ICU Ventilators (V2)
Hospital Medical Education
AIZAWL, MIZORAM
Bid Publish Date
26-Dec-2025, 4:29 pm
Bid End Date
16-Jan-2026, 5:00 pm
EMD
₹40,000
Value
₹40,00,000
Location
Progress
Quantity
1
Bid Type
Two Packet Bid
Organization: Dr. Ram Manohar Lohia Institute Of Medical Sciences, Lucknow, Uttar Pradesh. Procurement: ICU Ventilators (V2) with extensive patient assessment tools and onboard safety features. Estimated value: ₹4,000,000 with an EMD of ₹40,000. Location: Lucknow, UP 226001. Scope emphasizes advanced ventilation modes, integrated monitoring, and automatic compliance features. Unique aspects include a comprehensive set of patient assessment tools and mandatory integration with hospital-grade compressors and medical air quality per ISO standards. This tender targets suppliers capable of delivering a complete ICU ventilation solution with scalable features and robust clinical data reporting.
ICU ventilator with NIV in all modes, PRVC/Auto Flow, and Autowean
Lung mechanics visualization, RSBI, capnography, esophageal pressure monitoring
Display: touchscreen, minimum ±2% size, graphic wave scaling
RS-232/USB/HL7 outputs; remote alarm display
Compressor: oil-free, ≥150 LPM output; ISO air quality
Humidifier with digital temp monitoring; nebulizer <3 μm
UPS with specified backup time; built-in medical air source
EMD amount and payment terms must be strictly followed
Quantity may be increased up to 25% during contract
Bid must include Technical Specifications and ATC terms uploaded by institute
EMD₹40,000; payment terms aligned with delivery milestones per ATC; security and warranty terms to be finalized in ATC
Delivery period may be extended if option clause exercised; additional time formula: (Increased quantity ÷ Original quantity) × Original delivery period, minimum 30 days
LD terms to be specified in ATC; ensure compliance with delivery and performance benchmarks
Experience in supplying ICU ventilators or equivalent critical care equipment
GST registration and GST compliance in India
Financial capacity to execute multi-unit ICU ventilator supply
OEM authorization for the proposed ventilator brand
Hospital Medical Education
AIZAWL, MIZORAM
Vadodara District Panchayat
N/a
WEST TRIPURA, TRIPURA
Central Health Service Ministry Of Health And Family Welfare
NEW DELHI, DELHI
Tender Results
Loading results...
| Category | Specification | Requirement |
|---|---|---|
| PRODUCT INFORMATION | Patient Type | Adult & pediatric, Adult, Pediatric and Neonatal |
| PRODUCT INFORMATION | Type of technology | Turbine, Compressor |
| PRODUCT INFORMATION | Tidal volume in ml | (50 ml or less) to (1500 ml or more), (2 ml or less) to (1500 ml or more), (20 ml or less) to (1500 ml or more) |
| PRODUCT INFORMATION | Respiration rate, breaths/min | up to 80 or more, up to 120 or more, upto 150 or more |
| PRODUCT INFORMATION | Trigger mechanism | Pressure /flow, Both pressure and flow |
| PRODUCT INFORMATION | Inspiratory flow rate, L/min | upto 120 or more, up to 150 or more, upto 200 or more |
| PRODUCT INFORMATION | Inspiratory pressure, cm H2O | upto 50 or more, upto 80 or more |
| PRODUCT INFORMATION | IE ratio | 1:10 to 4:1, 1:6 to 3:1 |
| PRODUCT INFORMATION | Sigh breath function | Yes, No |
| PRODUCT INFORMATION | Control panel lock | Yes, No |
| PRODUCT INFORMATION | Facility for double lung ventilation | Yes, No |
| PATIENT ASSESSMENT TOOLS | Number of loops | 2 loops (P-V, F-V) with facility of saving of 1 Loops for reference, 2 loops (P-V, F-V) |
| PATIENT ASSESSMENT TOOLS | Lung recruitment tools (PV loops) / OLT (Open lung tools) | Yes, No |
| PATIENT ASSESSMENT TOOLS | Lung mechanics visualization tool | Yes, No |
| PATIENT ASSESSMENT TOOLS | Capnography/CO2 monitoring | Yes, No |
| PATIENT ASSESSMENT TOOLS | Esophageal / transpulmonary pressure monitoring | Yes, No |
| PATIENT ASSESSMENT TOOLS | APRV(Airway Pressure Release Ventilation) /Bivent/Bi level | Yes, No |
| PATIENT ASSESSMENT TOOLS | MMV(Mandatory Minute Ventilation) + PSV (Pressure Support Ventilation) /ASV( Adaptive Support Ventilation ) or ALPV(Adaptive Lung Protection Ventilation) or Automode | Yes, No |
| PATIENT ASSESSMENT TOOLS | PRVC(Pressure-Regulated Volume Control ventilation)/Auto flow | Yes, No |
| PATIENT ASSESSMENT TOOLS | High flow oxygen therapy | Yes, No |
| PATIENT ASSESSMENT TOOLS | Automatic weaning system provided | Yes, No |
| PATIENT ASSESSMENT TOOLS | Volume Support | Yes, No |
| PATIENT ASSESSMENT TOOLS | RSBI (Rapid Shallow Breathing Index) | Yes, No |
| PATIENT ASSESSMENT TOOLS | Expiratory Time constant ( Tcexp) | Yes, No |
| PATIENT ASSESSMENT TOOLS | Compliance static and/or dynamic | Yes, No |
| PATIENT ASSESSMENT TOOLS | Inspired and/or expired resistance | Yes, No |
| PATIENT ASSESSMENT TOOLS | Occlusion pressure | Yes, No |
| PATIENT ASSESSMENT TOOLS | Inspiratory and expiratory hold | Yes, No |
| PATIENT ASSESSMENT TOOLS | Spontaneous frequency | Yes, No |
