High End Multipara Monitor (V3)
N/a
PATAN, GUJARAT
Bid Publish Date
07-Jul-2026, 10:11 am
Bid End Date
28-Jul-2026, 5:00 pm
EMD
₹8,47,470
Location
Progress
Quantity
82
Bid Type
Two Packet Bid
All India Institute Of Medical Sciences (aiims) invites bids for High End Multipara Monitor (V3) (Q2) in NADIA, WEST BENGAL. Quantity: 82. Submission Deadline: 28-07-2026 17: 00: 00. Submit your proposal before the deadline.
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
Yes
Item Category
High End Multipara Monitor (V3) (Q2)
Advisory Bank
ICICI
ePBG Percentage
5%
ePBG Duration (Months)
26
Pre-Bid Date
17-07-2026 15:00:00
Pre-Bid Venue
Welcome Hall , Administration Building, Ground Floor, AIIMS KALYANI
Max Delivery Days
90
Delivery Locations
1
Delivery Cities
Nadia
Delivery Pincodes
741245
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Aditya Kumar Pandit | 741245,NH-34 Connector, Basantapur, Saguna, Kalyani, District Nadia, West Bengal 741245 | Nadia | West Bengal | 741245 | 82 | 90 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
OTHER
OTHER
GEM_CATEGORY_SPECIFICATION
OTHER
OTHER
GEM_GENERAL_TERMS_AND_CONDITIONS
N/a
PATAN, GUJARAT
Central University Of Gujarat
VADODARA, GUJARAT
Directorate Of Medical Education And Training
CHANDAULI, UTTAR PRADESH
N/a
LUCKNOW, UTTAR PRADESH
Ssg Hospital Vadodara
VADODARA, GUJARAT
Tender Results
Loading results...
| Category | Specification | Requirement |
|---|---|---|
| Product Specifications | Type of Monitor | Pre- Configured |
| Product Specifications | Monitoring Parameters | ECG, HR, Respiration Rate, SpO2, Pulse Rate, NIBP, Temperature, IBP |
| Product Specifications | EtCO2 | Yes |
| Product Specifications | Cardiac output | Non invasive (Continuous) |
| Product Specifications | Transport monitor module with basic vital parameters for transporting within the hospital | No |
| Product Specifications | Screen size of transport module( in inches ) | NA |
| Product Specifications | AGM (Anesthetic Gas Module) | Upgradable |
| Product Specifications | NMT (Neuro Muscular Transmission) | Upgradable |
| Product Specifications | EEG (Electroencephalogram) | Upgradable |
| Product Specifications | AEEG (Amplitude Integrated Electroencephalography)Trend Graph | No |
| Product Specifications | Channels in EEG | 8 |
| Product Specifications | Depth of Anesthesia Monitoring | Upgradable |
| Product Specifications | Spirometry | No |
| Product Specifications | Non invasive hemoglobin | No |
| Product Specifications | PVI (Pleth variability index) | No |
| Product Specifications | Perfusion Index | Yes |
| Product Specifications | Transcutaneous PO2 / PCO2 | No |
| Product Specifications | S T Analysis | Yes |
| Product Specifications | PPV (derived value during IBP measurement) | Yes |
| Product Specifications | Regional Oximetry | No |
| Product Specifications | Length of time of full disclosure of patient data in hr | ≥ 72 Or higher |
| Product Specifications | Length of time of trending in hr | ≥ 72 Or higher |
| ECG PARAMETERS | Number of leads of ECG | 3 |
| ECG PARAMETERS | 12-Lead interpretation for ECG | Yes |
| ECG PARAMETERS | ECG Waveform number of leads display | 7 |
| ECG PARAMETERS | Arrhythmia detect | Yes |
| RESPIRATION PARAMETERS | Respiration Method | Impedance pneumography |
| RESPIRATION PARAMETERS | Display waveform of respiration | Yes |
| RESPIRATION PARAMETERS | Respiration rate measurement range | 0 - 150 bpm or more |
| RESPIRATION PARAMETERS | Minimum gradation of respiratory measurement | 1 bpm or better |
| PULSE OXIMETERY PARAMETERS | Measurement technique (Declaration of the measurement technique by the seller at the time of bid submission/supply) | Low Perfusion Technology with Motion artifact detection |
| PULSE OXIMETERY PARAMETERS | Probe type | Clip |
| PULSE OXIMETERY PARAMETERS | Probe for Adult and Pediatric | Reusable |
| PULSE OXIMETERY PARAMETERS | Probe for neonates | Reusable |
| PULSE OXIMETERY PARAMETERS | SPO2 measurement range | 40 - 100% |
| PULSE OXIMETERY PARAMETERS | Accuracy of SPO2 measurement | ± 2% or better |
| PULSE OXIMETERY PARAMETERS | Minimum Gradation | 1% |
| NIBP PARAMETERS | Systolic Blood pressure monitoring range | 40 - 250 mmHg or more |
| NIBP PARAMETERS | Diastolic Blood pressure monitoring range | 30 - 220 mmHg or more |
| TEMPERATURE PARAMETERS | Type of probe | Skin |
| IBP PARAMETERS | Number of channels for IBP | 2 |
| ETCO2 PARAMETERS | Sensor type | Mainstream |
| ETCO2 PARAMETERS | Patient type | Intubated and non intubated |
| ETCO2 PARAMETERS | CO2 Measuring range ( in mmHg ) | 0 to 99 mmHg or better |
| ALARMS | Availability of Audio and visual alarms with high , medium and low intensity adjustment | Yes |
