N/a invites bids for Mobile Digital Radiography System (V2) (Q2) in WEST DELHI, DELHI. Quantity: 4. Submission Deadline: 07-11-2025 13: 00: 00. Submit your proposal before the deadline.
Government Tender Published for Mobile Digital Radiography System (V2) in WEST DELHI, DELHI
Posted
28 Oct 2025, 12:53 pm
Deadline
07 Nov 2025, 01:00 pm
EMD
₹2,88,000
Value
₹96,00,000
Location
Progress
Quantity
4
Bid Type
Two Packet Bid
   
Technical Specifications
 1 Item  
 Item #1 Details
View Catalog| Category | Specification | Requirement | 
|---|---|---|
| General | Product Description | Mobile Digital Radiography System | 
| General | Scope of supply | Compact, easily transportable, digital mobile radiographic unit with articulated/telescopic arm | 
| Digital Detector | Radiography Mode | Digital | 
| Digital Detector | Number of detector | 1 | 
| Digital Detector | Detector Material | Cesium Iodide | 
| Digital Detector | Detector Size | 14 inches x 17 inches(wireless ) | 
| Digital Detector | Matrix size | 2kx2k | 
| Digital Detector | Pixel size in µm | <=150 | 
| Digital Detector | Grey Scale resolution, bits/pixel | 14 or better | 
| Digital Detector | DQE at 0 LP/mm | 65% or more | 
| Digital Detector | Machine should have a detector storage compartment | Yes | 
| Digital Detector | Weight of detector in kg | <=5 | 
| Digital Detector | Detector should be designed and calibrated for General Radiography Purposes and must be fully integrated with the mobile unit including the controls | Yes | 
| Digital Detector | Detector storage capacity | 100 images or more | 
| Control panel for image acquisition | Image preview time in seconds | <=6 | 
| Control panel for image acquisition | Control panel | Microprocessor controlled | 
| Control panel for image acquisition | Automatic parameter selection | Yes | 
| Control panel for image acquisition | Control panel entry | Touch screen | 
| Control panel for image acquisition | Anatomic specific post processing | Yes | 
| Control panel for image acquisition | Display | Digital display of mAs and KV and an electronic timer | 
| Control panel for image acquisition | Inbuilt console should have touch TFT screen and control | Yes | 
| Control panel for image acquisition | Console should enable to view the image, and provide post processing features, using touch screen | Yes | 
| Control panel for image acquisition | Post processing features | Zoom, contrast and brightness, adjustment, panning annotate, mark and reporting | 
| X Ray Generator and Tubes | High Frequency X-ray Generator Power in KW | 3.5 KW | 
| X Ray Generator and Tubes | KV range | 40-125 or more | 
| X Ray Generator and Tubes | Maximum mA | 100 @ 40 KVP | 
| X Ray Generator and Tubes | Focal spot, mm | 0.8 - 2 | 
| X Ray Generator and Tubes | Heat Capacity, KHU | 40-100 | 
| X Ray Generator and Tubes | Anode Type | Stationary | 
| X Ray Generator and Tubes | Anode RPM | NA for stationary | 
| X Ray Generator and Tubes | Exposure time in ms | 0.1 to 3 Sec | 
| X Ray Generator and Tubes | Control Panel and X-Ray Tube | Yes | 
| Tube Suspension | X Ray tube output should match with generator | Yes | 
| Tube Suspension | Type of suspension | Articulated arm | 
| Tube Suspension | Collimator Type | Manual | 
| Tube Suspension | Collimator rotation | ± 90 degree with on off timer | 
| Ergonomics | Unit should have small foot print | Yes | 
| Ergonomics | Height of the column in cm (Should facilitate easy transportation in the lift and areas with small height doors etc) | <=150 | 
| Ergonomics | Weight of mobile X Ray unit in kg | <=600 | 
| Ergonomics | Cables should be concealed in the arm system | Yes | 
| Ergonomics | Exposure release switch | Detachable with a cord of at least 5 meters. | 
| Workstation | Inbuilt work station | Yes | 
| Workstation | Monitor size in inches | 17 | 
| Workstation | Dicom compatibility of work station | Yes | 
