ICU Ventilators (V2)
N/a
WEST TRIPURA, TRIPURA
Progress
Quantity
1
Bid Type
Two Packet Bid
N/a has released a public tender for ICU Ventilators (V2) (Q2) in AHMEDABAD, GUJARAT. Quantity: 1. Submission Deadline: 15-01-2026 13: 00: 00. Download documents and apply online.
N/a
WEST TRIPURA, TRIPURA
Directorate Of Medical Education And Training
BIJNOR, UTTAR PRADESH
Central Health Service Ministry Of Health And Family Welfare
NEW DELHI, DELHI
Directorate Of Medical Education And Training
JHANSI, UTTAR PRADESH
Tender Results
Loading results...
| Category | Specification | Requirement |
|---|---|---|
| PRODUCT INFORMATION | Patient Type | Adult & pediatric |
| PRODUCT INFORMATION | Type of technology | Compressor |
| PRODUCT INFORMATION | Tidal volume in ml | (20 ml or less) to (1500 ml or more) |
| PRODUCT INFORMATION | Respiration rate, breaths/min | upto 150 or more |
| PRODUCT INFORMATION | Trigger mechanism | Both pressure and flow |
| PRODUCT INFORMATION | Inspiratory flow rate, L/min | up to 150 or more |
| PRODUCT INFORMATION | Inspiratory pressure, cm H2O | upto 80 or more |
| PRODUCT INFORMATION | IE ratio | 1:10 to 4:1 |
| PRODUCT INFORMATION | Sigh breath function | Yes |
| PRODUCT INFORMATION | Control panel lock | Yes |
| PRODUCT INFORMATION | Facility for double lung ventilation | Yes |
| PATIENT ASSESSMENT TOOLS | Number of loops | 2 loops (P-V, F-V) with facility of saving of 1 Loops for reference |
| PATIENT ASSESSMENT TOOLS | Lung recruitment tools (PV loops) / OLT (Open lung tools) | Yes |
| PATIENT ASSESSMENT TOOLS | Lung mechanics visualization tool | Yes |
| PATIENT ASSESSMENT TOOLS | Capnography/CO2 monitoring | Yes |
| PATIENT ASSESSMENT TOOLS | Esophageal / transpulmonary pressure monitoring | Yes |
| PATIENT ASSESSMENT TOOLS | APRV(Airway Pressure Release Ventilation) /Bivent/Bi level | Yes |
| PATIENT ASSESSMENT TOOLS | MMV(Mandatory Minute Ventilation) + PSV (Pressure Support Ventilation) /ASV( Adaptive Support Ventilation ) or ALPV(Adaptive Lung Protection Ventilation) or Automode | Yes |
| PATIENT ASSESSMENT TOOLS | PRVC(Pressure-Regulated Volume Control ventilation)/Auto flow | Yes |
| PATIENT ASSESSMENT TOOLS | High flow oxygen therapy | Yes |
| PATIENT ASSESSMENT TOOLS | Automatic weaning system provided | Yes |
| PATIENT ASSESSMENT TOOLS | Volume Support | Yes |
| PATIENT ASSESSMENT TOOLS | RSBI (Rapid Shallow Breathing Index) | Yes |
| PATIENT ASSESSMENT TOOLS | Expiratory Time constant ( Tcexp) | Yes |
| PATIENT ASSESSMENT TOOLS | Compliance static and/or dynamic | Yes |
| PATIENT ASSESSMENT TOOLS | Inspired and/or expired resistance | Yes |
| PATIENT ASSESSMENT TOOLS | Occlusion pressure | Yes |
| PATIENT ASSESSMENT TOOLS | Inspiratory and expiratory hold | Yes |
| PATIENT ASSESSMENT TOOLS | Spontaneous frequency | Yes |
| PATIENT ASSESSMENT TOOLS | Total peep, intrinsic peep, extrinsic peep | Yes |
| PATIENT ASSESSMENT TOOLS | Expiratory block is autoclavable and no routine calibration required | Yes |
| PATIENT ASSESSMENT TOOLS | Auto compensation for ET Tube | Yes |
| PATIENT ASSESSMENT TOOLS | Facility for automatic compliance & Leakage compensation for circuit available | Yes |
| PATIENT ASSESSMENT TOOLS | RS 232 out put port | Yes |
| PATIENT ASSESSMENT TOOLS | USB/Memory card | Yes |
| PATIENT ASSESSMENT TOOLS | Availability of Remote alarm/display port | Yes |
| PATIENT ASSESSMENT TOOLS | Facility to Report (vent alarms and patient status) | Yes |
| PATIENT ASSESSMENT TOOLS | HL/7 Compliant out put | Yes |
| PATIENT ASSESSMENT TOOLS | Type of coloured display | LED/LCD/TFT |
| PATIENT ASSESSMENT TOOLS | Display should be touch screen | Yes |
| PATIENT ASSESSMENT TOOLS | NIV (Non Invasive Ventilation) to be possible in all modes of ventilation available | Yes |
| PATIENT ASSESSMENT TOOLS | Minimum Size of display (±2%) | 15.0 |
| PATIENT ASSESSMENT TOOLS | Graphic display have automatic scaling facility for waves | Yes |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Provision of UPS | Yes |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Backup time in hrs | 2 |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Built in air source | Air compressor (external) |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Availability of stand alone compressor | Yes |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Compressor shall provide an oil free Medical air, with Peak output flow should be minimum 150 LPM and Air quality complying with ISO compressed air purity class | Yes |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Seller shall ensure compatibility of compressor with ventilator | Yes |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Nebulizer with particle size less than 3 micron | Yes |
| ADDITIONAL ACCESSORIES | Humidifier: Servo controlled with digital monitoring of inspired gas temperature with heating wire | 1 |
| WARRANTY | Warranty (years) | 5 Or higher |
Discover companies most likely to bid on this tender
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
Extended Deadline
15-Jan-2026, 1:00 pm
Opening Date
15-Jan-2026, 1:30 pm
N/a
📍 LUCKNOW, UTTAR PRADESH
Sign up now to access all documents
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS