Ultrasound Machine (V2)
N/a
LUCKNOW, UTTAR PRADESH
Bid Publish Date
10-Jul-2026, 1:05 pm
Bid End Date
31-Jul-2026, 11:00 am
EMD
₹20,000
Progress
Quantity
1
Bid Type
Two Packet Bid
Public procurement opportunity for N/a Ultrasound Machine (V2) (Q2). Quantity: 1 issued by. Submission Deadline: 31-07-2026 11: 00: 00. View full details and respond.
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
No
Item Category
Ultrasound Machine (V2) (Q2)
Advisory Bank
Bank of India
ePBG Percentage
5%
ePBG Duration (Months)
37
Payment Timelines
Payments shall be made to the Seller within 30 days of issue of consignee receipt-cum-acceptance certificate (CRAC) and on-line submission of bills (This is in supersession of 10 days time as provided in clause 12 of GeM GTC)
Max Delivery Days
90
Delivery Locations
1
Delivery Cities
Valsad
Delivery Pincodes
396191
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Chandanikumari Anilbhai Patel | 396191,Government Hospital Vapi, Near Swaminarayan Gurukul, Chala,Vapi | Valsad | Gujarat | 396191 | 1 | 90 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
OTHER
OTHER
OTHER
OTHER
OTHER
GEM_GENERAL_TERMS_AND_CONDITIONS
N/a
LUCKNOW, UTTAR PRADESH
N/a
LUCKNOW, UTTAR PRADESH
N/a
RAJKOT, GUJARAT
N/a
NORTH EAST DELHI, DELHI
N/a
CENTRAL DELHI, DELHI
Tender Results
Loading results...
| Category | Specification | Requirement |
|---|---|---|
| Types of Probes & Features | Applications | Abdomen, Abdominal vascular, Urology, OB/GYN, Neonatal brain, Small parts (breast, thyroid, scrotal, prostate, musculoskeletal) |
| Types of Probes & Features | Convex Probe | 2.5 to 5 MHz (±2) |
| Types of Probes & Features | Microconvex probe | 4 to 8 MHz (±2), Not Provided |
| Types of Probes & Features | Linear array probe | 3 to 12 MHz (±2) |
| Types of Probes & Features | TVS Probe | 4 to 9 MHz (±2) |
| Types of Probes & Features | Cardiac Probe | 1 to 5 MHz (±2), 5 to 9 MHz (±2), Not Provided |
| Types of Probes & Features | Hockey Stick Probe | 6 to 17 MHz (±2), 6 to 13 MHz (±2), 8 to 20 MHz (±2), Not Provided |
| Types of Probes & Features | Volume Convex Probe | 2.5 to 5 MHz (±2) |
| Types of Probes & Features | Volume Linear Probe | 3 to 12 MHz (±2), 6 to 15 MHz (±2), 6 to 18 MHz (±2), 5 to 10 MHz (±2), Not Provided |
| Types of Probes & Features | Volume TVS Probe | 4 to 9 MHz (±2), Not Provided |
| Types of Probes & Features | Endorectal Probe | 5 to 9 MHz (±2), Not Provided |
| Types of Probes & Features | E-Breast Elastography technique | Yes, No |
| Types of Probes & Features | E-Thyroid Elastography technique | Yes, No |
| Types of Probes & Features | Liver Elastography technique | Yes, No |
| Types of Probes & Features | Gel Warmer | Yes, No |
| Types of Probes & Features | Dedicated 3-D/4-D | Yes, No |
| Scan Modes | B-mode | Yes |
| Scan Modes | M-mode | Yes |
| Scan Modes | Contrast harmonic imaging with angio mode | Yes |
| Scan Modes | Provision of Harmonic imaging in Linear Probes | Yes, No |
| Scan Modes | Color Doppler imaging (CDI) | Yes |
| Scan Modes | Color Doppler 3-D/4-D option | Yes, No |
| Scan Modes | Power Doppler imaging (PDI) | Yes |
| Scan Modes | Power Doppler 3-D/4-D option | Yes, No |
| Scan Modes | Continuous wave | Yes |
| Scan Modes | Pulsed wave | Yes |
| Scan Modes | Duplex mode | Yes |
| Scan Modes | Triplex mode | Yes, No |
| Scan Modes | Tissue Doppler imaging | Yes |
| Scan Modes | Contrast-enhanced ultrasound (CEUS) option | Yes |
| Scan Modes | Tissue synchronization, tissue velocity | Yes, No |
| Scan Modes | Gain control in two dimensions for additional level of flexibility for image quality control | Yes |
| Scan Modes | Provision for higher sensitivity for low frequency doppler in all probes | Yes |
| Scan Modes | Number of channels in color doppler system | 50000 or more (if channels are being designated as digital channels), 300000 or more (if channels are being designated as digital channels), 1000000 or more (if channels are being designated as digital channels), 2500000 or more (if channels are being designated as digital channels), 4000000 or more (if channels are being designated as digital channels), 128 or more (if channels are being designated as physical channels) |
| Scan Modes | Live 3 D imaging in color doppler system | Yes, No |