| ALARMS | Operator can set audio visual alarm levels for low or high levels of each parameter independently | Yes |
| ALARMS | Systolic BP - low and high | Yes |
| ALARMS | Diastolic BP - low and high | Yes |
| ALARMS | Mean BP - low and high | Yes |
| ALARMS | Heart rate - low and high | Yes |
| ALARMS | SPO2 - low and high | Yes |
| ALARMS | Respiratory rate - low and high | Yes |
| ALARMS | Temperature - low and high | Yes |
| ALARMS | Alarm Volume Adjustable | Yes |
| ALARMS | Activating silence of alarm | Yes |
| ALARMS | Duration of alarm silence ( in seconds ) | 120 |
| ALARMS | Suspension of all alarms | Yes |
| ALARMS | Latching of alarm | Yes |
| ALARMS | Cuff leakage alarm | Yes |
| ALARMS | Cuff disconnect alarm | Yes |
| ALARMS | Alarm for inflation/deflation errors | Yes |
| ALARMS | Indicator for low-battery | Yes |
| ALARMS | EtCO2 - low and high | Yes |
| ALARMS | Escalation of Alarm Priority | Yes |
| ALARMS | Sensor/ wire/probe disconnection alarm | Yes |
| NETWORK PARAMETERS | Connectivity with Network printer | Yes |
| NETWORK PARAMETERS | Connectivity provision with Central Nursing Station | Yes |
| DISPLAY PARAMETERS | Type of display | LCD/LED/TFT |
| DISPLAY PARAMETERS | Display Size in inches | 12, 15, 17, 19 Or higher |
| DISPLAY PARAMETERS | Traces of display | >12 |
| DISPLAY PARAMETERS | User Interface of display | Touchscreen, Touchscreen and Rotary knob Or higher |
| DISPLAY PARAMETERS | Inbuilt thermal Recorder (Printer) | No |
| DISPLAY PARAMETERS | Should be able to connect to additional display | Yes |
| STANDARD ACCESSORIES | Number of reusable adult probe provided | 1 |
| STANDARD ACCESSORIES | Number of reusable pediatric probe provided | 1 |
| STANDARD ACCESSORIES | Number of reusable neonatal probe provided | NA |
| STANDARD ACCESSORIES | Number of Disposable neonate probes provided | NA |
| STANDARD ACCESSORIES | Number of ECG 5/6/10 lead cable provided | 1 |
| STANDARD ACCESSORIES | Number of Neonatal ECG 3/5/6 lead cable provided | 1 |
| STANDARD ACCESSORIES | Number of IBP cables provided | 2 Reusable Cables |
| STANDARD ACCESSORIES | Number of disposable IBP Transducers provided | 10 |
| STANDARD ACCESSORIES | Number of Adult NIBP cuffs of Medium, Large, Extra large provided | 1 Set |
| STANDARD ACCESSORIES | Number of Pediatric NIBP cuffs of Small and medium provided | 1 Set |
| STANDARD ACCESSORIES | Number of Disposable Neonate NIBP cuffs of 1,2,3,4,5 (in cm) provided | NA |
| STANDARD ACCESSORIES | Number of Skin temperature probe provided | 1 for Adult / Paediatric |
| STANDARD ACCESSORIES | Number of Neonatal Rectal temperature Probe provided | NA |
| STANDARD ACCESSORIES | Number of Esophageal Temperature probe (Adult) provided | 1 |
| STANDARD ACCESSORIES | Number of EtCO2 adapter provided | NA |
| STANDARD ACCESSORIES | Number of EtCO2 Sampling Line provided | NA |
| STANDARD ACCESSORIES | Number of Water trap provided | NA |
| STANDARD ACCESSORIES | Accessories to wall mount the monitor | Yes |
| POWER REQUIREMENTS | Power supply | Single Phase 220-240 VAC, 50Hz |
| POWER REQUIREMENTS | Mains cable | 3 meter or more with Indian plug |
| POWER REQUIREMENTS | Battery | Inbuilt |
| POWER REQUIREMENTS | Minimum Back up time of UPS ( in minutes ) | NA |
| POWER REQUIREMENTS | Minimum Inbuilt Battery back up ( in minutes ) | 30 |
| POWER REQUIREMENTS | Recharging time of battery ( in hr ) | ≤ 6 |
| POWER REQUIREMENTS | Weight of the Machine in kgs | ≤ 15 |
| POWER REQUIREMENTS | Noise level ( in dB ) | ≤ 50 |
| POWER REQUIREMENTS | Device Earthing | Yes |
| CERTIFICATIONS | Compliance to Medical Device Rules (MDR) 2017 as amended till date | Yes |
| CERTIFICATIONS | Availability of valid Medical Device license for the product issued from the competent authority defined under Drugs and Cosmetic Act 1940 and Rules made there under as amended till date | Yes |
| CERTIFICATIONS | Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission or along with supplies as per buyer requirement | Yes |
| CERTIFICATIONS | Conformity to Electrical Safety Standards | IEC 60601 or BIS Equivalent |
Discover companies most likely to bid on this tender
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover
Additional Doc 1 (Requested in ATC) *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
All India Institute Of Medical Sciences (aiims)
📍 JODHPUR, RAJASTHAN
All India Institute Of Medical Sciences (aiims)
📍 NAGPUR, MAHARASHTRA
Directorate Of Medical Education
📍 DIBRUGARH, ASSAM
Directorate Of Medical Education And Training
📍 BASTI, UTTAR PRADESH
Gujarat Medical Education And Research Society (gmers)
📍 VALSAD, GUJARAT
Sign up now to access all documents
Main Document
OTHER
OTHER
GEM_CATEGORY_SPECIFICATION
OTHER
OTHER
GEM_GENERAL_TERMS_AND_CONDITIONS