke, Ellestad, Naughton) and user-defined protocols.
  • System software features include ECG resting/exercise interpretation, data management, auto reports, 12/16 lead support, NIBP integration, Borg scale, and PACEMAKER rhythm detection. Comprehensive reporting with configurable segments, password protection, and multi-user review workflows are requested.
  • Certification & MDR compliance requires MDR 2017 adherence, drug license validity, manufacturing certification, and test reports; OEM service presence in India is mandated along with spare parts availability for 10 years.
  • Accessories & storage cover cables, leads, cuffs in multiple sizes, trolley mounting, and storage for at least 100 patient records with external full-disclosure archival support.

Terms, Conditions & Eligibility

  • MDR compliance, drug license, and manufacturing certification are mandatory.
  • Inspection by nominated agency (RITES or consignee) before/after dispatch with buyer-borne testing charges; pre-dispatch inspection timeline specifies 7 days readiness window.
  • Scope: Bid price must include all cost components for supply; installation, demonstration, and training are required post-supply.
  • OEM direct presence or approved service center in India is required for service continuity and maintenance.
  • Warranty: Catalogue warranty with optional comprehensive warranty selected through bidding; spares availability for 10 years is expected.

Key Specifications

  • Cardiac TMT Machine (V2) with treadmill and integrated computer system

  • Treadmill: emergency stop, zero start, fold-down rails, remote keyboard control, operator-defined protocols

  • Power: compatible with minimum 3 KVA stabilizer

  • Computer system: touch screen, Windows OS licensed, 5+ protocols supported, 16-lead ECG display

  • Data handling: A/D sampling rate, impedance >90 MΩ, CMRR >100 dB, advanced noise reduction

  • Software: resting/exercise ECG, real-time ST analysis, auto report, data storage for at least 100 patient records

  • Accessories: cables/leads, varied cuff sizes, trolley, printer (laser + thermal), network capabilities

  • Compliance: MDR 2017, drug license, manufacturing certification, and test reports

Terms & Conditions

  • MDR 2017 compliance and drug license verified for product

  • Pre-dispatch and post-receipt inspection by nominated agency with specified timelines

  • OEM direct presence or approved Indian service center required for after-sales support

Important Clauses

Payment Terms

Not disclosed in data; payment terms to be confirmed in ATC and contract; expect standard government terms.

Delivery Schedule

Delivery and installation to be completed in agreed period post award; pre-dispatch inspection if applicable.

Penalties/Liquidated Damages

LD provisions not detailed; likely to align with Indian Railways standard L.D. policy for delayed delivery.

Bidder Eligibility

  • Direct OEM presence or approved Indian service center

  • MDR 2017 compliance and drug license validity

  • Experience in supplying/servicing cardio lab equipment or similar diagnostic systems

Tender Data

Bid Details

Quantity

1

Bid Type

Single Packet Bid

Bid Validity

60 (Days)

Bid Type

Service

Evaluation

Total value wise evaluation

Inspection Required

Yes

Tech Clarification Time

4 Days

Preference & Exemptions

EMD Required

No

MII Purchase Preference

Yes

MII Preference Band

L1+20%

MSE Purchase Preference

Yes

MSE Preference Band

L1+15%

MSE Exemption/Relaxation

No

Startup Exemption/Relaxation

No

Bid Splitting Applied

No

Requirements

Arbitration Clause

No

Mediation Clause

No

Additional Tender Data

Commercial Details

Tender Category

Goods

Bid To RA

No

Bid To RA Enabled

No

Item Category

Cardiac TMT Machine (V2) (Q2)

Delivery Details

Max Delivery Days

45

Delivery Locations

1

Delivery Cities

Vadodara

Delivery Pincodes

390004

Delivery Locations

ConsigneeAddressCityStatePincodeQuantityDelivery DaysAdditional Requirement
Desai Himanshu390004,WESTERN RAILWAY DIVISIONAL HOSPITAL, RAILWAY COLONY, PRATAPNAGAR, VADODARA GUJARAT 390004VadodaraGujarat390004145-

Authority Records

MINISTRY OF RAILWAYSWESTERN RAILWAY

BID & GeM Expert Consultancy

End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.