| Scan Modes | Colour Doppler system with all application packages ,Quad loop for serial studies with high frame rate revies, harmonic imaging capability in all modes and integrated stress echo package digital storage and retrieval | Yes |
| Scan Modes | High-definition acoustic zoom provision for enlarging sections of 2D and Colour flow images with more acoustic information for greater clarity and detail while maintaining an optimal frame rate | Yes |
| Scan Modes | Colour Flow Imaging Option | Yes |
| Scan Modes | Provision of Tissue Colorization (B-Colour) for improved contrast resolution | Yes |
| Scan Modes | All application packages should be inbuilt into the system | Yes |
| Scan Modes | Cine Loop Features | Yes |
| Scan Modes | Cine Loop Memory in MB | >=100 |
| Scan Modes | Frame grabber facility for post analysis | Yes, No |
| Scan Modes | Number of transducer/probe ports | 2, 3, 4 or More |
| Scan Modes | System should be capable of generating real time live 3- D images | Yes |
| Scan Modes | Contrast agent imaging | Yes |
| Scan Modes | Color Doppler system with all applications package, quad loop for serial studies with high frame rate, harmonic imaging capability in all modes and digital storage and retrieval | Yes |
| Image Display and Processing | DICOM 3 Compliant | Yes |
| Image Display and Processing | Automated B-mode (2-D) image | Yes, No |
| Image Display and Processing | Automated CDI(Color Doppler Imaging) image optimization | Yes |
| Image Display and Processing | Automated PW Doppler image optimization | Yes, No |
| Image Display and Processing | Dynamic receive focus | Yes |
| Image Display and Processing | Touch Screen | Yes, No |
| Image Display and Processing | Monitor display size (inch) | 21 or more Or higher |
| Image Display and Processing | Split screen | Twin View and Live Quad Screen |
| Image Display and Processing | Single/dual monitors | Dual (image display and touch panel) |
| Image Display and Processing | Max number frames | >=1000 |
| Image Display and Processing | Digital storage hard drive, GB | >=500, >=200 or more |
| Image Display and Processing | DVD/CD Writer/ USB | Yes |
| Image Display and Processing | Type of processor (in PC) | i7 or latest |
| Image Display and Processing | RAM capacity of PC in GB (hint: In case PC not provided put zero) | ≥ 4, NA |
| Image Display and Processing | Processor speed in GHz (PC) | ≥ 3, NA |
| Image Display and Processing | Frame Grabber incorporated(PC) | Yes, No |
| Image Display and Processing | Connectivity between USG and PC | Yes, No |
| Image Display and Processing | PC with all software inclusive interfaced with USG machine | Yes, No |
| Image Display and Processing | Laser Color printer provided | Yes, No |
| Image Display and Processing | Inkjet/Thermal printer provided | Yes, No |
| Image Display and Processing | CD/DVD/USB Produced should be playable on any system | Yes, No |
| Image Display and Processing | Glossy Colour Print Paper supplied | 500 sheets, 1000 sheets, NA |
| Image Display and Processing | Dicom Interfacability Dicom to PACS | Yes, No |
| Image Display and Processing | Interfacability Dicom to RIS | Yes, No |
| Power Requirements | Resettable over current breaker shall be fitted for protection | Yes, No |
| Power Requirements | Power backup | UPS, Inbuilt battery |
| Power Requirements | Back up time in minutes (minimum) | 30, 60, 90, 120 |
| Power Requirements | Suitable Servo Controlled Stabilizer / CVT should be provided | Yes, No |
| Warranty and Maintenance | Warranty in Years (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) | 3, 5 Or higher |
Discover companies most likely to bid on this tender
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover
Additional Doc 1 (Requested in ATC)
Compliance of BoQ specification and supporting document *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
All India Institute Of Medical Sciences (aiims)
📍 RAIPUR, CHHATTISGARH
Western Railway
📍 MUMBAI, MAHARASHTRA
Institute Of Kidney Diseases And Research Centre
📍 AHMEDABAD, GUJARAT
N/a
📍 CHAMBA, HIMACHAL PRADESH
Pimpri Chinchwad Municipal Corporation
📍 PUNE, MAHARASHTRA
Sign up now to access all documents
Main Document
CATALOG Specification
OTHER
OTHER
OTHER
OTHER
OTHER
GEM_GENERAL_TERMS_AND_CONDITIONS