Bid Publish Date
11-Aug-2026, 1:19 pm
Bid End Date
22-Aug-2026, 12:00 pm
EMD
₹40,000
Location
Progress
A tender has been published for Dr. Y S Parmar University Of Horticulture And Forestry, Nauni-solan Automated HPLC System for Separation of complex mixtures (Q3) in SOLAN, HIMACHAL PRADESH. Quantity: 1 by. Submission Deadline: 22-08-2026 12: 00: 00. Check eligibility and apply.
Quantity
1
Bid Type
Two Packet Bid
Bid Validity
60 (Days)
Bid Type
Service
Evaluation
Total value wise evaluation
Inspection Required
No
Tech Clarification Time
2 Days
EMD Required
Yes
MII Purchase Preference
No
MSE Purchase Preference
Yes
MSE Preference Band
L1+15%
MSE Exemption/Relaxation
Yes
Startup Exemption/Relaxation
Yes
Bid Splitting Applied
No
Past Performance
10 %
Warranty Period
1 years
Experience Required
3 Year (s)
Arbitration Clause
No
Mediation Clause
No
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
No
Item Category
Automated HPLC System for Separation of complex mixtures (Q3)
Advisory Bank
State Bank of India
ePBG Percentage
2%
ePBG Duration (Months)
3
Max Delivery Days
15
Delivery Locations
1
Delivery Cities
Solan
Delivery Pincodes
173230
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Kailash Chander Sharma | 173230,Dean College of Forestry Dr. Y S Parmar University of Horticulture and Forestry Nauni-Solan 173230 | Solan | Himachal Pradesh | 173230 | 1 | 15 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
OTHER
GEM_GENERAL_TERMS_AND_CONDITIONS
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
| Category | Specification | Requirement |
|---|---|---|
| Standard parameter | Instrument calibration standards | IEC 17025 |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Components of unit | Quaternary Low-Pressure Gradient pump& Parallel Double Plunger |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Number of solvents delivery | 4 |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Flow rate setting range in ml/min | 0.001 to 10 |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Flow rate accuracy | ± 1 % |
| UV-Vis Detector | UV-Vis Detector provided | Yes, No Or higher |
| UV-Vis Detector | Light source | NA if not provided |
| UV-Vis Detector | Spectrum Band width in nm | NA if not provided |
| UV-Vis Detector | Wavelength range in nm | NA if not provided |
| UV-Vis Detector | Sampling Rate in Hz | 80, 90 |
| UV-Vis Detector | The flow cell must be temperature Controlled | NA if not provided |
| UV-Vis Detector | Wavelength accuracy in nm | NA if not provided |
| UV-Vis Detector | Wavelength reproducibility in nm | NA if not provided |
| UV-Vis Detector | Drift in AU/Hour | NA if not provided |
| UV-Vis Detector | Noise level in AU | NA if not provided |
| UV-Vis Detector | Ability to monitor and quantitative 2 wavelengths simultaneously | NA if not provided |
| UV-Vis Detector | Linearity in AU | 2.5 or more |
| UV-Vis Detector | Temperature control provided for Flow cell | NA if not provided |
| UV-Vis Detector | Temperature control range for flow cell in Degree Celsius | NA if not provided |
| UV-Vis Detector | flow cell path length in mm | NA if not provided |
| UV-Vis Detector | Flow cell volume in ?l | NA if not provided |
| Auto Sample Injector with cooler | Injection Method | Total volume sample injection |
| Auto Sample Injector with cooler | Carry over of sample | 0.025% or lower |
| Auto Sample Injector with cooler | Injection precision in RSD | less than 0.20 % RSD |
| Software | Operating Windows | Win 7 or latest |
| Photo Diode Array (PDA) Detector | Provision of Photo diode array | Yes Or higher |
| Photo Diode Array (PDA) Detector | PDA detector Wavelength range in nm | 190- 800 or better |
| Photo Diode Array (PDA) Detector | Number of Diodes | 1024 |
| Photo Diode Array (PDA) Detector | PDA detector light source | Deuterium lamp and Tungsten lamp |
| Photo Diode Array (PDA) Detector | PDA detector Wavelength accuracy in nm | ± 1 |
| Photo Diode Array (PDA) Detector | Data Rate in Hz | 100, <100 |
| Photo Diode Array (PDA) Detector | Noise in AU | less than ± 0.6 x 10-5 or lower |
| Photo Diode Array (PDA) Detector | PDA detector Drift in AU/Hour | less than 1 x 10-3 |
| Photo Diode Array (PDA) Detector | Temperature controlled flow cell in C | 5 degree above room temp to 50 degree |
| Photo Diode Array (PDA) Detector | Flow cell | Flow cell |
| Warranty & Reports | Warranty (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) | 3, 4, 5 Or higher |
| Warranty & Reports | Number of installations in Central /State/PSU Govt Hospitals | <3, >3 |
| Warranty & Reports | Availability of toll free facility for technical support maintained by OEM or authorized agencies | Yes Or higher |