Bid Preparation GeM Registration Document Filing

Free consultation · 24h response

Documents 3

GeM-Bidding-9745593.pdf

Main Document

CATALOG-Specification-1

CATALOG Specification

GEM General Terms and Conditions Document

GEM_GENERAL_TERMS_AND_CONDITIONS

Authority & Contact

Organization

Western Railway

Department

Indian Railways

Office Name

Western Railway

Grievance Contact

[email protected]

🤖 AI-Powered Bidder Prediction

Discover companies most likely to bid on this tender

Live AI
Historical Data

Required Documents

1

GST certificate

2

PAN card

3

Experience certificates

4

Financial statements

5

EMD/Security deposit documentation (if applicable)

6

Technical compliance certificates

7

OEM authorizations (if reseller)

8

Manufacturing/ MDR compliance certificates

9

Drug license (as applicable to product)

10

Test reports per MDR 2017

11

Installation, maintenance, and user manuals (English)

12

Certification and license documents for the buyer's inspection needs

Technical Specifications 1 Item

Item #1 Details

View Catalog
Category Specification Requirement
GENERAL FEATURES Product Description Cardiac TMT Machine with Monitor and Printer
GENERAL FEATURES Product components Heavy-duty medical treadmill with online UPS for the whole system, Computer system, printer, Voltage stabilizer and software along with trolley system to mount computer & printer
GENERAL FEATURES Facility, whereby data should be able to be acquired and digitized near the patient, providing better ECG signal Yes
TREADMILL SPECIFICATIONS Treadmill Type Four tier heavy duty noise less medical treadmill with coasters having locking arrangement
TREADMILL SPECIFICATIONS Compatible online UPS for the whole system with minimum 3 hrs back up time Yes
TREADMILL SPECIFICATIONS Stop/start button on treadmill for emergency stop Yes
TREADMILL SPECIFICATIONS Emergency stop button (ESB) must have the ability to be located in the location of choice by the end user Yes
TREADMILL SPECIFICATIONS Tredmill should have zero start Yes
TREADMILL SPECIFICATIONS Automatic operation from the main unit as per test protocol Yes
TREADMILL SPECIFICATIONS Fold down hand rails on treadmill for allowing easy and convenient assemble and installation Yes
TREADMILL SPECIFICATIONS Length of Treadmill Floor (mm) 2001 to 2100mm
TREADMILL SPECIFICATIONS Width of Treadmill Floor (mm) 700 to 800mm
TREADMILL SPECIFICATIONS Length of walking area on treadmill (mm) 1500 to 1600mm
TREADMILL SPECIFICATIONS Width of walking area on treadmill (mm) 500 to 600mm
TREADMILL SPECIFICATIONS Maximum Patient Load Capacity of Treadmill (Kg) 150 Kg or more, 200 Kg or more Or higher
TREADMILL SPECIFICATIONS Speed range of Treadmill 0.1 to 14 km/Hr or higher range
TREADMILL SPECIFICATIONS Inclination (continuously variable) 0 to 25 or more
TREADMILL SPECIFICATIONS Treadmill motor AC Motor
TREADMILL SPECIFICATIONS Power requirements 220 - 240 VAC @ 50-60 Hz
TREADMILL SPECIFICATIONS Double sided polish on walking surface of treadmill for prolonged product life span Yes
TREADMILL SPECIFICATIONS Remote keyboard operation to start, stop and increase or decrease elevation and speed of treadmill Yes
TREADMILL SPECIFICATIONS Self-aligning anti skid conveyor belt Yes
TREADMILL SPECIFICATIONS Communication/Connectivity RS232, USB
TREADMILL SPECIFICATIONS Enhanced tolerance to electrical changes for greater patient safety Yes
TREADMILL SPECIFICATIONS Self-calibrating system for speed Yes
TREADMILL SPECIFICATIONS Inbuilt digital control issues, precise speed and grade control in the system Yes
TREADMILL SPECIFICATIONS Delivers high torque even at low speeds which ensures no lag when the foot strikes the belt Yes
TREADMILL SPECIFICATIONS Provide smooth power and accuracy even at slow speeds Yes
TREADMILL SPECIFICATIONS Normal, slow deceleration and emergency stop modes on treadmill Yes
TREADMILL SPECIFICATIONS Compatible Voltage Stabilizer provided (Minimum 3KVA) Yes
COMPUTER SYSTEM SPECIFICATIONS Type of Monitor LCD, LED Or higher
COMPUTER SYSTEM SPECIFICATIONS Touch Screen Monitor No
COMPUTER SYSTEM SPECIFICATIONS Size of Monitor (inch) 21, 22, 24 Or higher
COMPUTER SYSTEM SPECIFICATIONS RAM size 4 GB, 6 GB, 8 GB Or higher
COMPUTER SYSTEM SPECIFICATIONS Hard disk space 1 TB Or higher
COMPUTER SYSTEM SPECIFICATIONS Processer Intel core i3
COMPUTER SYSTEM SPECIFICATIONS Operating System Windows 10 or Higher
COMPUTER SYSTEM SPECIFICATIONS CD ROM Drive Yes
COMPUTER SYSTEM SPECIFICATIONS Number of serial port 2
COMPUTER SYSTEM SPECIFICATIONS Number of USB Ports 4
COMPUTER SYSTEM SPECIFICATIONS Alphanumeric multimedia PC keyboard with systems dedicated keypad Yes
COMPUTER SYSTEM SPECIFICATIONS Standard PC mouse Yes
COMPUTER SYSTEM SPECIFICATIONS Built in uninterrupted power supply for minimum of 3 minutes Yes
COMPUTER SYSTEM SPECIFICATIONS QRS triggered output Yes
