Public procurement opportunity for Kamdhenu University Automated HPLC System for Separation of complex mixtures (Q3) in ANAND, GUJARAT. Quantity: 1 issued by. Submission Deadline: 09-03-2026 18: 00: 00. View full details and respond.
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS
State Government
NASHIK, MAHARASHTRA
Indian Army
NORTH DELHI, DELHI
N/a
AKOLA, MAHARASHTRA
Kolhapur District Panchayats
KOLHAPUR, MAHARASHTRA
Kolhapur District Panchayats
KOLHAPUR, MAHARASHTRA
Tender Results
Loading results...
| Category | Specification | Requirement |
|---|---|---|
| Standard parameter | Instrument calibration standards | IEC 17025, NA |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Components of unit | Quaternary Low-Pressure Gradient pump& Parallel Double Plunger, Quaternary Low Pressure Gradient pump & Serial Double Plunger |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Number of solvents delivery | 4 |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Flow rate setting range in ml/min | 0.001 to 10 |
| Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Flow rate accuracy | ± 1 % |
| UV-Vis Detector | UV-Vis Detector provided | Yes, No Or higher |
| UV-Vis Detector | Light source | Deuterium lamp |
| UV-Vis Detector | Spectrum Band width in nm | 4, 5, 6, 8 |
| UV-Vis Detector | Wavelength range in nm | 190 - 700 |
| UV-Vis Detector | Sampling Rate in Hz | 80, 90, 100 |
| UV-Vis Detector | The flow cell must be temperature Controlled | Yes, NA if not provided |
| UV-Vis Detector | Wavelength accuracy in nm | ± 1 |
| UV-Vis Detector | Wavelength reproducibility in nm | ± 0.1 |
| UV-Vis Detector | Drift in AU/Hour | less than 1 x 10-4 |
| UV-Vis Detector | Noise level in AU | ± 2.5 x 10-6, NA if not provided, ±5 x 10-6 |
| UV-Vis Detector | Ability to monitor and quantitative 2 wavelengths simultaneously | Yes |
| UV-Vis Detector | Linearity in AU | 2.5 or more |
| UV-Vis Detector | Temperature control provided for Flow cell | Yes, NA if not provided |
| UV-Vis Detector | Temperature control range for flow cell in Degree Celsius | 5 degree above room temp to 50 degree, NA if not provided |
| UV-Vis Detector | flow cell path length in mm | 10 |
| UV-Vis Detector | Flow cell volume in ?l | 10 - 15 |
| Auto Sample Injector with cooler | Injection Method | Total volume sample injection |
| Auto Sample Injector with cooler | Carry over of sample | 0.025% or lower |
| Auto Sample Injector with cooler | Injection precision in RSD | less than 0.20 % RSD |
| Column Oven | Type of column oven | forced-air-circulation type |
| Software | Operating Windows | Win 10 or latest |
| Photo Diode Array (PDA) Detector | Provision of Photo diode array | Yes Or higher |
| Photo Diode Array (PDA) Detector | PDA detector Wavelength range in nm | 190- 800 or better |
| Photo Diode Array (PDA) Detector | Number of Diodes | 1024 |
| Photo Diode Array (PDA) Detector | PDA detector light source | Deuterium lamp and Tungsten lamp |
| Photo Diode Array (PDA) Detector | PDA detector Wavelength accuracy in nm | ± 1 |
| Photo Diode Array (PDA) Detector | Data Rate in Hz | 100, NA if not provided, <100 |
| Photo Diode Array (PDA) Detector | Noise in AU | less than ± 0.6 x 10-5 or lower |
| Photo Diode Array (PDA) Detector | PDA detector Drift in AU/Hour | less than 1 x 10-3 |
| Photo Diode Array (PDA) Detector | Temperature controlled flow cell in C | 5 degree above room temp to 50 degree, NA if not provided |
| Photo Diode Array (PDA) Detector | Flow cell | Flow cell |
| Warranty & Reports | Warranty (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) | 3, 4, 5 Or higher |
| Warranty & Reports | Number of installations in Central /State/PSU Govt Hospitals | <3, >3 |
| Warranty & Reports | Availability of toll free facility for technical support maintained by OEM or authorized agencies | Yes Or higher |
Discover companies most likely to bid on this tender
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
All India Institute Of Medical Sciences (aiims)
📍 JODHPUR, RAJASTHAN
National Rural Health Mission (2005-12)
Directorate Of Medical Education And Training
📍 Firozabad, UTTAR PRADESH
N/a
📍 RAIPUR, CHHATTISGARH
Hindustan Aeronautics Limited (hal)
Sign up now to access all documents
Main Document
CATALOG Specification
ATC
GEM_GENERAL_TERMS_AND_CONDITIONS