| PATIENT ASSESSMENT TOOLS | Total peep, intrinsic peep, extrinsic peep | Yes, No |
| PATIENT ASSESSMENT TOOLS | Expiratory block is autoclavable and no routine calibration required | Yes, No |
| PATIENT ASSESSMENT TOOLS | Auto compensation for ET Tube | Yes, No |
| PATIENT ASSESSMENT TOOLS | Facility for automatic compliance & Leakage compensation for circuit available | Yes, No |
| PATIENT ASSESSMENT TOOLS | RS 232 out put port | Yes, No |
| PATIENT ASSESSMENT TOOLS | USB/Memory card | Yes, No |
| PATIENT ASSESSMENT TOOLS | Availability of Remote alarm/display port | Yes, No |
| PATIENT ASSESSMENT TOOLS | Facility to Report (vent alarms and patient status) | Yes, No |
| PATIENT ASSESSMENT TOOLS | HL/7 Compliant out put | Yes, No |
| PATIENT ASSESSMENT TOOLS | Type of coloured display | LED/LCD/TFT |
| PATIENT ASSESSMENT TOOLS | Display should be touch screen | Yes, No |
| PATIENT ASSESSMENT TOOLS | NIV (Non Invasive Ventilation) to be possible in all modes of ventilation available | Yes, No, NIV available in pressure targeted modes only |
| PATIENT ASSESSMENT TOOLS | Minimum Size of display (±2%) | 12, 15, 17, 18 |
| PATIENT ASSESSMENT TOOLS | Graphic display have automatic scaling facility for waves | Yes, No |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Provision of UPS | Yes, No |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Backup time in hrs | 1, 2, NA if not provided |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Built in air source | Air compressor (external), Turbine (In built) |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Availability of stand alone compressor | Yes, No |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Compressor shall provide an oil free Medical air, with Peak output flow should be minimum 150 LPM and Air quality complying with ISO compressed air purity class | Yes, NA (for turbine) |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Seller shall ensure compatibility of compressor with ventilator | Yes, NA (for Turbine) |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Nebulizer with particle size less than 3 micron | Yes |
| ADDITIONAL ACCESSORIES | Humidifier: Servo controlled with digital monitoring of inspired gas temperature with heating wire | 1, 2, NA |
| WARRANTY | Warranty (years) | 3, 5 Or higher |
Discover companies most likely to bid on this tender
GST registration certificate
Permanent Account Number (PAN) card
Experience certificates for ICU ventilator supply
Financial statements (last 3 years or as specified)
EMD submission proof (₹40,000)
Technical bid documents showing compliance with ICU ventilator specifications
OEM authorization / dealership letter for ventilator brand
Product manuals or technical literature demonstrating features
Certificates of ISO/IEC or standard compliance referenced (if available)
Any other bidder qualification documents requested in Terms and Conditions
Key insights about UTTAR PRADESH tender market
To bid, bidders must submit Technical Specifications as uploaded by the institute, OEM authorizations, GST and PAN, ICU ventilator experience, and EMD ₹40,000. Ensure compliance with NIV, PRVC/AutoFlow, and HL7 outputs; delivery terms align with ATC and option clauses allow up to 25% quantity variation.
Required documents include GST certificate, PAN card, last 3 years financial statements, experience certificates in critical care equipment, EMD receipt of ₹40,000, OEM authorization, technical bid, and product literature proving compliance with listed features and standards.
Specifications require NIV capability in all modes, graphic wave scaling, capnography, RSPI, and HL7/RS-232/USB interfaces; display must be touchscreen with ±2% accuracy; compressor output ≥150 LPM; ISO air quality; nebulizer <3 μm particle size; and compliance with IS/ISO requirements per ATC.
EMD is ₹40,000; payment terms follow ATC, with security deposits and milestone payments tied to delivery. Ensure timely submission of EMD along with technical bid to participate; exact payment milestones are defined in the Buyer Added ATC.
The purchaser reserves right to increase quantity by up to 25% during contract; delivery period extends proportionally, calculated as (increased quantity/original quantity) × original delivery period, with a minimum extension of 30 days.
Bidders must provide OEM authorizations for the proposed ventilator; the institute requires technical compliance certificates and may supersede standard GeM terms per ATC; ensure compatibility with hospital compressor and medical air system.
Mandatory tools include capnography/CO2 monitoring, lung mechanics visualization, RSBI, auto compliance, UVC, and manual/automatic weaning support; EN compatibility with DL data reporting and HL7 outputs is required.
Suppliers must include humidifier with digital temperature monitoring, oil-free compressor output ≥150 LPM, built-in medical air source, stand-alone compressor option, and UPS with specified backup time to ensure uninterrupted ICU operation.