COMPUTER SYSTEM SPECIFICATIONS Parameters Displayed by the system 12 Lead raw ECG, 12 lead ST level & Slops, Enlarged QRS complex with maximum ST level, Exercise time, Target and Max Heart Rate with % of target achievement, Heart Rate & METS trends, NIBP Trends, Speed of tredmill, Protocol Name, Protocol stage, Patient info
COMPUTER SYSTEM SPECIFICATIONS Built in thermal printer with compatible additional laser black and white printer to print report on A4 Size thermal and A4 size ordinary bond papers respectively Yes
COMPUTER SYSTEM SPECIFICATIONS Minimum Printing Speed of laser printer 20 pages / minute
COMPUTER SYSTEM SPECIFICATIONS Minimum Paper tray capacity of laser printer 250 Sheets
COMPUTER SYSTEM SPECIFICATIONS Print Resolution 1200 x 600 dpi
COMPUTER SYSTEM SPECIFICATIONS Duplex Printing No
COMPUTER SYSTEM SPECIFICATIONS Connectivity / interface of Laser printer USB
COMPUTER SYSTEM SPECIFICATIONS Supported paper sizes A4
TRANSMISSION OF DATA Transmission of data available through Network, PDF
TRANSMISSION OF DATA Advanced signal processing algorithms to provide exceptional accuracy in beat detection and noise rejection resulting in superior identification of events Yes
TRANSMISSION OF DATA Unique source consistency filter, reducing noise and baseline artifact while preserving a diagnostic quality ECG signal Yes
TRANSMISSION OF DATA Provision to eliminate artifacts due to respiration, muscle rub and AC interference, baseline wandering without compromising and distortion in ST segment changes Yes
PROTOCOLS Test Protocols include Bruce, Modified Bruce, Balke, Ellestad, Naughton protocols Yes
PROTOCOLS Facility to add user defined protocols Yes
SYSTEM SOFTWARE FEATURES Licensed copy of windows operating system Yes
SYSTEM SOFTWARE FEATURES Provision to record resting ECG & Exercise ECG Yes
SYSTEM SOFTWARE FEATURES Resting ECG software with detailed automatic interpretation statement for all 12 or 16 lead resting ECG should be present Yes
SYSTEM SOFTWARE FEATURES ECG data management software for storage and review Yes
SYSTEM SOFTWARE FEATURES Facility to review, edit and add ECG from full discloser storage post exam Yes
SYSTEM SOFTWARE FEATURES 50mm sweep speed selection for ECG display in the system Yes
SYSTEM SOFTWARE FEATURES Enlarged QRS complexes with superimposition technique Yes
SYSTEM SOFTWARE FEATURES Automatic BP Measurement System integrated with the stress test system Yes
SYSTEM SOFTWARE FEATURES Provision of real time average complexes Yes
SYSTEM SOFTWARE FEATURES Provision of real time ST Analysis in bar graphs Yes
SYSTEM SOFTWARE FEATURES Provision of Re analysis of final summary report Yes
SYSTEM SOFTWARE FEATURES System should be able to present ST levels for 12 average complexes, slope for 12 average complexes and ST Profile Yes
SYSTEM SOFTWARE FEATURES Provision of online ST measurement adjustment Yes
SYSTEM SOFTWARE FEATURES Provision of reanalyzing the possibility of complete exercise test with ST/HR and NIBP trend graphically Yes
SYSTEM SOFTWARE FEATURES SCD predictive tools like T-wave alternans, ST /HR hysteresis analysis Yes
SYSTEM SOFTWARE FEATURES The system should have a Borg scale, symptom and point criteria table Yes
SYSTEM SOFTWARE FEATURES Full disclosure review/playback/scroll back and addition of past ECG events Yes
SYSTEM SOFTWARE FEATURES ST level and slope data for lead and worst case average beat update should be available continuously during the test Yes
SYSTEM SOFTWARE FEATURES Beat by beat review mode to allow full disclosure review during and post exam Yes
SYSTEM SOFTWARE FEATURES QRS signal averaging software, QT dispersion software, XYZ vector-cardiography software provided as standard part of the unit Yes
SYSTEM SOFTWARE FEATURES Facility to do reanalysis of scored ECG by changing line measurement point title ST and J Yes
SYSTEM SOFTWARE FEATURES The system should present comprehensive final report on minute by minute record of ST segment trend Yes
SYSTEM SOFTWARE FEATURES Facility of automatic arrhythmia detection Yes
SYSTEM SOFTWARE FEATURES Auto calibration with digital micro process control in the system and should also be manually operable Yes
SYSTEM SOFTWARE FEATURES Facility of pacemaker rhythm detection Yes
SYSTEM SOFTWARE FEATURES System should be able to view comprehensive ST segment and morphology analysis in an anatomically intelligent format Yes
SYSTEM SOFTWARE FEATURES ST profile, ST elevation, ST depression, heart rate, Mets, NIBP, ST index should be able to be clearly analyzed Yes
SYSTEM SOFTWARE FEATURES The unit should be able to quickly identify anomalies with the dynamic zoom ST display Yes
SYSTEM SOFTWARE FEATURES The system should be able to view ST maps, patented visual anatomical representations of ST deviations in-frontal planes Yes
SYSTEM SOFTWARE FEATURES Monitoring of ST changes with auto comparison of current and reference beats Yes
SYSTEM SOFTWARE FEATURES The system should be able to display the traces incrementally Yes