| Workstation | Inbuilt console should have touch screen control | Yes | 
| Workstation | Number of images which can be stored | 10000 or more | 
| Workstation | Machine should be fully network ready and it should be possible to transfer images and patient data from and to hospital network using LAN connectivity or wireless LAN | Yes | 
| Accessories | Integration | DICOM 3.0 storage SOP Classes | 
| Accessories | DAP Meter provided | No | 
| Accessories | Light weight ‘zero lead’ radiation protection aprons provided | Yes | 
| Accessories | Thyroid shields provided | Yes | 
| Accessories | Gonadsheet Male provided | Yes | 
| Accessories | Gonad sheet Female provided | Yes | 
| Accessories | Extractable measurable tape should be available | Yes | 
| Accessories | Castor wheel | Yes | 
| Accessories | Unit should have effective breaking system for parking | Yes | 
| Accessories | Power Requirements | 230 V 50 HZ Single phase -16 Ampls socket | 
| Accessories | Battery charging time in hours | <=6 | 
| Certifications | Compliance to Medical Device Rules (MDR) 2017 as amended till date | Yes | 
| Certifications | Availability of valid Medical Device license for the product issued from the competent authority defined under Drugs and Cosmetic Act 1940 and Rules made there under as amended till date | Yes | 
| Certifications | Certification for manufacturing unit | ISO:13485 (Latest) | 
| Certifications | Availability of Test Report for each supplied batch/product as per Medical Device Rule (MDR) 2017 as amended till date | Yes | 
| Certifications | Conformity to Safety Standards | IEC 60601-2-54 or equivalent BIS | 
| Certifications | Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission or along with supplies as per buyer requirement | Yes | 
| Certifications | AERB Type Approved Equipment | Yes | 
| Certifications | Conformity to Indian Standard IS 7620- Part 1, Part 2, Part 3 | Yes | 
| Miscellaneous Parameters | OEM/Reseller (if supplied to reseller shall ensure uninterputed availabilty of all spares for 10 years) | Yes | 
| Miscellaneous Parameters | Availability of toll free facility for technical support maintened by OEM or authorized agencies | Yes | 
| Miscellaneous Parameters | User/Technical/Maintenance manuals to be supplied in English in hard and soft copy | Yes | 
| Miscellaneous Parameters | Demonstration of equipment and training to be provided after completing supplies before acceptance | Yes | 
| Miscellaneous Parameters | Copies of reports and certifications to be furnished to buyer on demand at time of supplies | Yes | 
| Miscellaneous Parameters | QA, QC and other tests as per NABH shall be carried out by OEM or authorized agency of OEM | Yes | 
| Miscellaneous Parameters | Copies of all calibration and test reports to be submitted at the time of Supply of Machine | Yes | 
| Warranty | Warranty in Years | 5 Or higher | 
AI-Powered Bidder Prediction
Companies most likely to bid
Unlock Bidder Insights
AI predictions on likely bidders
Required Documents
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover
Additional Doc 1 (Requested in ATC)
Additional Doc 2 (Requested in ATC)
Additional Doc 3 (Requested in ATC)
Additional Doc 4 (Requested in ATC) *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
   
Similar Tenders
 Digital Radiography System (V2)
DEPARTMENT OF HEALTH AND FAMILY WELFARE
CHANDIGARH, CHANDIGARH
Download FREE!
Access all tender documents at no cost
    Documents  4  
 GeM-Bidding-8498493.pdf
Main Document
CATALOG-Specification-1
CATALOG Specification
Buyer uploaded ATC document
ATC
GEM General Terms and Conditions Document
GEM_GENERAL_TERMS_AND_CONDITIONS
GeM-Bidding-8498493.pdf
Main Document
CATALOG-Specification-1
CATALOG Specification
Buyer uploaded ATC document
ATC
GEM General Terms and Conditions Document
GEM_GENERAL_TERMS_AND_CONDITIONS