SYSTEM SOFTWARE FEATURES Provision of updating ECG and ST segment changes and compare them on a pre-exercise and during exercise basis on a high resolution display Yes
SYSTEM SOFTWARE FEATURES Compact digital acquisition module that virtually eliminates non-cardiac electrical noises Yes
SYSTEM SOFTWARE FEATURES Provision of alarm on HR, BP & ST level Yes
SYSTEM SOFTWARE FEATURES Facility to mark the ECG strip to enter comments at any stage Yes
DATA ACQUISITION AND PROCESSING Data Acquisition 12 Leads ECG Simultaneously and Display 3x4, 4x6, 6x12, 12 lead data indicate significant changes
DATA ACQUISITION AND PROCESSING A/D 12 bits or more
DATA ACQUISITION AND PROCESSING Sampling Rate 300/Sec/channel or more
DATA ACQUISITION AND PROCESSING Input impedance should be more than 90M Ohms Yes
DATA ACQUISITION AND PROCESSING CMRR Should be more than 100 dB Yes
DATA ACQUISITION AND PROCESSING Frequency response 0.05Hz to 100Hz
DATA ACQUISITION AND PROCESSING Patient leakage Should be less then 100?A Yes
DATA ACQUISITION AND PROCESSING The system should have baseline correction Yes
DATA ACQUISITION AND PROCESSING Digital filters 50Hz, 20Hz, 35Hz
REPORTING Facility to get system generated auto report Yes
REPORTING Analysis should work on toal of atleast 5 protocols Yes
REPORTING No startup delay in the system ie instant ECG Reports at the push of a button should be available Yes
REPORTING Final report with configurable lead groups Yes
REPORTING Patient report include protocol, Indications, medications, target heart rate reasons for end, symptoms, diagnosis, notes and conclusions with space for personnel and physician electronic signature Yes
REPORTING Examination summary include heart rate/BP/Workload trends page, ST level trends/ST slope trends, Average QRS by stage or by minute, Mets, Maximum predicted heart rate, THR formula selection Yes
REPORTING Ability to select final report segments including patient information, exam summary, rate, BP, work trends, ST level trends, ST slope trends, average QRS and events should be provided in the system Yes
REPORTING Events should include 12 lead ECG analysis for arrhythmias and user added ECGs Yes
REPORTING Provision of customizable system generated summary report to determine the contents of report with data storage facility Yes
REPORTING Provision of Password protection Yes
REPORTING Option to create network review workstations for viewing, editing and report printing of the stress test examinations should be there Yes
ACCESSORIES Standard Accessories, consumables provided 2 sets of 12 lead ECG patient cables, 100 Electrodes, 10 rim A4 Reporting papers
ACCESSORIES Detachable/replaceable cables and leads Yes
ACCESSORIES Minimum warranty(months) on accessories i e lead set 6, 12 Or higher
ACCESSORIES Appropriate sized cuffs of small, medium and large should be provided with the system Yes
ACCESSORIES Trolley system to mount computer & printer Yes
STORAGE OF RECORDS Storage of atleast 100 patients records Yes
STORAGE OF RECORDS Full disclosure archive to external drive Yes
STORAGE OF RECORDS Events/12 lead ECGs/Full disclosure Yes
CERTIFICATION & REPORTS Compliance to Medical Device Rules (MDR) 2017 as amended till date Yes
CERTIFICATION & REPORTS Availability of valid drug license for the product issued from the competent authority defined under Drugs and Cosmetic Act 1940 and Rules made there under as amended till date Yes
CERTIFICATION & REPORTS Manufacturing unit certification ISO:13485 (Latest)
CERTIFICATION & REPORTS Availability of Test Report for product as per Medical Device Rules (MDR) 2017 as amended till date Yes
CERTIFICATION & REPORTS Compliance to Safety Standards IEC-60601-2-25 or equivalent BIS Standard
CERTIFICATION & REPORTS Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission and/or along with supplies as per buyer requirement Yes
WARRANTY Warranty in Years (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) 3, 4, 5 Or higher
MISCELLANEOUS PARAMETERS OEM/Reseller shall ensure uninterrupted availability of all spares for 10 years Yes
MISCELLANEOUS PARAMETERS Availability of toll free facility for technical support maintened by OEM or authorized agencies Yes
MISCELLANEOUS PARAMETERS User/Technical/Maintenance manuals to be supplied in English in hard and soft copy Yes
MISCELLANEOUS PARAMETERS Details of equipments and procedures required for local calibration and routine maintenance to be supplied and advanced maintenance task documentation also to be furnished Yes
MISCELLANEOUS PARAMETERS List of important spares and accessories, with their part numbers to be supplied to the buyer at the time of supplying the equipment Yes
MISCELLANEOUS PARAMETERS Installation and Demonstration of equipment and training to be provided after completing supplies before acceptance Yes
MISCELLANEOUS PARAMETERS Principal Manufacturer must have direct Presence/approved service center In India Yes

Frequently Asked Questions

Key insights about GUJARAT tender market

How to bid for the Cardiac TMT machine tender in Vadodara Gujarat 2026?

Bidders must meet MDR 2017 compliance, provide drug license and manufacturing certification, and submit RSVs including GST, PAN, and experience certificates. The bid should include OEM authorizations, test reports, and installation/demo plan. Ensure OEM direct presence in India and readiness for pre/post-dispatch inspections by RITES or consignee.

What documents are required for the Western Railway cardio lab bid 2026?

Prepare GST certificate, PAN, experience certificates, financial statements, EMD documentation, technical bid documents, OEM authorization, MDR compliance certificates, drug license, manufacturing certification, test reports, and English operation/maintenance manuals. Include installation/training plan and spare parts list for 10-year availability.

What are the key technical specifications for the treadmill in this tender?

Treadmill must have emergency stop, zero-start, fold-down rails, remote keyboard control, auto-protocol operation, 3 KVA stabilizer compatibility, maximum patient load, programmable speed/incline with self-calibration, and anti-skid belt with digital speed control and back-up power features.

What are the MDR compliance requirements referenced in the tender?

Bidders must provide MDR 2017 compliant product certifications, valid drug license, and manufacturing certifications. Also required are MDR-compliant test reports and ongoing service/support capability from the OEM or authorized Indian service center.

When is pre-dispatch inspection conducted for this tender in India?

Pre-dispatch inspection may be conducted by nominated external agencies or buyer, with a 7-day readiness window after the inspection offer email. Any delays beyond 7 days are regularised without liquidated damages.

What ongoing support and spares availability are expected for the Cardiac TMT system?

OEM/authorized service center must ensure uninterrupted availability of all spares for 10 years. Toll-free technical support must be maintained, and user/maintenance manuals provided in English. Training and installation support are mandatory after supply before acceptance.

What software features must the system provide for ECG analysis?

System must record resting and exercise ECG, offer automatic interpretation for 12/16 leads, real-time ST analysis with ST profile, beat-by-beat review, noise reduction, data management, and configurable final reports with password protection.

What are the data/storage requirements for the system?

System must store at least 100 patient records with full-disclosure archive to external drive, and provide network review workstations for viewing, editing, and printing stress test reports.